RADOBIO AirSafe 1500A2 Class II Type A2 Biological Safety Cabinet

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand RADOBIO
Origin Shanghai, China
Model AirSafe 1500A2
Cabinet Class Class II, Type A2
External Dimensions (L×W×H) 1500×810×2290 mm
Internal Dimensions (L×W×H) 1380×580×798 mm
Filter Efficiency ≥99.9995% @ 0.3 µm
Downflow Velocity 0.35 m/s
Inflow Velocity 0.55 m/s
Noise Level <65 dB(A)
Vibration at Work Surface Center ≤5 µm
Lighting Illuminance ≥900 lx
UV Lamp 30 W × 1
Fluorescent Lamp 18 W × 1
Power Supply AC 220 V, 1φ, 50 Hz
Rated Power (excl. outlet) 0.33 kW
Max. Power (incl. outlet) 1.65 kW
Net Weight 312 kg
Compliance YY 0569–2011, GB 4793.1, GB/T 18268.1
Air Cleanliness Class ISO 4 (equivalent to Fed Std 209E Class 10)
Personnel Protection ≤10 CFU/plate (impactor), ≤5 CFU/plate (slit sampler)
Product Protection ≤5 CFU/plate
Cross-Contamination Protection ≤2 CFU/plate
Filter Type ULPA (Ultra-Low Penetration Air)

Overview

The RADOBIO AirSafe 1500A2 is a certified Class II, Type A2 Biological Safety Cabinet engineered for simultaneous personnel, product, and environmental protection in microbiological and biomedical laboratories. It operates on the principle of laminar airflow combined with recirculation and exhaust filtration: approximately 70% of the inward airflow is recirculated through a ULPA filter after passing across the work surface, while the remaining 30% is exhausted via a dedicated ducted or thimble-connected system after dual-stage ULPA filtration. This configuration meets the stringent performance criteria defined in YY 0569–2011 (Chinese national standard aligned with NSF/ANSI 49 and EN 12469) and ensures ISO 4 (Class 10) air cleanliness within the work area. Designed for routine handling of Risk Group 2 (RG2) agents and select RG3 materials under appropriate containment protocols, the AirSafe 1500A2 integrates aerodynamic stability, real-time monitoring, and energy-efficient operation without compromising safety integrity.

Key Features

  • Intelligent 7-inch touchscreen HMI: Provides unified real-time visualization of sash position, fan runtime, ULPA filter lifespan (% remaining), UV lamp status, lighting control, and power outlet management; supports audit-trail-capable event logging and configurable alarm thresholds.
  • EC brushless DC constant-volume blower system: Delivers stable inflow (0.55 m/s) and downflow (0.35 m/s) velocities independent of filter loading; achieves up to 60% energy reduction versus conventional AC motor systems and incorporates a low-power standby mode (<10% nominal consumption) during sash closure.
  • Ergonomic structural design: Features a 15° inclined sash with easy-clean positioning stops, seamless 304 stainless steel monolithic work surface, adjustable ergonomic armrests, and optimized depth-to-height ratio for prolonged operator comfort and reduced fatigue during extended procedures.
  • Power-loss memory & sash-responsive operation: Automatically transitions to low-speed recirculation mode upon sash closure—maintaining ISO 4 air quality without full shutdown; resumes pre-interruption operational parameters instantly upon sash reopening. In the event of unexpected power failure, the system restores prior setpoints and fan states upon re-energization.
  • ULPA filtration with leak-resistant construction: Dual ULPA filters (supply and exhaust) utilize low-boron, high-strength glass fiber media with ≤0.15 kPa initial pressure drop; proprietary gasketed sealing technology prevents bypass leakage, ensuring validated containment integrity per YY 0569 Annex D testing protocols.
  • Programmable UV sterilization: One-touch scheduled UV activation allows users to define start time and duration; integrated motion-sensing interlock disables UV emission when sash is open or motion is detected within the cabinet volume.

Sample Compatibility & Compliance

The AirSafe 1500A2 accommodates standard laboratory equipment including centrifuges (up to 3000 rpm), pipetting stations, microscopes, and small incubators—provided proper placement maintains unobstructed front intake and rear exhaust grilles. Its airflow dynamics are validated for compatibility with volatile organic compounds (VOCs) at concentrations below 5% LEL when used with external exhaust ducting. Regulatory compliance includes YY 0569–2011 (mandatory for PRC medical device registration), GB 4793.1 (safety requirements for electrical equipment), and GB/T 18268.1 (EMC for in vitro diagnostic devices). While not inherently compliant with FDA 21 CFR Part 11, its event log functionality supports integration into GLP/GMP environments when paired with validated third-party data archiving solutions.

Software & Data Management

The embedded microcontroller firmware records timestamped events—including sash movements, filter runtime, UV cycles, alarm triggers, and manual overrides—with local storage capacity for ≥12 months of operational history. Data export is supported via USB port in CSV format for traceability and internal QA review. Optional RS485 Modbus RTU interface enables integration into centralized building management systems (BMS) or laboratory information management systems (LIMS) for remote status monitoring and preventive maintenance scheduling. All user-accessible settings are protected by multi-level password authentication (operator, technician, administrator).

Applications

This cabinet is routinely deployed in cell culture facilities, clinical microbiology labs, pharmaceutical QC units, and academic biosafety level 2 (BSL-2) teaching laboratories. Typical use cases include aseptic transfer of primary mammalian cells, preparation of sterile media and reagents, handling of genetically modified organisms (GMOs) requiring physical containment, and manipulation of diagnostic specimens containing bloodborne pathogens (e.g., HBV, HCV, HIV). It is not suitable for work involving radionuclides or high-volatility toxic chemicals unless externally vented to a dedicated scrubbed exhaust system.

FAQ

What certification standards does the AirSafe 1500A2 meet?
It complies with YY 0569–2011 (China), which references performance and safety requirements equivalent to NSF/ANSI 49–2022 and EN 12469:2000 for Class II Type A2 cabinets.
Can this cabinet be installed without external ducting?
Yes—it supports both ducted and recirculating (thimble) configurations; however, recirculation mode requires installation in well-ventilated rooms with adequate makeup air supply to prevent negative room pressure.
How often must the ULPA filters be replaced?
Filter service life depends on usage intensity and ambient particulate load; typical replacement intervals range from 3 to 5 years, with real-time monitoring providing % remaining lifespan based on cumulative fan runtime and differential pressure trends.
Is the UV lamp germicidal output validated?
The 30 W low-pressure mercury UV-C lamp (254 nm peak) delivers ≥100 µW/cm² at 15 cm distance; intensity verification is recommended annually using NIST-traceable radiometry as part of routine performance qualification.
Does the unit support remote diagnostics or firmware updates?
Firmware updates require local USB connection and authorized technician access; remote diagnostics are not natively supported but can be enabled via optional Ethernet gateway modules compatible with industrial IoT platforms.