Jiangsu Schuetz Measurement & Control Equipment Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS
OriginSwitzerland
ModelPAROX 1200
Measurement PrincipleParamagnetic Detection
O₂ Range Options0–25 %, 0–100 %, or Custom (e.g., 0–x %, x–100 %)
Output Signals0–1 V, 0–4 V, 4–20 mA
Gas Inlet Ports3 mm, 1/8", or 5 mm internal thread
Flow RequirementsDirect flow (10–150 mL/h), Internal fixed bypass (17–70 L/h), External flow control (10–90 L/h, optional module)
Gas PreheatingOptional integrated heater
Lifetime>5 years typical operational service life under specified conditions
Cross-sensitivityNegligible to CO₂, N₂, Ar, CH₄, H₂O (vapor), and common industrial gases
Operating Temperature-10 °C to +50 °C
Enclosure RatingIP65
ComplianceCE, RoHS, ATEX Category 3G (optional variant)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS (Schütz Mess- und Regeltechnik)
OriginGermany
ModelSilarex
Detection PrincipleDual-Beam Non-Dispersive Infrared (NDIR)
Measurement RangeCO₂ up to 20 vol.%, CO up to 1000 ppm, NO up to 2000 ppm, SO₂ up to 1000 ppm
Response Time (T90)< 8 s @ 0.7 L/min
Repeatability≤ ±1 % FS
Linearity Error≤ ±1 % FS (standard), ≤ ±1.5 % FS (select versions)
Stability< ±2 % FS over 1000 h
Operating Temperature0–40 °C (up to 45 °C with thermostatic housing)
Storage Temperature−20–60 °C
Humidity0–95 % RH (non-condensing)
Pressure Range800–1150 mbar
Flow Rate0.2–1.0 L/min (stable)
Warm-up Time< 2 min (cold start), < 30 min (full thermal stabilization)
Power Supply24 VDC ±0.3 V
Max. Current Draw1.5 A (heating chamber active)
Heating Setpoint42 °C (sensor core), 45 °C (with aluminum thermostatic housing)
Digital InterfaceModbus ASCII/RTU over TTL, auto-baud & auto-frame
Gas Inlet3 mm ID / 5 mm OD tubing connection
Optical CellGold-coated reflective chamber
Cross-Interference SuppressionEngineered spectral deconvolution and multi-wavelength compensation
Zero-Point ResolutionCO₂ 0.01 vol.%, CO 0.1 ppm, NO 1 ppm, SO₂ 1 ppm
LODCO₂ 0.1 %FS, CO 0.5 %FS, NO 1 %FS, SO₂ 1 %FS
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS
ModelSA-TR96/91
Detection PrincipleElectrochemical & Frequency-Modulated Ion Current
Measurement Range0–10 ppm to 0–25% O₂ (selectable)
Accuracy±1% FS (0–1000 ppm and above), ±3% FS (0–100 ppm), ±5% FS (0–10 ppm)
Repeatability & Stability±1% FS (≥100 ppm), ±1.5% FS (0–100 ppm), ±2.5% FS (0–10 ppm)
Response Time (T90)<60 s at 25°C
Recovery Time (20.94% → 10 ppm)≤60 min
Operating Pressure35–210 kPa (absolute), flow-controlled at 1.0–1.5 L/min (recommended: 1.2 L/min)
Ambient Temp.0–45°C
Humidity<80% RH
Power Supply85–264 VAC, 50/60 Hz, <20 VA
Output SignalsIsolated 4–20 mA (configurable to 0–20 mA / 0–1 V / 0–5 V / 0–10 V / 1–5 V)
CommunicationRS485 (standard), RS232 (optional)
Gas Inlet1/8″ stainless steel
Enclosure Dimensions144 × 144 × 251 mm (L×W×D)
Mounting Hole Pattern136 × 136 mm
Sensor Lifetime>2 years (typical use)
Analyzer Lifetime>5 years (typical use)
