- All
- Favorite
- Popular
- Most rated
| Brand | SIGAS (Schütz Measuring & Control) |
|---|---|
| Origin | Germany |
| Model | PR-Station |
| Inlet Pressure Range | 0.1–1.6 MPa |
| Outlet Pressure Range | 0.1–0.8 MPa (adjustable per circuit) |
| Construction | De-greased Oxygen-Service Copper Tubing (Φ15×1 mm or Φ22×1.2 mm standard |
| Configuration | Dual independent pressure-regulated circuits with isolation valves, digital pressure gauges (0–1.6 MPa), optional explosion-proof enclosure, optional safety valves, optional emergency door release mechanism, optional medical-grade audible/visual alarm with remote telemetry module |
| Brand | SIGAS |
|---|---|
| Origin | Switzerland |
| Model | PAROX 1200 |
| Measurement Principle | Paramagnetic Detection |
| O₂ Range Options | 0–25 %, 0–100 %, or Custom (e.g., 0–x %, x–100 %) |
| Output Signals | 0–1 V, 0–4 V, 4–20 mA |
| Gas Inlet Ports | 3 mm, 1/8", or 5 mm internal thread |
| Flow Requirements | Direct flow (10–150 mL/h), Internal fixed bypass (17–70 L/h), External flow control (10–90 L/h, optional module) |
| Gas Preheating | Optional integrated heater |
| Lifetime | >5 years typical operational service life under specified conditions |
| Cross-sensitivity | Negligible to CO₂, N₂, Ar, CH₄, H₂O (vapor), and common industrial gases |
| Operating Temperature | -10 °C to +50 °C |
| Enclosure Rating | IP65 |
| Compliance | CE, RoHS, ATEX Category 3G (optional variant) |
| Brand | Schutz SIGAS |
|---|---|
| Origin | Germany |
| Model | CoMo 170 Portable High-Sensitivity Alpha/Beta/Gamma Surface Contamination Monitor |
| Detector Type | Plastic Scintillation Detector (Alpha/Beta/Gamma Sensitive) |
| Detection Area | 170 cm² |
| Measurement Units | cps, Bq, Bq/cm² |
| Power Supply | Rechargeable Battery (Portable Operation) |
| Compliance | CE, DIN 6818-1, IEC 60846-1 |
| Optional Accessories | External GM Dose Rate Probe, Desktop Stand, Data Export Software |
| Brand | Schütz SIGAS |
|---|---|
| Origin | Germany |
| Model | SPTr-GAS PORT 200 |
| Type | Portable Multicomponent Gas Analyzer |
| Detection Principle | Photoacoustic Spectroscopy (PAS) |
| Response Time | ≤10 s |
| Repeatability | ±1% FS for SO₂F₂ and CH₃Br |
| Detection Gases | Sulfuryl Fluoride (SO₂F₂), Methyl Bromide (CH₃Br), Phosphine (PH₃) |
| Measurement Ranges | SO₂F₂: 1–150 g/m³ (PRO) / 0.5–200 ppm (CLIR) |
| CH₃Br & PH₃ | matching trace/high-range configurations |
| Accuracy | ±1% FS (full scale) or ±0.5 ppm (<10 ppm) |
| Resolution | 0.1 g/m³ (PRO) / 0.5 ppm (CLIR) |
| Warm-up Time | 5 min (PRO) / 15 min (CLIR) |
| Display | 1×8-character LCD |
| Power Supply | NiMH Battery Pack (19.2 V, 72 Ah) |
| AC Input | 100–240 VAC, 47–64 Hz, max. 110 W |
| Battery Runtime | >26 h (PRO) / >5 h (CLIR) |
| Charging Time | 3.5 h |
| Operating Temperature | +5 °C to +40 °C |
| Storage Temperature | −10 °C to +60 °C |
| Relative Humidity | 0–95% RH, non-condensing |
| Dimensions | 230 × 105 × 270 mm |
| Weight | 3.5 kg (PRO) / 4.3 kg (CLIR) |
| Brand | SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | SGC-100 CEMS-VOC |
| Pricing | Upon Request |
| Brand | SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Model | SA-200 TCD |
| Detection Principle | Thermal Conductivity Detection (TCD) |
| Measuring Range | H₂ 0–99.999 vol% (configurable per background gas) |
