| Brand | All Precision Instrument |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Manufacturer |
| Model | BP-19 |
| Power Supply | 170–240 V AC, 50/60 Hz, 100 W |
| Max Operating Pressure | 7999 mbar |
| Minimum Inlet Pressure | 3000 mbar |
| Test Pressure Range | 500–6900 mbar |
| Pressure Sensitivity | ±1.0 mbar |
| Bubble Point Accuracy | ±60 mbar |
| Dimensions (L×W×H) | 500 × 200 × 180 mm |
| Display | 5.7-inch 32-bit touchscreen |
| Language Options | English / Chinese |
| Data Storage | 32 GB (≥1 million test records) |
| Print Output | Bilingual (English/Chinese), including test parameters, results, and pressure decay curves |
| Environmental Operating Conditions | 0–50 °C, 10–80% RH |
| Test Duration | Bubble point: 2 ± 2 min |
| Diffusion flow | 7 ± 2 min |
| Water intrusion | 7 ± 2 min |
| System leak check | 3 ± 1 min |
| Communication Interface | RS232/USB for PC connectivity and remote supervision via proprietary software |
| Compliance Support | Fully compatible with FDA 21 CFR Part 11 audit trail requirements, EU Annex 1, USP <1207>, ISO 13408-2, and GMP Annex 15 integrity testing protocols |
| Brand | All Precision Instrument |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | OEM Manufacturer |
| Country of Origin | China |
| Model | BQS-19 |
| Power Supply | 170–240 V AC, 50/60 Hz, 100 W |
| Max Test Pressure | 6900 mbar (sensor customizable) |
| Pressure Sensitivity | ±1.0 mbar |
| Bubble Point Accuracy | ±60 mbar |
| Operating Temperature | 0–50 °C |
| Relative Humidity | 10–80% RH |
| Dimensions (W×D×H) | 240 × 320 × 260 mm |
| Display | 5.7–7-inch color touchscreen, 256-color, GUI in English/Chinese |
| Data Storage | 32 GB internal, ≥1 million test records |
| Print Output | Thermal printer, bilingual (English/Chinese), includes test parameters, results, and pressure decay curves |
| Test Duration | Bubble Point: 2 ± 2 min |
| Diffusion Flow | 7 ± 2 min |
| Water Intrusion | 7 ± 2 min |
| System Leak Check | 3 ± 1 min |
| Brand | All Precision Instrument |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | OEM Manufacturer |
| Country of Origin | China |
| Model | BQS-19 Series |
| Power Supply | 170–240 V AC, 50/60 Hz, 100 W |
| Max Operating Pressure | 7999 mbar |
| Minimum Inlet Pressure | 3000 mbar |
| Test Pressure Range | 500–6900 mbar |
| Pressure Sensitivity | ±1.0 mbar |
| Bubble Point Accuracy | ±60 mbar |
| Test Duration (Bubble Point) | 2 ± 2 min |
| Test Duration (Diffusion Flow / Water Intrusion) | 7 ± 2 min |
| System Leak Test Time | 3 ± 1 min |
| Display | 5.7-inch 32-bit touchscreen |
| Data Storage | 32 GB (≥1 million records) |
| Interface | RS232/USB/Ethernet for PC connectivity |
| Compliance Support | FDA 21 CFR Part 11 (audit trail enabled), GMP/GLP-ready |
| Language Options | English & Chinese |
| Dimensions | 500 × 200 × 180 mm (L×W×H) |
| Environmental Operating Range | 0–50 °C, 10–80% RH |
| Brand | All Precision Instrument |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Country of Origin | China |
| Model | BQS-19+ |
| Power Supply | 100–240 V AC, 50/60 Hz, ≤100 W |
| Max Operating Pressure | 9999 mbar |
| Min Inlet Pressure | 4000 mbar |
| Pressure Range for Testing | 100–8000 mbar (1.5–116 psi) |
| Bubble Point Accuracy | ±50 mbar |
| Diffusion Flow Accuracy | ±4% |
| Net Volume Test Accuracy | ±4% |
| Test Duration (Bubble Point) | 15 ± 2 min |
| Test Duration (Enhanced Bubble Point) | 20 ± 2 min |
| Test Duration (Diffusion Flow) | 10 ± 2 min |
| Display | 5.7" monochrome TFT LCD |
| Interface | RS232 serial port |
| Data Storage | 500 test records (configurable extension via PC connection) |
| Environmental Operating Conditions | 5–35 °C, 10–80% RH |
| Dimensions (W × D × H1 × H2) | 400 × 380 × 100 × 235 mm |
