Analysis Tech
Filter
Showing all 10 results
| Brand | Analysis Tech |
|---|---|
| Origin | USA |
| Model | 128-105 |
| Channel Capacity | 128 independent DC resistance monitoring channels |
| Minimum Event Duration | 0.1 / 0.5 / 1.0 µs (field-replaceable clock modules support additional durations) |
| Threshold Resistance Range | 0.1–2000 Ω |
| Sensing Current Options | 1, 5, 10, 20, 50, 100 mA |
| Max Compliance Voltage | 2.5 V |
| Threshold Tolerance | ±5% |
| Event Detection Sensitivity | 100 ns |
| Operating Temperature | ≤35 °C |
| Power Supply | 120 VAC, 60 Hz, 6 A |
| Dimensions | 11.5" H × 18" W × 15.5" D |
| Weight | 51 lb |
| Interface | USB or RS-232C |
| Software | WinDatalog v5.x for data acquisition, event logging, and statistical analysis |
| Compliance | Designed for GLP/GMP-aligned test environments |
| Brand | ANALYSIS TECH |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Import Status | Imported |
| Model | 32-106 |
| Price | Upon Request |
| Channel Count | 32 |
| Threshold Resistance Range | 0.1–2000 Ω |
| Minimum Event Duration | 100 ns (standard settings: 0.1, 0.5, 1.0 µs) |
| Channel Test Current Options | 1, 5, 10, 20, 50, 100 mA |
| Max Channel Source Voltage | 2.5 V |
| Max Sensing Current | 100 mA |
| Threshold Tolerance | ±5% |
| Operating Temperature | ≤35 °C |
| Power Supply | 120 VAC, 60 Hz, 3 A |
| Enclosure Dimensions | 7.5" H × 18" W × 16.5" D |
| Weight | 26 lbs |
| Input Cable Standard | PVC-shielded ribbon cable (rated to 105 °C) |
| Optional Cable | High-temp shielded cable (rated to 180 °C) |
| Compatibility | HLV Amplifier for milliohm-range sensing |
| Brand | Analysis Tech |
|---|---|
| Origin | USA |
| Model | 32-106 |
| Channel Capacity | 32 |
| Minimum Event Duration | 0.1 / 0.5 / 1.0 µs (field-replaceable clock module supports other durations) |
| Threshold Resistance Range | 0.1–2000 Ω |
| Channel Test Current | 1, 5, 10, 20, 50, 100 mA |
| Compliance Voltage | ≤2.5 V |
| Threshold Tolerance | ±5% |
| Interface | USB & RS-232C |
| Software | WinDatalog v5.x |
| Operating Temperature | ≤35 °C |
| Power Supply | 120 VAC, 60 Hz, 3 A |
| Dimensions (H×W×D) | 7.5″ × 18″ × 16.5″ |
| Weight | 26 lb |
| Warranty | 1 year parts and labor |
| Brand | Analysis Tech |
|---|---|
| Origin | USA |
| Model | CTE-360 |
| Temperature Range | –25 °C to +100 °C |
| Compliance | ASTM D696, ASTM E228 |
| Sample Max Cross-Section | 12.5 mm (0.5 in) |
| Recommended Sample Length | 50 mm (2 in) |
| CTE Accuracy | ±3% (full 100 °C range) |
| Power Supply | 100–230 VAC, 50/60 Hz, ≤400 W |
| Dimensions | 43 × 43 × 43 cm (17 × 17 × 17 in) |
| Weight | 9 kg (20 lb) |
| Software | WinCTE v1.x (Windows 7–11, 32/64-bit) |
| Brand | Analysis Tech |
|---|---|
| Origin | USA |
| Model | Phase12 |
| Heating Current Options | 2 A / 10 A / 20 A / 100 A (Phase12B) |
| Sampling Rate | 1 MHz |
| Temperature Resolution | ±0.007°C (Tj digital), ±0.1°C (display), ±0.2–0.5°C (absolute) |
| Thermocouple Accuracy | ±0.15°C (typ.), ±0.3°C (max.) |
| Heating Voltage Accuracy | ±0.2% rdg + 0.025% FS (50 V) for Phase12 |
| Transient Time Resolution | 1 µs |
| Transient Span | 1 µs – 1000 s |
| Brand | Analysis Tech |
|---|---|
| Origin | USA |
| Model | PHASE12 |
| Compliance | JEDEC JESD51-1, JESD51-14, JESD51-8, JESD51-12 |
| Thermal Resolution | ±0.007°C |
| Junction Temperature Accuracy | ±0.1°C |
| Voltage Measurement Accuracy | ±0.25% (0–30 V) |
| Current Measurement Accuracy | ±0.25% |
| Sampling Rate | 1 MHz |
| Max Standard Current Output | 18 A |
| Optional High-Current Power Booster | up to 1000 A |
| Junction Sensing Current Options | 1–50 mA |
| Calibration Temperature Range | Ambient to 180°C |
| Input Power | 220 VAC, 5 A, 50/60 Hz |
| Brand | Analysis Tech |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Product Category | Imported |
| Model | STD Event Detectors |
| Pricing | Upon Request |
| Brand | Analysis Tech |
|---|---|
| Origin | USA |
| Model | STD Event Detectors Accessories |
| Type | Field-Upgradeable Hardware Modules and Interconnect Solutions for Electrical Continuity Interruption Testing |
| Compliance | Designed for use with ASTM F1978, IPC-TM-650 2.6.23, MIL-STD-883H Method 2014, and IEC 60068-2 environmental test correlation workflows |
| Brand | Analysis Tech |
|---|---|
| Origin | USA |
| Model | TIM1300 / TIM1400 |
| Test Method | ASTM D5470 |
| Sample Diameter | 1.3 in (33 mm) |
| Thickness Range | 0–20 mm |
| Temperature Range | 20–65°C (TIM1300), 20–130°C (TIM1400) |
| Contact Pressure Range | 5–380 psi (34–2600 kPa) |
| Thickness Measurement Accuracy | ±1% + 0.001 in (±25 µm) |
| Thermal Resistance Range | 0.005–8 K/W |
| Power Supply | 100–230 VAC, 50/60 Hz |
| Cooling Requirement | 6 L/min @ 10–35°C, ±0.1°C stability |
| Compressed Air | ≥90 psi (600 kPa), 10 µm filtration |
| Dimensions (H×W×D) | 58×41×43 cm |
| Weight | 19 kg (TIM1300), 21 kg (TIM1400) |
| Brand | ANALYSIS TECH |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model Series | TIMTester 1300, TIMTester 1400 |
| Pricing | Upon Request |