| Brand | FTT |
|---|---|
| Origin | United Kingdom |
| Model | ABD0031 |
| Application | Integrated with NBS-type smoke density chamber for aircraft cabin material qualification |
| Core Components | Vacuum chamber, vacuum pump, manual metering pump, gas analysis detector tubes |
| Compliance Context | Airbus ABD0031 specification, ASTM E662, ISO 5659-2, FAA AC 20-135, EN 45545-2 |
| Brand | FTT |
|---|---|
| Origin | United Kingdom |
| Model | FRP |
| Standards Compliance | EN ISO 9239-1, ASTM E648, ASTM E970, NFPA 253, GB/T 11785 |
| Radiation Source | Gas-fired flat-panel radiator inclined at 30° |
| Radiant Heat Flux Range | 1.1–10.9 kW/m² (extendable to 25.0 kW/m²) |
| Critical Radiant Flux Resolution | 0.1 kW/m² |
| Sample Orientation | Horizontal |
| Ignition | Pilot burner with stainless steel nozzle |
| Smoke Measurement | DIN 50055-compliant optical smoke density system mounted on exhaust duct |
| Dimensions (W×D×H) | 1900 × 750 × 1900 mm |
| Construction | Stainless steel hood, calibrated angular positioning plate, dual-access observation door with sliding platform |
| Brand | FTT |
|---|---|
| Origin | United Kingdom |
| Model | iCone |
| Compliance | ISO 5660-1 & -2, ASTM E1354, ASTM E1740, ASTM F1550, ASTM D5485, ASTM D6113, NFPA 271, NFPA 264, CAN/ULC-S135, BS 476-15, GB/T 16172 |
| Heat Flux Range | Up to 100 kW/m² |
| Sample Size | 100 mm × 100 mm (max thickness 50 mm) |
| Mass Loss Resolution | 0.01 g |
| Exhaust Flow Rate | 24 L/s |
| Oxygen Analysis | Paramagnetic or Zirconia-based O₂ sensor |
| Smoke Obscuration | He–Ne laser (0.5 mW), Si photodiode detector with ND filters (0.3 & 0.8) |
| Ignition Source | 10 kV spark igniter with auto-positioning and safety cutoff |
| Data Acquisition | Synchronized multi-channel, 100 Hz sampling rate |
| Software | ConeCalc v6.x (Windows-based, multilingual UI) |
| Brand | FTT |
|---|---|
| Origin | United Kingdom |
| Model | NBS |
| Compliance | ASTM E662, ASTM F814, ISO 5659-2, NFPA 258, BS 6401, GB/T 8323, IMO FTPC Part 2, NES 711 |
| Chamber Dimensions | 914 mm × 914 mm × 610 mm (L×H×W) |
| Radiant Heat Flux Range | 10–50 kW/m² (calibrated at 25 kW/m² ±1 kW/m²) |
| Optical System | 6.5 V Tungsten Lamp + S-4 Spectral Response Photomultiplier Tube (dark current <10⁻⁹ A) |
| Light Beam Transmission Accuracy | 0.01% |
| Optical Density Resolution | 0.01 |
| Mass Loss Measurement Accuracy | ±0.1% |
| Timer Resolution | 0.1 s |
| Cone Heater Power | 2600 W |
| Thermocouples | Dual K-type armored (Cu/Alumel) |
| Heat Flux Meter | Thermopile-based, calibrated to ±1 kW/m² at 25 & 50 kW/m² |
| Ignition | Horizontal single-flame burner (30 mm ±5 mm), spark-ignited without chamber opening |
| Brand | FTT |
|---|---|
| Origin | United Kingdom |
| Model | Oxygen Bomb Calorimeter |
| Standards Compliance | ISO 1716, ISO 1928, ASTM D240, ASTM D4809, ASTM D5468, ASTM D5865, ASTM E711 |
| Application Scope | Gross Calorific Value (GCV) determination of solid, liquid, and heterogeneous fuels and combustible materials |
| Configuration | Includes main calorimeter unit, high-pressure oxygen bomb assembly, oxygen charging station, calibration accessories (benzoic acid standard, ignition wire, crucibles), and installation kit |
| Brand | FTT |
|---|---|
| Origin | United Kingdom |
| Model | Radiant Panel Flame Spread Apparatus |
| Compliance Standards | ASTM E162, ASTM D3675 |
| Radiation Panel Dimensions | 12 in × 18 in (305 mm × 457 mm) |
| Panel Material | Porous cement and cast iron |
| Ignition System | Electronic spark igniter with auto-shutdown safety monitoring |
| Sample Holder | Stainless steel, calibrated at 3-inch (76 mm) intervals for flame front observation |
| Thermocouple Array | Eight calibrated stainless steel thermocouples integrated into exhaust duct |
| Surface Temperature Measurement | Calibrated high-temperature pyrometer with mounting fixture |
| Gas Control | Precision methane flowmeter and adjustable gas mixture valve |
| Exhaust System | Removable-panel stainless steel ducting for maintenance access |
| Safety Systems | Interlocked gas shutoff valve and circuit-breaker activated by flame failure or overtemperature |