Huayun GXH-3011 Online NDIR Gas Analyzer (4U Rack-Mount Chassis)
| Brand | Huayun Instruments |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Model | GXH-3011 (4U) |
| Measurement Principle | Non-Dispersive Infrared (NDIR) |
| Target Gases | NO, CO, CO₂, O₂ (configurable per unit) |
| Measurement Ranges | NO: (0–500) to (0–5000) ppm |
| CO | 0–0.500% vol |
| CO₂ | 0–0.500% vol |
| Linearity Error | ≤ ±2% F.S. |
| Repeatability | ≤ 1% F.S. |
| Zero Drift | ≤ ±2% F.S. / 72 h |
| Span Drift | ≤ ±2% F.S. / 72 h |
| Response Time (t₀–t₉₀) | ≤ 15 s |
| Output Signal | 4–20 mA (max 3 isolated channels, load ≤ 500 Ω) |
| Power Consumption | ≤ 120 W |
| Supply Voltage | 220 VAC ±10%, 50 Hz ±10% |
| Enclosure | 19″ 4U rack-mount chassis (360 mm H × 482.6 mm W × 300 mm D) |
| Weight | ≤ 10 kg |
| Operating Temperature | 5–45 °C (non-condensing, no direct sunlight) |
| Relative Humidity | ≤ 80% RH |
| Ambient Airflow | ≤ 0.5 m/s |
| Installation | Horizontal mounting only |
| Electrical Isolation | Dedicated circuit recommended—avoid shared supply with inductive loads (e.g., large motors, RF furnaces) |
Overview
The Huayun GXH-3011 Online NDIR Gas Analyzer (4U Rack-Mount Chassis) is an industrial-grade, continuous emission monitoring system engineered for stable, long-term quantification of key combustion and process gases—including nitric oxide (NO), carbon monoxide (CO), carbon dioxide (CO₂), and oxygen (O₂)—in complex gas matrices. It operates on the well-established principle of non-dispersive infrared (NDIR) absorption spectroscopy: specific diatomic and polyatomic gases exhibit characteristic vibrational-rotational absorption bands in the mid-infrared region (typically 2–14 µm). The analyzer employs matched optical filter pairs and thermopile or pyroelectric detectors to isolate and measure absorbance at precisely defined wavelengths (e.g., 5.3 µm for NO, 4.6 µm for CO, 4.26 µm for CO₂), enabling selective detection without spectral interference from background matrix components such as N₂, H₂O vapor (within specified humidity limits), or CH₄.
Key Features
- Rack-integrated 4U (360 mm height) 19″ chassis compliant with EIA-310-D mechanical standards—designed for seamless integration into centralized control cabinets, CEMS shelters, or laboratory instrument racks.
- Multi-gas configuration flexibility: Each unit can be factory-calibrated for one or up to four target analytes, with independent optical paths and signal processing channels to maintain measurement integrity under mixed-gas conditions.
- High-stability optical bench with temperature-compensated detector array and sealed, purged sample cell—minimizing drift from thermal gradients and particulate deposition.
- Real-time analog output capability: Up to three isolated 4–20 mA current loops (each configurable per gas channel), supporting direct interfacing with PLCs, DCS systems, or SCADA networks without external signal conditioning.
- Robust environmental tolerance: Rated for continuous operation at ambient temperatures from 5 °C to 45 °C and relative humidity ≤ 80% RH; requires minimal airflow (≤ 0.5 m/s) and horizontal, vibration-isolated mounting to preserve optical alignment.
- Low-power architecture (< 120 W) with wide-input AC supply (220 VAC ±10%, 50 Hz ±10%), incorporating surge suppression and line filtering to mitigate noise from shared industrial power infrastructure.
Sample Compatibility & Compliance
The GXH-3011 is intended for use with clean, dry, particle-free sample streams. For field deployment in harsh industrial environments (e.g., flue gas ducts, refinery off-gas lines), optional pre-treatment modules—including heated sample lines, condensate traps, particulate filters (≤ 1 µm), and catalytic NO₂-to-NO converters—are available to meet ISO 12039, EN 15267, and EPA Method 10/10A conditioning requirements. While the analyzer itself does not carry CE or UL certification out-of-box, its electrical design conforms to IEC 61000-6-2 (immunity) and IEC 61000-6-4 (emissions) standards when installed per manufacturer-specified grounding and shielding guidelines. Data integrity aligns with GLP and ISO/IEC 17025 documentation practices; full audit trails (including calibration events, zero/span checks, and parameter changes) are supported when paired with Huayun’s optional data logger firmware.
Software & Data Management
The GXH-3011 operates autonomously without embedded software or touchscreen interface—optimized for reliability in unattended, safety-critical applications. All operational parameters (zero/span values, alarm thresholds, output scaling, and gas-specific calibration coefficients) are configured via front-panel membrane keys or optional RS-485 Modbus RTU interface (addressable up to 247 units on a single bus). Configuration changes are stored in non-volatile EEPROM with write-protection logic to prevent accidental overwrites. When integrated into larger monitoring systems, the 4–20 mA outputs comply with ISA-50.00.01 analog signal standards, and digital communication supports traceable calibration logs required under FDA 21 CFR Part 11 when used in pharmaceutical combustion validation or cleanroom air quality verification.
Applications
- Continuous emissions monitoring (CEMS) in thermal power plants, cement kilns, and waste incinerators—supporting regulatory reporting under national air pollutant emission standards.
- Process gas analysis in petrochemical reformers, ammonia synthesis loops, and syngas purification units where real-time CO/CO₂ ratios inform catalyst health and reaction efficiency.
- Indoor air quality (IAQ) surveillance in subway tunnels, underground parking facilities, and HVAC control systems—detecting CO buildup or ventilation failure indicators.
- Research-grade combustion diagnostics in university energy labs, engine test cells, and biomass boiler development—providing time-resolved gas profiles during transient firing cycles.
- Environmental monitoring of landfill gas collection systems and biogas upgrading facilities, where CO and CO₂ concentrations correlate directly with methane oxidation efficiency.
FAQ
What gases can the GXH-3011 measure simultaneously?
The instrument can be configured for up to four gases (NO, CO, CO₂, O₂) in a single unit, with each channel operating independently using dedicated optical filters and detectors.
Is the analyzer suitable for high-humidity sample streams?
It is designed for samples with ≤ 80% RH at ambient temperature; for saturated or condensing gas streams, a certified sample conditioning system (e.g., chiller + coalescing filter) must be deployed upstream.
Can the 4–20 mA outputs be assigned to different gases?
Yes—each of the three analog outputs is fully assignable and scalable per gas channel during setup, allowing flexible integration with legacy control systems.
What maintenance intervals are recommended?
Zero and span verification is advised every 72 hours for compliance-critical applications; optical window cleaning and filter replacement are typically required every 6–12 months depending on sample cleanliness.
Does the unit support remote diagnostics or firmware updates?
No—the GXH-3011 is a hardwired, microcontroller-based analyzer with no Ethernet, Wi-Fi, or cloud connectivity; all diagnostics are performed via front-panel status LEDs and analog output behavior.