Formulaction
Filter
Showing all 8 results
| Brand | Formulaction |
|---|---|
| Origin | France |
| Model | Turbiscan AGS |
| Particle Size Range | 0.05–1000 µm |
| Concentration Range | up to 95% v/v |
| Temperature Control | 25–60 °C (±0.5 °C) |
| Sample Capacity | 54 positions (3 × 18-well thermostated modules) |
| Scan Resolution | 40 µm along 55 mm height |
| Scan Interval | as low as 20 s per scan |
| Light Source | Pulsed NIR LED (880 nm) |
| Detectors | Synchronized transmission and backscattering photodiodes |
| Compliance | ASTM D7809, ISO 13321, USP <729>, GLP-ready data audit trail |
| Brand | Formulaction |
|---|---|
| Origin | France |
| Model | TURBISCAN LAB |
| Measurement Principle | Static Multiple Light Scattering (MLS) |
| Wavelength | 880 nm (pulsed NIR source) |
| Scan Height | 55 mm |
| Spatial Resolution | 20 µm |
| Temp. Range | +5 °C above ambient to 60 °C |
| Temp. Control Accuracy | ±0.5 °C |
| Particle Size Range | 0.01–1000 µm |
| Max. Concentration | 95% v/v |
| Detectors | Dual photodiode (transmission + backscattering) |
| Sample Cell | Cylindrical glass cell (Ø 27.5 mm × H 55 mm), threaded black cap with butyl/PTFE seal |
| Data Acquisition Modes | Full-height scanning (1 scan/20 s) or fixed-position kinetics (1 measurement/s) |
| Max. Scan Programs | 250 |
| Software Output | BS/T vs. time, layer thickness vs. time, particle migration velocity, hydrodynamic diameter (d), volume fraction (φ), mean free path, TSI Stability Index, dispersion index |
| Brand | Formulaction |
|---|---|
| Origin | France |
| Model | TURBISCAN TOWER |
| Detection Principle | Multiple Light Scattering (MLS) |
| Wavelength | 880 nm (pulsed NIR source) |
| Sample Capacity | 6 positions |
| Temperature Range | 4–80 °C |
| Temp. Control Accuracy | ±0.1 °C |
| Scan Height | 55 mm |
| Spatial Resolution | 20 µm per data point |
| Scan Interval | ≥20 s (user-configurable) |
| Particle Size Range | 0.01–1000 µm |
| Max. Concentration | 95% v/v |
| Measurement Modes | Transmission (T) and Backscattering (BS) |
| Software Compliance | FDA 21 CFR Part 11 ready, audit trail enabled, GLP/GMP-supportive workflows |
| Brand | Formulaction |
|---|---|
| Origin | France |
| Model | TRI-LAB |
| Light Source | 880 nm NIR LED |
| Sample Volume | 1.5–30 mL |
| Temperature Range | 20–60 °C |
| Sample Positions | 1–3 |
| Concentration Range | 0.0001–95% v/v |
| Detection Principle | Synchronized Transmittance (T) and Backscattering (BS) Profiling |
| Measurement Mode | Non-invasive, label-free, no dilution required |
| Compliance | Designed for GLP/GMP-aligned pharmaceutical stability assessment |
| Brand | Formulaction |
|---|---|
| Origin | France |
| Model | TURBISCAN TRI-LAB |
| Light Source | 880 nm NIR LED |
| Sample Volume | 1.5–30 mL |
| Temperature Range | 20–60 °C |
| Sample Capacity | 1–3 positions |
| Concentration Range | 0.0001–95% v/v |
| Measurement Principle | Static Multiple Light Scattering (SMLS) |
| Detection Modes | Simultaneous Transmitted (T) and Backscattered (BS) Light |
| Output Metrics | Turbiscan Stability Index (TSI), Migration Rate, Phase Thickness, Kinetic Index, Mean Hydrodynamic Diameter, Dispersion Ratio, Redispersion Efficiency |
| Brand | Formulaction |
|---|---|
| Origin | France |
| Model | TURBISCAN DNS |
| Measurement Principle | Static Multiple Light Scattering (SMLS) |
| Detection Modes | Transmittance (T) and Backscattering (BS) |
| Sample Volume | 1.5–30 mL |
| Concentration Range | 0.0001–95% v/v |
| Particle Size Range | 10 nm–1 mm |
| Spatial Resolution | Vertical scanning with high-frequency acquisition |
| In-situ Options | T-MIX (stirred mode), T-LOOP (circulation mode) |
| Output Metric | Turbiscan Stability Index (TSI) |
| Brand | Formulaction |
|---|---|
| Origin | France |
| Model | DNS |
| Measurement Principle | Static Multiple Light Scattering (SMLS) |
| Sample Volume | 1.5–30 mL |
| Concentration Range | 0.0001–95% v/v |
| Particle Size Range | 10 nm–1 mm |
| Detection Modes | Transmitted Light (T) & Backscattered Light (BS) |
| Scan Mode | Vertical high-frequency scanning of static sample |
| Optional Functions | T-MIX (in-situ stirring), T-LOOP (circulation coupling) |
| Output Metric | Turbiscan Stability Index (TSI) |
| Compliance Context | Designed for GLP-compliant formulation development and stability assessment per ICH Q1/Q5, USP <729>, ISO 13321 |
| Brand | Formulaction |
|---|---|
| Origin | France |
| Model | Turbiscan TMIX |
| Measurement Range | 0.05–1000 µm |
| Repeatability | ≤1% |
| Accuracy | ≤1% |
| Measurement Interval | 20 s |
| Spatial Resolution | 40 µm |