| Brand | All Precision Instrument |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Model | BQS-40 |
| Power Supply | 170–240 V AC, 50/60 Hz, 100 W |
| Maximum Operating Pressure | 7999 mbar |
| Inlet Gas Pressure | 3000 mbar |
| Test Pressure Range | 500–6900 mbar |
| Pressure Sensitivity | ±1.0 mbar |
| Bubble Point Accuracy | ±60 mbar |
| Dimensions (L×W×H) | 500 × 200 × 180 mm |
| Display | 5.7-inch 32-bit touchscreen |
| Language Options | English / Chinese |
| Data Storage | 32 GB (≥1 million test records) |
| Compliance | CE certified, FDA 21 CFR Part 11 compliant architecture, GMP-aligned audit trail |
| Enclosure Rating | IP54 |
| Supported Test Methods | Bubble Point, Diffusion Flow, Pressure Hold (Decay), Water Intrusion |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | ME2 |
| Price | USD 198,000 (FOB) |
| Brand | NeuronBC |
|---|---|
| Model | BGT-100 |
| Origin | Beijing, China |
| Power Supply | AC 220 V, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 0–5000 Pa |
| Pressure Resolution | 0.1 Pa |
| Response Threshold | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Dimensions | 370 × 220 × 210 mm |
| Weight | 4.5 kg |
| Data Storage Capacity | ≥10,000 test records |
| External Gas Source | Compressed air, nitrogen, or inert gas (≤0.5 MPa inlet pressure) |
| Audit Trail | Optional |
| Printer | Integrated thermal printer |
| Brand | NeuronBC |
|---|---|
| Model | BGT-120@ |
| Origin | Beijing, China |
| Power Supply | AC 220 V, 50 Hz with built-in Li-ion battery |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 0–5000 Pa |
| Pressure Resolution | 0.1 Pa |
| Response Threshold | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Dimensions | 370 × 220 × 210 mm |
| Weight | 4.5 kg |
| Data Storage Capacity | ≥10,000 test records |
| Compliance | ASTM F2095-01, FDA 21 CFR Part 11 (electronic records & signatures), GMP-aligned validation support (IQ/OQ/PQ/SOP) |
| Brand | NeuronBC |
|---|---|
| Model | BGT-120 |
| Principle | Pressure Decay Method (ASTM F2095-01) |
| Power Supply | AC 220 V, 50 Hz with Li-ion backup |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 0–5000 Pa |
| Pressure Resolution | 0.1 Pa |
| Response Threshold | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 370 × 220 × 210 mm |
| Weight | 4.5 kg |
| Audit Trail | Optional |
| Compliance | ASTM F2095-01, FDA 21 CFR Part 11 (electronic records & signatures), GMP-aligned validation support (IQ/OQ/PQ/SOP) |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | BGT-120 |
| Power Supply | AC 220 V, 50 Hz with built-in Li-ion battery |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 0–5000 Pa |
| Pressure Resolution | 0.1 Pa |
| Response Threshold | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 370 mm × 220 mm × 210 mm |
| Weight | 4.5 kg |
| External Gas Source | Compressed air, nitrogen, or inert gas (≤0.5 MPa inlet pressure) |
| Audit Trail | Optional module compliant with FDA 21 CFR Part 11 requirements |
| Origin | Beijing, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | GT-2.0 |
| Pricing | Available Upon Request |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | GT-2.0 |
| Power Supply | 100–220 VAC, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Response Threshold | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Operating Mode | Continuous |
| External Gas Source | Compressed Air, Nitrogen, or Inert Gas |
| Max Seal Pressure | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 400 × 260 × 250 mm |
| Weight | 7.5 kg |
| Brand | NeuronBC |
|---|---|
| Model | GT-2.0 |
| Type | Online & Offline Glove Integrity Tester |
| Compliance | GB/T 25915.7–2010 / ISO 14644-7:2004 |