ComplianceDesigned for continuous industrial gas monitoring per IEC 61000-6-2/6-4 EMI immunity & emission requirements
CertificationsCE-marked
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS (Schütz Mess- und Regeltechnik)
OriginSwitzerland
ModelFoxsense PAROX 2000 Plus
Measurement PrincipleHigh-Sensitivity Paramagnetic Detection
Operating Temperature RangeUp to 55°C continuous
Enclosure19" 3U Rack-Mounted
Dimensions (W×H×D)482 × 133 × 245 mm
IP RatingIP20
Sensor CoreIntegrated Parox1200 Paramagnetic Sensor
CalibrationFully Parameterized Auto-Calibration
Channel ConfigurationSingle or Dual (MULTI2000Plus option)
Gas Path MaterialsViton® or PVDF
Gas Inlet/Outlet4 mm or 6 mm PVDF tubing with stainless steel compression fittings
Pressure Compensation OptionsNone / Electronic / Integrated Pressure Regulator / External Pressure Input (sensor not included)
Flow Monitoring OptionsNone / Mechanical Flowmeter / Electronic Flowmeter with Display & Alarm
Oxygen Range Options0–25% O₂
Pump OptionsNone / Integrated Sampling Pump
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSchuetz SIGAS
OriginJiangsu, China
Manufacturer TypeAuthorized Distributor
Product OriginDomestic (China)
ModelSILAREX-TOC
Price RangeUSD 2,800 – 7,000
Detection PrincipleNon-Dispersive Infrared (NDIR), Dual-Beam
Measured GasCO₂ (as TOC surrogate)
Ranges0–100 ppm (±2 ppm), 0–1,000 ppm (±20 ppm), 0–10,000 ppm (±100 ppm)
LOD0.15 ppm
Response Time (T90)<8 s @ 0.7 L/min
Repeatability≤ ±1 % FS
Linearity Error≤ ±1 % FS (≤ ±1.5 % FS optional)
Stability< ±2 % FS over 1,000 h
Operating Temp.0–40 °C
Storage Temp.−20–60 °C
Humidity0–95 % RH (non-condensing)
Pressure Range800–1,150 mbar
Flow Rate0.2–1.0 L/min
Warm-up Time<2 min (cold start), <30 min (thermal equilibrium)
Power Supply24 VDC ±0.3 V
Max. Current Draw1.5 A (heating active)
Power Consumption<800 mW (idle), up to 36 W (heating)
Digital OutputModbus ASCII/RTU (TTL, auto-baud 2400–115.2 kbps)
Analog OutputOptional 4–20 mA or 0–10 V
Internal Heater Setpoint42 °C
Gas Inlet/Outlet3 mm ID / 5 mm OD tubing
Chamber MaterialElectropolished Aluminum
Optical WindowFluorinated Glass
Wavelength4.26 µm (2350 cm⁻¹)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS
OriginGermany
ModelS-MODULE BasicEVO
Detection PrincipleNon-Dispersive Infrared (NDIR), Dual-Beam
Measurement RangeSee Standard Model List
Sampling MethodDiffusion
Response Time (T90)≤30 s
Resolution (@zero)1 ppm to 0.1 vol.%
LOD (3σ)≤±1 % FS
Repeatability≤±1.5 % FS
Linearity Error≤±2 % FS
Zero Stability≤±3.0 % FS / 12 months (typ.)
Span Stability≤±2.5 % FS / 12 months (typ.)
Zero Temp. Drift≤±0.15 % FS / °C (typ.)
Span Temp. Drift≤±0.3 % FS / °C (typ.)
Pressure Drift±0.1–0.2 % of reading per mbar
Operating Temp.–20 °C to +40 °C (model-dependent)
Storage Temp.–20 °C to +60 °C
Humidity0–95 % RH (non-condensing)
Pressure Range800–1150 mbar
Warm-up Time<2 min (cold start), <11 min (slow start), <30 min (full spec.)