| Accuracy | ±2% FS |
| Repeatability | ±1% FS |
| Stability | ±1% FS / 7 days |
| Response Time (T90) | <30 s |
| Display | 240×128 LCD |
| Data Storage | SD card (standard), USB flash drive (optional) |
| Output | 4–20 mA (default), configurable to 0–20 mA / 0–1 V / 0–5 V / 0–10 V / 1–5 V |
| Communication | RS485 (standard), RS232 (optional) |
| Power Supply | 85–264 VAC, 50/60 Hz |
| Power Consumption | <35 VA |
| Operating Temperature | –10 to +45 °C |
| Relative Humidity | <80% RH |
| Sample Flow Rate | 200–400 mL/min |
| Sample Pressure | Ambient ±10% (at outlet) |
| Inlet/Outlet Port | 1/4" NPT internal thread, stainless steel |
| Enclosure Dimensions | 483 × 300 × 177 mm (L×W×H) |
| Mounting | Panel mount, cutout 445 × 180 mm (rectangular), fixing holes Φ6 × 4 at 465 × 101.5 mm |
| Brand | Schütz SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Model | S-ANALYZER 200 V5.0 |
| Instrument Type | Online Fixed-Mount Analyzer |
| Detection Principle | Non-Dispersive Infrared (NDIR) Spectroscopy |
| Number of Channels | Up to 5 Simultaneous Gas Components |
| Measurement Range (Min) | 0–50 ppm |
| Repeatability | ±1% FS |
| Resolution | 0.1 ppm |
| Response Time (T90) | <2 s at 60 L/h flow rate |
| Zero Drift | <±1% FS / 7 days |
| Span Drift | <±1% FS / 7 days |
| Linearity Error | <±2% FS |
| Operating Flow Rate | 0.2–1.2 L/min (±0.02 L/min stability required) |
| Inlet Temperature | 5–35 °C |
| Inlet Pressure | ≤116 kPa (abs) |
| Inlet Humidity | Non-condensing |
| Particulate Load | ≤100 µg/m³, particle size ≤1 µm |
| Display | 5.6" TFT LCD touch screen |
| Analog Output | 4–20 mA (configurable), 2 mA fault signal |
| Digital Communication | RS232 (DB9 female), RS485 Modbus RTU |
| Relay Outputs | 6 configurable dry-contact relays (1 A @ 250 V AC / 2 A resistive) |
| Power Supply | 198–242 V AC, 50/60 Hz, <125 W |
| Enclosure Rating | IP41 (EN 60529) |
| Weight | ~10 kg (net), ~19 kg (packaged) |
| Calibration Gases | Zero gas — 99.999% N₂ |
| Brand | Heyer Aerotech (SIGAS/Schütz Meßtechnik) |
|---|---|
| Origin | Germany |
| Product Type | Medical Gas Outlet Terminal |
| Compliance | DIN EN ISO 7396-1 |
| Mounting Options | Wall-mounted (load-bearing & non-load-bearing), equipment rail, ICU headwall, ceiling-suspended pendants, gas supply boxes from multiple OEMs |
| Gas Interface | Dual-seal, keyed coupling system preventing cross-gas misconnection |
| Certification | CE-marked per EU MDR 2017/745 |
| Brand | SIGAS (Schütz Mess- und Regeltechnik) |
|---|---|
| Origin | Germany |
| Model | Silarex |
| Detection Principle | Dual-Beam Non-Dispersive Infrared (NDIR) |
| Measurement Range | CO₂ up to 20 vol.%, CO up to 1000 ppm, NO up to 2000 ppm, SO₂ up to 1000 ppm |
| Response Time (T90) | < 8 s @ 0.7 L/min |
| Repeatability | ≤ ±1 % FS |
| Linearity Error | ≤ ±1 % FS (standard), ≤ ±1.5 % FS (select versions) |
| Stability | < ±2 % FS over 1000 h |
| Operating Temperature | 0–40 °C (up to 45 °C with thermostatic housing) |
| Storage Temperature | −20–60 °C |
| Humidity | 0–95 % RH (non-condensing) |
| Pressure Range | 800–1150 mbar |
| Flow Rate | 0.2–1.0 L/min (stable) |
| Warm-up Time | < 2 min (cold start), < 30 min (full thermal stabilization) |
| Power Supply | 24 VDC ±0.3 V |
| Max. Current Draw | 1.5 A (heating chamber active) |