| Power Supply | 100–220 VAC, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Pressure Resolution | ≤50 Pa |
| Operating Noise | ≤50 dB(A) |
| Max Seal Pressure | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 400 × 260 × 250 mm |
| Weight | 7.5 kg |
| External Gas Source | Compressed air, nitrogen, or inert gas |
| Mobility | Portable with rapid-connect fittings |
| Interface | Color touchscreen with time/date stamping |
| Regulatory Conformance | 21 CFR Part 11 (electronic records & signatures), audit trail, role-based access control |
| Brand | NeuronBC |
|---|---|
| Model | GT-2.0 |
| Origin | Beijing, China |
| Power Supply | AC 220 V, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Pressure Resolution | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Operating Mode | Continuous |
| External Gas Source | Compressed Air, Nitrogen, or Inert Gas |
| Max Sealing Pressure | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions (L×W×H₁×H₂) | 400 × 260 × 270 × 250 mm |
| Glove Tray Configuration | Up to 6 trays per unit (customizable) |
| Optional Features | Audit Trail, USB Data Export, Printer Integration |
| Compliance | Designed per GB/T 25915.7–2010 / ISO 14644-7:2004, FDA 21 CFR Part 11–ready (with electronic signature and role-based access control) |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | NeuronBC GT-2.0 Online |
| Power Supply | 100–220 VAC, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Pressure Resolution | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Operating Mode | Continuous |
| External Gas Source | Compressed air, nitrogen, or inert gas (max. inlet pressure ≤0.5 MPa) |
| Seal Pressure Capacity | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 400 × 260 × 250 mm |
| Weight | 7.5 kg |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | GT-5.0 |
| Power Supply | AC 220 V, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Pressure Resolution | ≤50 Pa |
| Operating Noise | ≤50 dB(A) |
| Maximum Seal Pressure | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Gas Supply | Compressed air, nitrogen, or inert gas |
| Tray Configuration | Up to 6 glove trays (customizable) |
| Compliance | Designed per GB/T 25915.7–2010 / ISO 14644-7:2004 |
| Regulatory Alignment | 21 CFR Part 11–ready (electronic signatures, audit trail, user access control) |
| Optional Features | USB data export, audit trail module, backup data archiving |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | IntegTest 4.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (115 psi) |
| Test Accuracy | Bubble Point ±50 mbar |
| Bubble Point Range | 590–6900 mbar |
| Diffusion Flow Range | 1–180 mL/min |
| Water Intrusion Range | 0.01–30 mL/min |
| Display | 5.7″ TFT monochrome LCD with backlight |
| Data Storage | 500 test records |
| Interface | RS232 serial port |
| Language Support | English & Chinese |
| Operating Environment | 5–40 °C, 10–80% RH |
| Dimensions (L×W×H₁×H₂) | 400 × 240 × 270 × 220 mm |
| Weight | 7.5 kg |
| Compliance | Suitable for ISO 13408-2, ASTM F838-22, USP <1207>, and GMP/GLP environments requiring filter integrity validation |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | IntegTest 6.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W (optional backup battery) |
| Operating Pressure Range | 100–10,000 mbar (145 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (mm) | 400 (D) × 240 (W) × 270/220 (H1/H2) |
| Weight | 8 kg |
| Bubble Point Range | 100–8,000 mbar |
| Diffusion Flow Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Test Accuracy | Net Volume ±4% |
| Supported Configurations | Symmetric & asymmetric membranes, syringe filters, capsule filters, flat-sheet filters, cartridge filters (up to 9-core, 20-inch), ultrafiltration cassettes & columns, irregular-format filters |
