Empowering Scientific Discovery

Jiangsu Schuetz Measurement & Control Equipment Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSchütz SIGAS
OriginJiangsu, China
ModelS-ANALYZER 200 V5.0
Instrument TypeOnline Fixed-Mount Analyzer
Detection PrincipleNon-Dispersive Infrared (NDIR) Spectroscopy
Number of ChannelsUp to 5 Simultaneous Gas Components
Measurement Range (Min)0–50 ppm
Repeatability±1% FS
Resolution0.1 ppm
Response Time (T90)<2 s at 60 L/h flow rate
Zero Drift<±1% FS / 7 days
Span Drift<±1% FS / 7 days
Linearity Error<±2% FS
Operating Flow Rate0.2–1.2 L/min (±0.02 L/min stability required)
Inlet Temperature5–35 °C
Inlet Pressure≤116 kPa (abs)
Inlet HumidityNon-condensing
Particulate Load≤100 µg/m³, particle size ≤1 µm
Display5.6" TFT LCD touch screen
Analog Output4–20 mA (configurable), 2 mA fault signal
Digital CommunicationRS232 (DB9 female), RS485 Modbus RTU
Relay Outputs6 configurable dry-contact relays (1 A @ 250 V AC / 2 A resistive)
Power Supply198–242 V AC, 50/60 Hz, <125 W
Enclosure RatingIP41 (EN 60529)
Weight~10 kg (net), ~19 kg (packaged)
Calibration GasesZero gas — 99.999% N₂