| Heating Setpoint | 42 °C (sensor core), 45 °C (with aluminum thermostatic housing) |
| Digital Interface | Modbus ASCII/RTU over TTL, auto-baud & auto-frame |
| Gas Inlet | 3 mm ID / 5 mm OD tubing connection |
| Optical Cell | Gold-coated reflective chamber |
| Cross-Interference Suppression | Engineered spectral deconvolution and multi-wavelength compensation |
| Zero-Point Resolution | CO₂ 0.01 vol.%, CO 0.1 ppm, NO 1 ppm, SO₂ 1 ppm |
| LOD | CO₂ 0.1 %FS, CO 0.5 %FS, NO 1 %FS, SO₂ 1 %FS |
| Brand | SIGAS |
|---|---|
| Model | SPTr-GAS® ANALYZER V3.0 |
| Detection Principle | Photoacoustic Spectroscopy (PAS) |
| Target Gases | NH₃, CH₃Br, C₂H₆O, C₃H₈O, CH₄O, SO₂, SF₆, C₂Cl₄, NF₃, N₂O, C₈H₈, C₂HCl₃F₂, C₂HCl₂F₃, CClF₃, CF₄, CHClF₂, CHF₃, C₂F₆, CCl₂F₂, C₂HCl₃, C₈H₂₀O₄Si, etc. |
| Response Time | < 20 s |
| Detection Limit | < 0.01 % FS (varies by range) |
| Measurement Range | Configurable per application (e.g., NH₃: 0–100 ppm or 0–1 g/m³ |
| CH₃Br | 0–2 g/m³) |
| Accuracy | ±0.1 % FS typical (±0.5 ppm for sub-ppm ranges) |
| Zero Stability | < 0.1 % FS/24 h |
| Operating Temperature | Maintained at 50 °C (±0.5 °C) internally |
| Humidity Compensation | Integrated IR humidity sensor inside PAS cell |
| Display | 5.6” resistive touchscreen LCD |
| Analog Output | 4–20 mA (2 mA fault signal) |
| Digital Interface | RS232 (9600 baud, 8N1) |
| Power | 198–242 V AC, 50/60 Hz |
| Weight | ~15 kg |
| Enclosure Rating | IP42 |
| Compliance | EN60320-C1 power inlet, EN60529, CE-marked |
| Brand | SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Product Origin | Domestic (China) |
| Model | S-Handheld 10 |
| Pricing | Upon Request |
| Detection Principles | Infrared (NDIR), Electrochemical, PID, Catalytic Bead |
| Sensor Lifespan | Electrochemical: 2–3 years |
| NDIR | 5–10 years |
| Catalytic Bead | ~3 years |
| PID | 2–3 years |
| Response Time (T90) | ≤60 s |
| Operating Temperature | 5–45 °C |
| Relative Humidity Range | 10–95 %RH (non-condensing) |
| Barometric Pressure Measurement | 300–1100 hPa (±1 hPa accuracy) |
| Power Supply | 5 VDC, rechargeable 6800 mAh Li-ion battery |
| Runtime | Up to 8 h (pump-on continuous operation) |
| Display | 4.3″ TFT-LCD, 800 × 480 resolution |
| Sampling Method | Integrated active pump, 300 mL/min flow rate |
| Filtration | Built-in hydrophobic particulate filter and water trap |
| Data Storage | 100,000 records (concentration + timestamp + alarm flags) |
| Connectivity | USB Type-C (data transfer & charging), optional Bluetooth 5.0 |
| Dimensions | 217 × 92 × 80 mm (L × W × H) |
| Weight | ~850 g |
| Alarm Types | Low/High, STEL, TWA |
| Certifications | CE, RoHS, IP65-rated enclosure |
| Compliance | Designed for alignment with ISO 10156, IEC 60079-0/-11 (for intrinsic safety configurations), and applicable regional occupational exposure standards (e.g., OSHA PELs, ACGIH TLVs) |
| Brand | Schütz SIGAS |
|---|---|
| Origin | Germany |
| Model | SPTr-GAS PORT 100 |
| Type | Portable Analyzer |
| Response Time | ≤10 s |
| Target Gases | Sulfuryl Fluoride (SO₂F₂), Methyl Bromide (CH₃Br), Phosphine (PH₃), Ethylene Oxide (C₂H₄O) |
| Repeatability | ±2% of reading (for SO₂F₂, CH₃Br, PH₃) |
| Calibration | Auto-zero and manual zero options |
| Display | Integrated touchscreen interface with analog + digital readout |