| Pre-stored Protocols | 80 |
| Audit Trail Events | ≥16 logged parameters (user, timestamp, method, result, etc.) |
| Data Storage | >5,000 test records (unlimited with USB backup) |
| Display | 7" high-resolution color touchscreen |
| Interface | RS232, USB |
| Language | English / Chinese |
| Compliance | Fully aligned with USP <1207>, EP 5.1.1, ISO 13408-2, and FDA 21 CFR Part 11 (electronic signatures, role-based access, audit trail) |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | FIT-V4.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (115 psi) |
| Test Methods | Bubble Point, Diffusion Flow, Pressure Hold, Water Intrusion |
| Accuracy | Bubble Point ±50 mbar |
| Test Range | Bubble Point 590–6900 mbar |
| Display | 5.7″ TFT monochrome LCD with backlight |
| Data Storage | 500 test records |
| Connectivity | RS232 serial interface |
| Language | English / Chinese |
| Operating Environment | ISO Class B cleanroom compliant |
| Dimensions (W×D×H) | 240 × 400 × 270 mm (base) + 220 mm (top module) |
| Weight | 7.5 kg |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | BP1.2 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (max. 115 psi) |
| Bubble Point Accuracy | ±75 mbar |
| Diffusion Flow Accuracy | ±4% |
| Bubble Point Range | 590–6900 mbar |
| Diffusion Flow Range | 1–180 mL/min |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (L×W×H) | 400 × 240 × 270 mm (main unit) + 220 mm (height with printer module) |
| Weight | 7.5 kg |
| Display | 3.2″ monochrome LCD |
| Language | Chinese only |
| Data Storage | 20 test records |
| Test Modes | Automated bubble point, semi-automated diffusion flow |
| Filter Compatibility | Single- and multi-cartridge filters (≤3 cartridges, ≤20-inch length), syringe filters, capsule filters, flat-sheet filters, and cylindrical membrane filters (symmetric & asymmetric membranes) |
| Cleanroom Suitability | ISO Class B (or equivalent) environments |
| Regulatory Alignment | Designed to support ASTM F838-22, ISO 13408-2, USP <1207>, and EU Annex 1 requirements for filter validation and integrity testing |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | Integtest V1.2 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (max. 115 psi) |
| Bubble Point Accuracy | ±75 mbar |
| Diffusion Flow Accuracy | ±4% |
| Bubble Point Range | 590–6900 mbar |
| Diffusion Flow Rate Range | 1–180 mL/min |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (L×W×H₁×H₂) | 400 × 240 × 270 × 220 mm |
| Weight | 7.5 kg |
| Display | 3.2″ monochrome LCD |
| Data Storage | 20 test records |
| Language | Chinese only |
| Test Modes | Bubble point (fully automated), diffusion flow (semi-automated), pressure hold (manual setup) |
| Membrane Compatibility | Symmetric and asymmetric membranes |
| Filter Formats | Disc membranes (Φ25–300 mm), standard cartridges (2.5″–40″, 1–15 cores), capsule filters, mini-cartridges, flat-sheet filters, pleated cylindrical filters (≤20″, ≤3 cores), air/gas terminal filters |
| Installation Mode | Offline (benchtop) and online (with optional manifold interface) |
| Cleanroom Class Compatibility | ISO Class 5 (Grade B) and above |
| Compliance Context | Designed to support execution of ASTM F838-22, ISO 13408-2:2018, USP <1207>, EU GMP Annex 1 (2022), and Ph. Eur. 2.6.27 |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | Integtest V4.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (115 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (L×W×H) | 400 × 240 × 270 mm (base) + 220 mm (tower height) |
| Weight | 7.5 kg |
| Bubble Point Accuracy | ±50 mbar |
| Diffusive Flow Accuracy | ±4% |
| Water Intrusion Accuracy | ±0.01 mL |
