IKA RV 10 auto V Starter Solution Rotary Evaporator System
| Brand | IKA |
|---|---|
| Origin | Germany |
| Manufacturer | IKA-Werke GmbH & Co. KG |
| Product Type | Imported Laboratory Equipment |
| Model | RV 10 auto V Starter Solution |
| Evaporation Flask Capacity | 3 L |
| Ultimate Vacuum | 9 mbar |
| Rotation Speed Range | 5–300 rpm |
| Heating Bath Temperature Range | Ambient to 180 °C |
| Temperature Control Accuracy | ±1 K |
| Condenser Cooling Area | 500 cm² (base unit) |
| Heating Bath Volume | 4 L |
| Heating Power | 1350 W |
| Motorized Lift Travel | 140 mm |
| Interface | RS232, Micro-USB, Analog Output |
| Compliance | DIN EN 60529 IP20, CE, RoHS |
Overview
The IKA RV 10 auto V Starter Solution is a fully integrated, laboratory-grade rotary evaporator system engineered for reproducible, safe, and programmable solvent removal in research, QC, and pilot-scale synthesis environments. Based on the proven Couette flow principle—where controlled rotation of the evaporation flask within a heated bath induces thin-film formation—the system maximizes surface-area-to-volume ratio for efficient heat transfer and rapid vaporization under reduced pressure. Its core architecture integrates motorized lifting, vacuum-controlled distillation, real-time temperature monitoring, and intelligent condensation management. Unlike basic manual or semi-automatic systems, the RV 10 auto V operates as a coordinated platform: the built-in vacuum controller (VC 10 lite), variable-speed diaphragm pump (Vacstar lite), and recirculating chiller (RC 2 lite) communicate bidirectionally to maintain process stability across volatile and high-boiling solvents—from ethyl acetate to dimethylformamide (DMF) and N-methyl-2-pyrrolidone (NMP). The system complies with fundamental safety and electromagnetic compatibility standards (CE, RoHS, DIN EN 60529 IP20), and its design supports alignment with GLP documentation workflows through timestamped parameter logging and firmware-updatable control logic.
Key Features
- Triple-layer vertical condenser delivering 1500 cm² effective cooling area—optimized for high-efficiency condensation of low- and medium-volatility solvents
- Motorized lift with 140 mm travel and endpoint positioning—ensures precise, repeatable flask immersion depth and eliminates mechanical impact during descent
- Programmable distillation protocols: users define multi-step vacuum ramps, temperature gradients, rotation profiles (including forward/reverse intervals), and timed hold phases via intuitive TFT color touchscreen
- Integrated vacuum control (VC 10 lite) with auto-clean cycle and adaptive pump speed regulation—reduces energy consumption by >40% during steady-state operation and extends pump service life
- Heating bath with conical geometry, 4 L capacity, and adjustable safety limit (50–190 °C)—minimizes thermal inertia and prevents dry-run conditions via dual-level temperature cutoff
- PTFE-sealed PTFE/stainless-steel spring-loaded joints—resistant to aggressive solvents including halogenated hydrocarbons and strong acids
- Dual communication interfaces (RS232 and Micro-USB) enabling remote operation, firmware updates, and integration with IKA labworldsoft software for audit-trail-compliant data export
Sample Compatibility & Compliance
The RV 10 auto V accommodates standard ISO-K 29/32 and ISO-K 45/50 glassware configurations—including 3 L evaporation flasks, 2 L receiving flasks, and optional bump traps. Its sealed glass train meets DIN 12337 specifications for borosilicate 3.3 glass (e.g., Schott Duran), and all wetted seals conform to USP Class VI biocompatibility requirements. While not certified for GMP manufacturing use out-of-the-box, the system’s deterministic control logic, parameter traceability, and programmable setpoints align with FDA 21 CFR Part 11 readiness when deployed with validated labworldsoft configurations. It supports ASTM D129, ISO 11348, and USP evaporation-related test methods through consistent film thickness control and vacuum stability (<±0.5 mbar drift over 60 min at 15 mbar setpoint).
Software & Data Management
Control and documentation are managed via the embedded 4.3″ TFT interface or externally through IKA labworldsoft (v3.5+), a Windows-based application supporting method import/export, real-time graphing, and CSV/Excel report generation. All operational parameters—including actual vs. setpoint values for rotation speed, bath temperature, vacuum level, and elapsed time—are logged with millisecond-resolution timestamps. The system records up to 100 user-defined methods, each with version history and operator ID fields. Analog output (0–10 V) enables connection to third-party SCADA or LIMS platforms for centralized monitoring. Firmware updates preserve calibration constants and retain user method libraries without requiring reconfiguration.
Applications
This system is routinely deployed in synthetic organic chemistry labs for post-reaction solvent removal prior to crystallization or chromatography; in natural product isolation for gentle concentration of thermolabile extracts (e.g., essential oils, flavonoid fractions); in polymer characterization for residual monomer analysis per ISO 11358; and in pharmaceutical development for API intermediate purification under controlled thermal stress. Its programmable vacuum ramping capability is particularly valuable for fractionated distillation of multi-solvent mixtures, while the reverse-rotation function mitigates foaming during aqueous or surfactant-containing solutions.
FAQ
Does the RV 10 auto V support automatic boiling point detection?
Yes—it uses dynamic vacuum and temperature feedback to estimate solvent boiling points in real time and adjusts distillation parameters accordingly.
Can the system operate unattended overnight?
Yes, provided appropriate safety interlocks (e.g., coolant level, bath temperature limits) are enabled and the receiving flask volume is verified against evaporation rate estimates.
Is the Vacstar lite pump compatible with corrosive vapors?
It is rated for intermittent exposure to common organic solvents but requires an inline cold trap and gas ballast valve for sustained use with acidic or halogenated vapors.
What is the maximum recommended bath fill volume for stable temperature control?
The 4 L heating bath achieves optimal thermal uniformity and response time when filled between 2.5 L and 3.8 L with silicone oil or water-glycol mixture.
How does the system handle bumping during rapid solvent removal?
Bumping suppression is achieved via programmable rotation acceleration profiles, intermittent reverse agitation, and vacuum ramping—eliminating reliance on anti-bumping granules in most applications.