| Compliance | Designed for field-deployable fumigant residue verification per ISPM No. 18, EN 14603, and EPA Method TO-15 derivations |
| Brand | Schutz SIGAS |
|---|---|
| Origin | Germany |
| Type | Portable |
| Detector | Energy-Compensated Geiger-Müller Tube |
| Radiation Types | Simultaneous γ and X-ray Measurement |
| Display | Dual LCD with backlight (dose & dose rate) |
| Dose Range | 0.5 µSv to 1 Sv |
| Dose Rate Range | 0.5 µSv/h to 50 mSv/h |
| Environmental Dose Equivalent | H*(10) and Ḣ*(10) |
| IP Rating | IP54 |
| Dimensions | 28 × 67 × 105 mm |
| Weight | ~200 g (with batteries) |
| Data Logging | Internal non-volatile memory |
| Alarm Thresholds | 4 factory-preset, user-adjustable |
| External Probe Interface | Yes (e.g., ABG170, ZP1083D, NaI-2002) |
| Brand | SIGAS (Schütz Measurement & Control) |
|---|---|
| Origin | Germany |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Ambulance Mobile Oxygen System |
| Pricing | Available Upon Request |
| Brand | SIGAS |
|---|---|
| Model | SA-TR96/91 |
| Detection Principle | Electrochemical & Frequency-Modulated Ion Current |
| Measurement Range | 0–10 ppm to 0–25% O₂ (selectable) |
| Accuracy | ±1% FS (0–1000 ppm and above), ±3% FS (0–100 ppm), ±5% FS (0–10 ppm) |
| Repeatability & Stability | ±1% FS (≥100 ppm), ±1.5% FS (0–100 ppm), ±2.5% FS (0–10 ppm) |
| Response Time (T90) | <60 s at 25°C |
| Recovery Time (20.94% → 10 ppm) | ≤60 min |
| Operating Pressure | 35–210 kPa (absolute), flow-controlled at 1.0–1.5 L/min (recommended: 1.2 L/min) |
| Ambient Temp. | 0–45°C |
| Humidity | <80% RH |
| Power Supply | 85–264 VAC, 50/60 Hz, <20 VA |
| Output Signals | Isolated 4–20 mA (configurable to 0–20 mA / 0–1 V / 0–5 V / 0–10 V / 1–5 V) |
| Communication | RS485 (standard), RS232 (optional) |
| Gas Inlet | 1/8″ stainless steel |
| Enclosure Dimensions | 144 × 144 × 251 mm (L×W×D) |
| Mounting Hole Pattern | 136 × 136 mm |
| Sensor Lifetime | >2 years (typical use) |
| Analyzer Lifetime | >5 years (typical use) |
| Compliance | Designed for continuous industrial gas monitoring per IEC 61000-6-2/6-4 EMI immunity & emission requirements |
| Certifications | CE-marked |
| Brand | Schütz SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | S-ANALYZER 200 BMBT |
| Price Range | USD 5,600–8,400 (FOB) |
| Response Time | <25 s (at 60 L/h stable flow) |
| Warm-up Time | 3–30 min (at 20°C, dependent on detector configuration) |
| O₂ Resolution | 0.001% (BMBT O₂ channel) |
| CO/CO₂/CH₄ Resolution | 0.1 µmol/mol |
| Display | 5.6" Touchscreen LCD |
| Analog Output | 4–20 mA (normal operation), 2 mA (alarm/preheat) |
| Digital Outputs | 6 configurable relay channels (fault, auto-calibration control, concentration high/low alarms) |
| Communication | RS232 (DB9 female), RS485 Modbus RTU |
| Power Supply | 198–242 V AC, 50/60 Hz, <125 W |
| Enclosure Rating | IP41 (IEC 60529) |
| Weight | ~10 kg (net), ~19 kg (packed) |
| Sample Flow Rate | 0.2–1.2 L/min (±0.02 L/min stability required) |
| Sample Temperature | 5–35 °C |
| Sample Pressure | ≤116 kPa (abs) |
| Sample Humidity | Non-condensing |
| Particulate Load | ≤100 µg/m³, particle size ≤1 µm |
| Zero Gas | 99.999% N₂ |
| Span Gas | 75–125% of full scale |
| Mid-point Calibration Gas | 37.5–62.5% of full scale |
| Brand | SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Model | SPTr-GAS®300 Tracer |
| Type | Portable |
