AEMIC
Filter
Showing all 2 results
| Brand | AEMIC |
|---|---|
| Model | AE-1136D |
| Origin | Japan |
| Measurement Principle | Ratio-based comparative resistance measurement against external reference thermistor |
| Temperature Compensation | Automatic thermoelectric EMF cancellation |
| Measurement Speed | Ultra-high-speed (7-class sorting capability) |
| Output | Percentage deviation from reference standard (%) |
| Compliance | Designed for industrial QC environments |
| Power Supply | AC 100–240 V, 50/60 Hz |
| Brand | AEMIC |
|---|---|
| Origin | Japan |
| Model | AE-1152D |
| Measurement Principle | 4-Terminal Kelvin (DC) |
| Basic Accuracy | ±(0.01% rdg + 1 μΩ) on 1 mΩ range |
| Minimum Resolution | 10 nΩ |
| Measurement Ranges | 0.0000 mΩ to 1.5000 kΩ |
| Test Currents | 3 A (1 mΩ), 1 A (10 mΩ), 100 mA (100 mΩ–10 Ω), 10 mA (100 Ω–1 kΩ), 1 mA (1 kΩ) |
| Open-Circuit Voltage | ≤5 V |
| Sampling Rate | 2–10 Hz (free-run) |
| Interfaces | Standard GPIB/RS-232C/Centronics (one optional interface built-in) |
| Comparator Outputs | Open-collector (max 40 V, 100 mA) for LO/GO/HI, CONT-E, EOC |
| Operating Environment | 5–40°C, ≤85% RH |
| Power Supply | AC 85–265 V, 50–60 Hz, ~60 VA |
| Dimensions | 333 W × 99 H × 300 D mm (excl. protrusions) |
| Weight | ~3 kg |
| Compliance | CE-marked |