Digital OutputModbus ASCII/RTU over TTL, auto-baud (2400–115.2 kbps)
Analog Outputs0/4–20 mA (with S-Connect), 0/0.2–1 V (with S-Connect V), RS-232/RS-485 (with respective S-Connect modules)
Supply Voltage3.3–6 VDC ±5 %
Avg. Current Draw120 mA @5 VDC, peak <400 mA @3.3 VDC / <240 mA @5 VDC
Power Consumption<0.8 W (avg.)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSchütz SIGAS
OriginJiangsu, China
Manufacturer TypeAuthorized Distributor
Regional ClassificationDomestic (China)
ModelFLOWEVO
PricingUpon Request
Detection PrincipleNon-Dispersive Infrared (NDIR), Dual-Beam
Gas TargetsMethyl Bromide (CH₃Br), Sulfuryl Fluoride (SO₂F₂)
Response Time (T90)≤12 s @ 0.7 L/min
Resolution (@ zero)1 ppm to 0.1 vol.%
LOD (3σ)≤1% FS
Repeatability≤±1% FS
Linearity Error≤±1.0% FS (standard), ≤±1.5% FS (select versions)
Zero Drift≤±2% FS over 12 months
Span Drift≤±2% FS over 12 months
Temp. Drift (Zero)≤±0.1% FS/°C
Temp. Drift (Span)≤±0.15% FS/°C
Pressure Drift0.1–0.2% reading/mbar
Operating Temp.0–50°C
Storage Temp.−20–60°C
Humidity0–95% RH (non-condensing)
Pressure Range800–1200 mbar
Warm-up Time<2 min (cold start), <30 min (full spec.)
Flow Rate0.1–1.0 L/min (stable)
Digital OutputModbus ASCII RTU (TTL, 2400–115.2 kbaud, auto-baud/auto-frame)
Analog Output (optional w/ FLOWEVO plus controller)4–20 mA, 0–1 V, 0–5 V, 0–10 V DC
Signal OutputTTL-level single-channel
Chamber Pressure CompensationActive
HeatingIntegrated active heating for condensation & drift mitigation
Supply Voltage3.3–6 V DC ±5% (sensor only)
Current Draw<240 mA @ 5 V (avg., no heating)
Power Consumption<800 mW (no heating)
HousingAnodized aluminum
Inlet Connection3 mm ID / 5 mm OD flexible tubing
Expected Service Life≥10 years
Maintenance RequirementMinimal (no consumables, no chemical degradation)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS
ModelS-MODULE EVO
Detection PrincipleNon-Dispersive Infrared (NDIR), Dual-Beam
Measurement RangeConfigurable per gas (e.g., CO₂: 100 ppm to 100% vol
CO2,000 ppm to 100% vol
CH₄5,000 ppm to 100% vol
SF₆50 ppm to 100% vol
SO₂2,000 ppm
N₂O500 ppm)
Operating Temperature0 °C to 50 °C
Storage Temperature−20 °C to 60 °C
Relative Humidity0–95% RH (non-condensing)
Pressure Range800–1200 mbar
Response Time (T90)<3 sec (Fast Mode), <14 sec (Standard Mode)
Zero Drift≤ ±2% FS / 12 months
Span Drift≤ ±2% FS / 12 months
Temperature Drift (Zero)≤ ±0.1% FS/°C
Temperature Drift (Span)≤ ±0.15% FS/°C
Pressure Drift0.1–0.2% of reading per mbar
Power Supply10–26 VDC (recommended 24 VDC)
Digital OutputRS232, RS485, Modbus ASCII/RTU, EVO-proprietary protocols
Analog Output Options0/4–20 mA (isolated), 0–1/2/5/10 VDC
Flow Rate0.1–1.0 L/min (pump-suction)
Tubing Connection3 mm ID / 5 mm OD
HousingAnodized aluminum with integrated thermal insulation and optional active heating
Gas CellGold-coated, temperature-stabilized
Pressure CompensationOptional internal chamber or ambient pressure compensation
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS
ModelS-UltraSense
OriginGermany
Detection PrincipleNon-Dispersive Ultraviolet (NDUV) Absorption
Measurable GasesNO, NO₂, SO₂, O₃, H₂S, N₂O, CnHm
Detection Range Options0–100 ppm to 0–30,000 ppm (gas-specific)
LOD1 ppm (typical, 3σ)
T₉₀ Response Time<3 s @ 1 L/min
Operating Temperature10–40 °C
Power Supply24 VDC ±0.3 V
Average Power Consumption4 W (55 W max during warm-up)
Digital Resolution1 ppm – 0.1 vol.%
Zero Stability±1% FS
Span Stability±1% FS
Linearity±1% FS
Repeatability±1% FS
Temperature Drift±1% FS / 10 K (typ.)
