| Brand | HEYL (Neomeris) |
|---|---|
| Origin | Germany |
| Model | Testomat ECO / Testomat 2000 |
| Measurement Parameter | Total Calcium and Magnesium Hardness (as CaCO₃) |
| Power Supply | 230 V or 24 V ±10 %, 50–60 Hz |
| Power Consumption | ≤30 VA |
| Ingress Protection | IP65 |
| Electrical Safety Class | Class I |
| Compliance | EN 50081-1, EN 50082-2, EN 61010-1 |
| Operating Ambient Temperature | 10–45 °C |
| Sample Inlet Temperature | 10–40 °C |
| Operating Pressure Range | 0.1–3 bar (optional pressure regulator: 1–8 bar) |
| Dimensions (W×H×D) | 380 × 480 × 280 mm |
| Weight | ~9.5 kg |
| Display | Backlit LCD with text-based menu navigation |
| Output Signals | Two programmable relay outputs |
| Maintenance Alert Output | Yes |
| Analog Output | 4–20 mA |
| Mounting | Wall-mounted |
| Units Supported | °dH, °f, ppm CaCO₃, mmol/L |
| Measurement Principle | Colorimetric photometric analysis using EDTA-based titration chemistry |
| Optional Configurations | Dual-channel measurement (Testomat 2000 only) |
| Sampling Frequency Modes | Time-based interval, pulse-triggered (e.g., flow meter signal), volume-proportional, or factory preset cycle |