- All
- Favorite
- Popular
- Most rated
| Brand | Schütz SIGAS |
|---|---|
| Origin | Jiangsu, China |
| Model | S-ANALYZER 200 V5.0 |
| Instrument Type | Online Fixed-Mount Analyzer |
| Detection Principle | Non-Dispersive Infrared (NDIR) Spectroscopy |
| Number of Channels | Up to 5 Simultaneous Gas Components |
| Measurement Range (Min) | 0–50 ppm |
| Repeatability | ±1% FS |
| Resolution | 0.1 ppm |
| Response Time (T90) | <2 s at 60 L/h flow rate |
| Zero Drift | <±1% FS / 7 days |
| Span Drift | <±1% FS / 7 days |
| Linearity Error | <±2% FS |
| Operating Flow Rate | 0.2–1.2 L/min (±0.02 L/min stability required) |
| Inlet Temperature | 5–35 °C |
| Inlet Pressure | ≤116 kPa (abs) |
| Inlet Humidity | Non-condensing |
| Particulate Load | ≤100 µg/m³, particle size ≤1 µm |
| Display | 5.6" TFT LCD touch screen |
| Analog Output | 4–20 mA (configurable), 2 mA fault signal |
| Digital Communication | RS232 (DB9 female), RS485 Modbus RTU |
| Relay Outputs | 6 configurable dry-contact relays (1 A @ 250 V AC / 2 A resistive) |
| Power Supply | 198–242 V AC, 50/60 Hz, <125 W |
| Enclosure Rating | IP41 (EN 60529) |
| Weight | ~10 kg (net), ~19 kg (packaged) |
| Calibration Gases | Zero gas — 99.999% N₂ |