Pressure Sensitivity±0.015% / mbar
Inlet Gas Temp.−2–40 °C
Inlet Humidity Limitdew point ≤10 °C (non-condensing)
Flow Rate0.2–1 L/min
Output InterfacesRS232 (9-pin Sub-D), CANopen (2×5-pin, 2.45 mm pitch)
Heater Setpoint42 °C
Warm-up Time<60 min (environment-dependent), <30 min @ 20 °C ambient
Ready Time after Power-On<2 min (post-warm-up)
Housing MaterialStainless Steel with Protective Coating
Optical Path Length100 mm
Light SourceAlGaN-based UV LED (230–360 nm) with Active Thermal & Current Control
Lifetime>3 years (LED source)
SealingO-ring interconnection between modular components
Cross-Interference ImmunityH₂O-free baseline stability
Power Dissipation<1 W (sensor core only)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS
ModelSPM-O2-MODULE III
Detection PrincipleParamagnetic Oxygen Sensing
Measurement Range0–3 / 0–10 / 0–25 / 0–100 vol.% O₂
Response Time (t₉₀)≤3 s
Zero Drift<±0.1 % O₂/week
Span Drift<±1 % O₂/week
Linearity Error<±1 % FS
Detection Limit (3σ)<1 % FS
Temperature Drift (Zero)<±0.05 % O₂/°C
Flow Rate Dependence<±0.2 % O₂ (20–100 mL/min)
Tilt Sensitivity<±0.02 % O₂/degree
Humidity Influence<0.5 % O₂ @ 20°C
Analog Output4–20 mA
Supply Voltage24 VDC
Inlet Flow Rate100 mL/min (optimal), max. 250 mL/min
Inlet Pressure Drop≤±300 hPa
Operating Temperature5–45 °C
Inlet Gas Dew Point≥5 °C
Particulate Limit≤100 µg/m³, particle size ≤1 µm (dust filter required)
Zero Calibration Gas99.999% N₂
Span Calibration Gas85–110% of full scale (e.g., dry air for 0–25% range)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSIGAS
ModelS-Infra.Sens
Detection PrincipleNon-Dispersive Infrared (NDIR)
T90 Response Time< 3 s @ 1 L/min
Digital Resolution1 ppm to 0.1 vol.%
Accuracy (3σ)±0.5% FS
Repeatability±1% FS
Linearity±1% FS
Zero Stability±1% FS / 30 days
Span Stability±1% FS / 30 days
Zero Temp. Drift±1% FS / 10 K
Span Temp. Drift±1% FS / 10 K
Pressure Drift±0.015% / mbar
Operating Temperature5–45 °C
Storage Temperature−10–60 °C
Relative Humidity0–95% RH (non-condensing)
Ambient Pressure800–1200 mbar
Warm-up Time< 10 min
Power-On Time< 2 min
Sample Gas Temperature+5–35 °C
Sample Gas Humidity≤10 °C dew point (non-condensing)
Flow Rate0.2–1 L/min (stable)
Output InterfacesRS232 (9-pin Sub-D), CANopen, Optional Analog (0–10 V / 4–20 mA)
Supply Voltage24 VDC ±0.3 V
Power Consumption2 W (typ., without heater)
Optical Path Length5–250 mm (modular)
Housing MaterialAluminum or Stainless Steel (optional)
Cell Wall CoatingGold-plated (optional, for enhanced reflectivity)
Response Time (Signal Rise)≈200 ms
Power Consumption (Sensor Only)< 1 W
Show next