| Response Time | <20 s |
| Repeatability | ±0.1% FS for NH₃, CH₃Br, SO₂, etc. |
| Operating Temperature | 50 °C (thermostatically stabilized) |
| Detection Principle | Photoacoustic Spectroscopy (PAS) |
| Dimensions | 483 mm × 390 mm × 177 mm (19″ 1/2-rack form factor) |
| Weight | ~15 kg |
| Enclosure Rating | IP42 |
| Power Supply | 198–242 V AC, 50/60 Hz |
| Analog Output | 4–20 mA (250–350 Ω load) |
| Communication | RS223 (7-pin blue/black cable) |
| Preheat Time | ~30 min |
| Resolution | 0.1 ppm (typical, gas-dependent) |
| Zero Calibration | Auto or manual air-based calibration |
| Brand | SIGAS (Schütz Mess- und Regeltechnik) |
|---|---|
| Origin | Switzerland |
| Model | Foxsense PAROX 2000 Plus |
| Measurement Principle | High-Sensitivity Paramagnetic Detection |
| Operating Temperature Range | Up to 55°C continuous |
| Enclosure | 19" 3U Rack-Mounted |
| Dimensions (W×H×D) | 482 × 133 × 245 mm |
| IP Rating | IP20 |
| Sensor Core | Integrated Parox1200 Paramagnetic Sensor |
| Calibration | Fully Parameterized Auto-Calibration |
| Channel Configuration | Single or Dual (MULTI2000Plus option) |
| Gas Path Materials | Viton® or PVDF |
| Gas Inlet/Outlet | 4 mm or 6 mm PVDF tubing with stainless steel compression fittings |
| Pressure Compensation Options | None / Electronic / Integrated Pressure Regulator / External Pressure Input (sensor not included) |
| Flow Monitoring Options | None / Mechanical Flowmeter / Electronic Flowmeter with Display & Alarm |
| Oxygen Range Options | 0–25% O₂ |
| Pump Options | None / Integrated Sampling Pump |
| Brand | Schütz Measurement & Control / SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Regional Classification | Domestic (China) |
| Model | SGS-Type Gas Analysis System |
| Pricing | Upon Request |
| Brand | Schuetz SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Product Origin | Domestic (China) |
| Model | SILAREX-TOC |
| Price Range | USD 2,800 – 7,000 |
| Detection Principle | Non-Dispersive Infrared (NDIR), Dual-Beam |
| Measured Gas | CO₂ (as TOC surrogate) |
| Ranges | 0–100 ppm (±2 ppm), 0–1,000 ppm (±20 ppm), 0–10,000 ppm (±100 ppm) |
| LOD | 0.15 ppm |
| Response Time (T90) | <8 s @ 0.7 L/min |
| Repeatability | ≤ ±1 % FS |
| Linearity Error | ≤ ±1 % FS (≤ ±1.5 % FS optional) |
| Stability | < ±2 % FS over 1,000 h |
| Operating Temp. | 0–40 °C |
| Storage Temp. | −20–60 °C |
| Humidity | 0–95 % RH (non-condensing) |
| Pressure Range | 800–1,150 mbar |
| Flow Rate | 0.2–1.0 L/min |
| Warm-up Time | <2 min (cold start), <30 min (thermal equilibrium) |
| Power Supply | 24 VDC ±0.3 V |
| Max. Current Draw | 1.5 A (heating active) |
| Power Consumption | <800 mW (idle), up to 36 W (heating) |
| Digital Output | Modbus ASCII/RTU (TTL, auto-baud 2400–115.2 kbps) |
| Analog Output | Optional 4–20 mA or 0–10 V |
| Internal Heater Setpoint | 42 °C |
| Gas Inlet/Outlet | 3 mm ID / 5 mm OD tubing |
| Chamber Material | Electropolished Aluminum |
| Optical Window | Fluorinated Glass |
| Wavelength | 4.26 µm (2350 cm⁻¹) |
| Brand | SIGAS (Schütz Mess- und Regeltechnik) |
|---|---|
| Origin | Germany |
| Model | AGS 35-50 |
| Supply Configuration | Dual-bank cylinder auto-switching with priority liquid oxygen inlet (AGS 80P compatible) |
| Pressure Monitoring | Continuous real-time differential pressure sensing |
