| Brand | BAUR (Austria) |
|---|---|
| Origin | Austria |
| Model | Locator Set |
| Output Power (AC Mains) | 0–50 VA |
| Output Power (Battery) | 0–20 VA |
| Max Output Current | 8 A |
| Frequencies | 2 kHz ±0.1% (k-factor ≤1.5% at 4 VA), 10 kHz ±0.1% (k-factor ≤3% at 4 VA) |
| Output Impedance Steps | 1 Ω / 3 Ω / 10 Ω / 30 Ω / 100 Ω / 300 Ω / 1000 Ω |
| Built-in Rechargeable Battery | 12 V, 9.5 Ah |
| Operating Temperature | −20 °C to +50 °C |
| Dimensions (W×H×D) | 290 × 130 × 240 mm |
| Weight | 9.5 kg |
| Input Voltage (Mains) | 110–120 V / 220–230 V / 240 V ±10%, 50–60 Hz |
| DC External Power Input | 12–24 V DC (reverse-polarity protected) |
| Runtime at 20 VA | ~2.5 h |
| Brand | BAUR (Austria) |
|---|---|
| Origin | Austria |
| Model | shirla |
| Operating Principle | Automatic Murray/Glaser Bridge-Based Pre-location & Pulse Current Injection |
| Max Test Voltage | 10 kV |
| Bridge Resistance Measurement Range | 4-wire Kelvin, ±0.1% accuracy |
| Pulse Output | 100 V–10 kV, 700 mA peak, 4 selectable pulse widths |
| Battery Operation | Internal rechargeable battery, full functionality on battery |
| Environmental Rating | -20°C to +50°C operating, IP2X enclosure |
| Dimensions | 440 × 490 × 220 mm |
| Weight | <20 kg (incl. accessories) |
| Compliance | IEC 61000-4 series EMC, IEC 61010-1 Safety, EN 61326-1 |
| Reporting | Automated test report generation with USB 2.0 export |
| Cable Segment Definition | Up to 50 user-defined segments (length, conductor cross-section, material) for distance compensation |
| Leakage Compensation | Automatic detection and offset correction during tanδ and insulation resistance measurement |
| Display | Backlit 320 × 240 pixel LCD with auto-brightness adjustment |