| Net Volume Measurement Accuracy | ±4% |
| Bubble Point Range | 590–6900 mbar |
| Diffusive Flow Range | 1–180 mL/min |
| Water Intrusion Flow Range | 0.01–30 mL/min |
| Display | 5.7″ monochrome TFT LCD with backlight |
| Data Storage | 500 test records |
| Interface | RS232 serial port |
| Language Support | English / Chinese |
| Test Modes | Manual Bubble Point, Standard Bubble Point, Enhanced Bubble Point, Diffusive Flow, Water Intrusion |
| Printer | Integrated thermal micro-printer |
| Application Scope | Symmetric & asymmetric membranes, syringe filters, capsule filters, flat-sheet filters, cartridge filters (≤3 elements, ≤20 inches) |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | Integtest V4.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (115 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (L×W×H) | 400 × 240 × 270 mm (base) + 220 mm (height extension) |
| Weight | 7.5 kg |
| Display | 5.7″ monochrome TFT LCD with backlight |
| Data Storage | 500 test records |
| Interface | RS232 serial port |
| Language Support | English & Chinese |
| Test Modes | Manual Bubble Point, Standard Bubble Point, Enhanced Bubble Point, Diffusion Flow, Water Intrusion |
| Accuracy | Net Volume ±4%, Bubble Point ±50 mbar, Diffusion Flow ±4%, Water Intrusion ±0.01 mL |
| Measurement Ranges | Bubble Point 590–6900 mbar, Diffusion Flow 1–180 mL/min, Water Intrusion 0.01–30 mL/min |
| Print Output | Integrated thermal micro-printer |
| Filter Compatibility | Symmetric & asymmetric membranes, syringe filters, capsule filters, flat-sheet filters, cartridge filters (≤3 elements, ≤20 inches) |
| Compliance | Designed for ISO 13408-2, ASTM F838-22, and USP <1207> water intrusion method validation |
| Application Context | Offline and online integrity verification in Grade B cleanroom environments |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | Integtest V6.5 |
| Price | USD 11,200 (approx.) |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (14.5–145 psi) |
| Temperature Range | +5 °C to +40 °C |
| Relative Humidity | 10–80 % RH |
| Dimensions (L×W×H) | 400 × 240 × 270 mm (main unit), +220 mm (height with display stand) |
| Weight | 8 kg |
| Bubble Point Accuracy | ±50 mbar |
| Diffusive Flow Accuracy | ±4 % |
| Water Intrusion Accuracy | ±0.01 mL |
| Net Volume Measurement Accuracy | ±4 % |
| Bubble Point Range | 100–8,000 mbar |
| Diffusive Flow Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Pre-stored Protocols | 80 |
| Audit Trail Events | 16+ traceable parameters (user, timestamp, method, pressure ramp, pass/fail status, etc.) |
| Data Storage | >5,000 test records (unlimited with USB backup) |
| Display | 7" high-resolution color touchscreen |
| Interface | RS232, USB, optional industrial bus modules (e.g., Modbus RTU/TCP, Profibus DP, analog 4–20 mA output) |
| Language Support | English & Chinese |
| Compliance | Fully aligned with FDA 21 CFR Part 11 (electronic signatures, role-based access control, audit trail, data immutability) and supports GLP/GMP validation protocols |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V8.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (145 psi) |
| Display | 10-inch true-color capacitive touchscreen |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Enclosure Rating | IP54 |
| Dimensions | 480 × 300 × 190 mm |
| Weight | 12 kg |
| Compliance | CE certified, FDA 21 CFR Part 11 compliant, GMP-aligned audit trail & electronic signature support |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V8.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (145 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Display | 10-inch true-color touchscreen |
| Enclosure Rating | IP54 |
| Weight | 12 kg |
