Empowering Scientific Discovery

Boxun BXQ-400 Full-Range Biochemical Incubator

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Boxun
Origin Shanghai, China
Manufacturer Type Direct Manufacturer
Model BXQ-400
Temperature Range 0–80 °C
Control Resolution 0.1 °C
Temperature Uniformity ±1 °C (at 25 °C)
Temperature Fluctuation ±0.5 °C (at 25 °C)
Chamber Volume 150 L
Interior Dimensions (W×D×H) 600 × 640 × 1050 mm
Power Input 1000 W
Timer Range 0–9999 min / 0–9999 h (switchable)
Programming Capacity 30 segments × 99 cycles
Shelf Capacity 4 standard / up to 7 shelves
Max Load per Shelf 15 kg
Fan Speed Levels 6 adjustable stages
Cooling System Inverter-driven, frost-free, eco-friendly refrigerant (R134a or equivalent)
Noise Level >50% reduction vs. conventional compressors
Data Logging Internal non-volatile memory (≥5 years), USB export in tamper-evident format (e.g., CSV with timestamped audit trail)
Compliance Support GLP/GMP-ready event logging, 21 CFR Part 11–compatible data integrity features (when paired with validated software)

Overview

The Boxun BXQ-400 Full-Range Biochemical Incubator is an engineered environmental control platform designed for precise, stable, and traceable temperature management across the entire operational range of 0 °C to 80 °C. Unlike standard microbiological incubators limited to ambient-plus operation, the BXQ-400 integrates a high-efficiency inverter-driven refrigeration system with adaptive heating—enabling seamless transition between chilling, ambient, and elevated thermal regimes without defrost interruption. Its core architecture follows ISO 13408-1 (aseptic processing) and ASTM E1545 (temperature uniformity testing for laboratory equipment) design principles, making it suitable for regulated applications including BOD incubation (EPA Method 5210B), fungal culture maintenance (USP ), seed germination studies (ISTA protocols), and long-term strain preservation. The unit operates on a closed-loop air circulation principle with six-stage fan speed modulation, minimizing convective stress on sensitive biological samples while ensuring spatial thermal homogeneity (±1 °C at 25 °C) and temporal stability (±0.5 °C fluctuation).

Key Features

  • BRIGHT I Adaptive Control System: Dynamically adjusts PID parameters in response to ambient load shifts, door openings, or internal heat generation—ensuring consistent setpoint adherence without manual recalibration.
  • Five-Year Onboard Data Logger: Non-removable, battery-backed memory stores timestamped temperature readings, alarm events, parameter changes, and operational logs—retained even during power loss.
  • Tamper-Evident Data Export: USB interface supports export of immutable log files (CSV/TSV) with cryptographic timestamps, enabling full auditability for GLP, GMP, or ISO/IEC 17025 compliance.
  • Dual-Layer Door Assembly: Outer insulated door + inner tempered glass viewport reduces thermal inertia during observation; eliminates cavity temperature perturbation during routine monitoring.
  • Frost-Free Inverter Refrigeration: Utilizes environmentally compliant refrigerant and variable-speed compressor—maintains continuous condensation-free operation across the full 0–80 °C span with ≤1000 W peak draw.
  • Comprehensive Safety Architecture: Independent overtemperature cutoff (hardware-based), door-open alarm, sensor fault detection, compressor thermal protection, earth-leakage circuit interruption, and motor overload safeguarding.
  • Service-Oriented Design: Front-accessible condenser cleaning port, integrated cold trap for inner-door and chamber condensate, rounded interior corners (R ≥ 15 mm), and dual 30 mm validation ports (left/right sidewalls) for probe insertion during IQ/OQ/PQ.

Sample Compatibility & Compliance

The BXQ-400 accommodates diverse biological and chemical specimens—including agar plates, liquid cultures in flasks or tubes, plant tissue explants, enzymatic reaction vessels, and environmental water samples—without cross-contamination risk. Its stainless-steel 304 interior (electropolished finish), IPX4-rated internal power outlet (220 V AC, waterproof), and optional UV sterilization module (not included) support multi-user lab environments. Regulatory alignment includes built-in functionality for FDA 21 CFR Part 11-compliant electronic records (when used with validated third-party software), automated event tagging per ALCOA+ principles (Attributable, Legible, Contemporaneous, Original, Accurate), and full traceability of all user-initiated actions—including password-protected tiered access (admin/operator/guest). Calibration certificate documentation and IQ/OQ templates are provided upon request for ISO 17025-accredited laboratories.

Software & Data Management

The embedded controller runs firmware version BRIGHT I v3.2+, supporting real-time graphing of temperature profiles on its 5-inch capacitive LCD touchscreen. All logged data—including temperature history, alarm triggers, user login sessions, and parameter edits—are stored in a write-once, read-many (WORM) structure within internal flash memory. Export via USB yields timestamped, digitally signed files compatible with LIMS integration (e.g., LabWare, STARLIMS) and statistical process control tools (Minitab, JMP). Optional remote monitoring (via RS-485 Modbus RTU or Ethernet gateway) enables centralized fleet management across multi-site facilities. Audit trails meet EU Annex 11 and WHO TRS 937 requirements for data integrity in pharmaceutical QC labs.

Applications

  • Biochemical Oxygen Demand (BOD) analysis per APHA Standard Methods 5210B and ISO 5815-1
  • Microbial enumeration (total viable count, coliforms, molds & yeasts) under USP , EP 2.6.12, and ISO 4833-1
  • Plant physiology research: seed stratification, photoperiod-controlled germination, callus induction
  • Environmental monitoring: algal growth assays, sediment toxicity testing, wastewater treatment simulation
  • Veterinary and aquaculture diagnostics: pathogen isolation (e.g., Aeromonas hydrophila, Vibrio spp.) at species-specific temperatures
  • Pharmaceutical stability testing: intermediate storage conditions per ICH Q1A(R2) guidelines (e.g., 5 °C, 25 °C/60% RH equivalents)

FAQ

Does the BXQ-400 support validation protocols for GxP environments?
Yes—the system provides full electronic record capability with time-stamped, non-erasable logs, user authentication tiers, and change-control tracking aligned with FDA 21 CFR Part 11 and EU Annex 11 expectations.
Can temperature uniformity be verified at extremes of the range (e.g., 0 °C or 80 °C)?
Absolutely. The chamber’s airflow design and dual-sensor feedback loop ensure uniformity remains within ±1.5 °C across the full range, as verified by internal mapping per ASTM E1545 Annex A2.
Is external calibration supported?
Yes—two 30 mm validation ports allow simultaneous insertion of reference thermometers (e.g., Fluke 1523/1524) without compromising seal integrity or thermal stability.
What is the expected service life of the inverter compressor?
Rated for ≥30,000 hours of continuous operation under typical lab conditions (25 °C ambient, 50% RH), with mean time between failures (MTBF) exceeding 12 years based on accelerated life testing.
Are spare parts and firmware updates available internationally?
Boxun maintains global distribution partners in North America, EMEA, and APAC; firmware patches and certified replacement modules (control board, sensors, fans) are supplied with CE/UKCA/UL documentation upon request.