| Control Logic | Automatic switchover triggered by user-defined high/low pressure thresholds |
| Compliance | Designed to meet ISO 8573-1 (compressed air quality), EN ISO 7396-1 (medical gas pipeline systems), and IEC 60601-1 (medical electrical equipment safety) |
| Brand | Schütz SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Model | SPGAS PORT 110 |
| Type | Portable |
| Response Time | <30 s (t₉₀ @ 36 L/h stable flow) |
| Repeatability | H₂S, SO₂F₂, CH₃Br, CO, CO₂, O₂, etc. |
| Measurement Channels | Up to 3 |
| Display | 4.9" TFT LCD |
| Power | Rechargeable Li-ion battery (≥8 h operation @25°C) |
| Communication | RS232 (DB-9F), USB, optional Bluetooth 2.0 (10 m range, BQB/FCC certified) |
| Safety Certifications | UL 62368-1, CSA C22.2 No. 62368-1, TÜV EN 62368-1, CCC GB4943.1 |
| Preheat Time | NDUV ≤1 h, NDIR ≤15 min, EC ≤10 min (cold start @25°C) |
| Sampling Flow | 3 ± 0.5 L/min (integrated low-noise diaphragm pump) |
| Operating Temp | 5–40°C (measurement), 0–45°C (charging) |
| Humidity | 0–95% RH (non-condensing) |
| Pressure Range | 76–116 kPa |
| Sample Gas Requirements | 5–40°C, no condensate, particulates <100 µg/m³ (<1 µm), external filtration recommended |
| Data Storage | Internal memory + USB export (FAT32, up to 16 GB) |
| Brand | SIGAS (Schütz Mess- und Regeltechnik) |
|---|---|
| Origin | Switzerland |
| Model | Foxsense MULTi 2000 Plus |
| Form Factor | Rack-Mount (19" 3U) |
| Channel Configuration | Single or Dual Channel |
| Enclosure Options | IP20 (bare rack unit) or IP52 (enclosed rack version) |
| Dimensions (IP20) | 482 × 133 × 245 mm |
| Dimensions (IP52) | 500 × 143 × 260 mm |
| Response Time (T90) | ≤30 s |
| Measured Gases | CH₄, CO₂, CO, O₂ |
| Gas Sampling | Continuous online flow-through |
| Compliance | Designed for industrial process monitoring and biogas quality control per EN 16726, ISO 8573-1 (compressed air purity), and applicable IEC 61000-6-2/4 EMC standards |
| Regulatory Context | Supports GLP/GMP-aligned data integrity when integrated with audit-trail-enabled SCADA or DCS systems |
| Brand | SIGAS |
|---|---|
| Origin | Germany |
| Model | S-MODULE BasicEVO |
| Detection Principle | Non-Dispersive Infrared (NDIR), Dual-Beam |
| Measurement Range | See Standard Model List |
| Sampling Method | Diffusion |
| Response Time (T90) | ≤30 s |
| Resolution (@zero) | 1 ppm to 0.1 vol.% |
| LOD (3σ) | ≤±1 % FS |
| Repeatability | ≤±1.5 % FS |
| Linearity Error | ≤±2 % FS |
| Zero Stability | ≤±3.0 % FS / 12 months (typ.) |
| Span Stability | ≤±2.5 % FS / 12 months (typ.) |
| Zero Temp. Drift | ≤±0.15 % FS / °C (typ.) |
| Span Temp. Drift | ≤±0.3 % FS / °C (typ.) |
| Pressure Drift | ±0.1–0.2 % of reading per mbar |
| Operating Temp. | –20 °C to +40 °C (model-dependent) |
| Storage Temp. | –20 °C to +60 °C |
| Humidity | 0–95 % RH (non-condensing) |
| Pressure Range | 800–1150 mbar |
| Warm-up Time | <2 min (cold start), <11 min (slow start), <30 min (full spec.) |
| Digital Output | Modbus ASCII/RTU over TTL, auto-baud (2400–115.2 kbps) |
| Analog Outputs | 0/4–20 mA (with S-Connect), 0/0.2–1 V (with S-Connect V), RS-232/RS-485 (with respective S-Connect modules) |
| Supply Voltage | 3.3–6 VDC ±5 % |
| Avg. Current Draw | 120 mA @5 VDC, peak <400 mA @3.3 VDC / <240 mA @5 VDC |
| Power Consumption | <0.8 W (avg.) |
Show next