| Dimensions (L×W×H) | 480 × 300 × 190 mm |
| Compliance | CE certified, compliant with EU GMP Annex 1, USP <1207>, EP 5.1.1, FDA 21 CFR Part 11, ISO 13408-2 |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V8.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (145 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Display | 10-inch true-color touchscreen |
| Data Storage Capacity | >300,000 test records |
| Ingress Protection Rating | IP54 |
| Dimensions | 480 × 300 × 190 mm |
| Weight | 12 kg |
| Interface Options | RS232, USB, configurable industrial bus (optional), wireless communication (optional) |
| Regulatory Compliance | CE certified, FDA 21 CFR Part 11 compliant, GMP-aligned audit trail and electronic signature support |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V8.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (max. 145 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Display | 10-inch true-color touchscreen |
| Enclosure Rating | IP54 |
| Dimensions | 480 × 300 × 190 mm |
| Weight | 12 kg |
| Compliance | CE certified, FDA 21 CFR Part 11 compliant, GMP-aligned |
| Data Interface | RS232, USB, configurable industrial bus & wireless options (optional) |
| Maximum Test Configuration | 12-cartridge, 20-inch filter housing |
| Sensor Accuracy | High-precision pressure transducers with <±0.5% FS deviation |
| Audit Trail | Built-in, time-stamped, immutable event logging with user attribution |
| Language Support | English / Chinese |
| Pressure Units | mbar, kPa, psi, kgf/cm² |
| User Management | Four-tier role-based access control with customizable permissions and electronic signature support |
| Brand | NeuronBC |
|---|---|
| Model | JWG-8A |
| Origin | Beijing, China |
| Compliance | Chinese Pharmacopoeia (2015 Edition), USP <788>, USP <789>, ISO 21501-4 |
| Detection Principle | Light Obscuration (LO) / Single-Particle Optical Sizing (SPOS) |
| Sample Volume Range | 0.1–25 mL (adjustable via interchangeable high-pressure syringe pumps) |
| Laser Source | Imported stabilized diode laser (650 nm, Class 1) |
| Particle Size Range | 2–100 µm (calibrated with NIST-traceable polystyrene latex standards) |
| Sensitivity | ≥90% counting efficiency for particles ≥10 µm |
| Flow Cell Material | 316L stainless steel + PTFE-wetted path |
| Stirring System | Variable-speed glass paddle stirrer (0–300 rpm, ±2 rpm control accuracy) |
| Display | 10.1″ capacitive color touchscreen (1280×800) |
| Data Storage | Internal SSD (≥64 GB), supports CSV/PDF export |
| Software | NeuronBC PartiView™ v3.2 (audit-trail enabled in “S” variant), 21 CFR Part 11 compliant modules available |
| Connectivity | USB 3.0, Ethernet, RS-232 |
| Environmental Operation | 5–40 °C, 20–85% RH, altitude-independent performance (0–5000 m ASL) |
| Regulatory Support | IQ/OQ/PQ documentation package available |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | MFT-1000 |
| Detection Principles | Vacuum Decay & Pressure Decay |
| Compliance | USP <1207>, ASTM F2338-22, FDA 21 CFR Part 11, GMP Annex 1 (2022), ISO 15378:2017 |
| Test Range | −0.1 to 1.0 MPa (selectable ranges) |
| Differential Pressure Sensor | ±2 kPa, accuracy ≤0.5% FS, resolution 0.1 Pa |
| Minimum Detectable Equivalent Orifice Diameter | 1–2 µm |
| Display | 10-inch color capacitive touchscreen |
| Power | AC 100–240 V, 50/60 Hz, ≤200 W |
| Enclosure Rating | IP54 |
| Data Retention | ≥10 years (audit trail + event logs) |
| User Management | 4-tier role-based access control with customizable permissions |
| Output Interfaces | 4–20 mA, RS485, Ethernet, USB, 12 V alarm |
| Data Export | Raw test data, metadata, configuration files via USB |
| Optional Connectivity | Wireless module (Wi-Fi/Bluetooth), OPC UA, MQTT support |