| Brand | Hinds Instruments |
|---|---|
| Origin | USA |
| Model | DV-193 |
| Wavelength Range | 180–200 nm (optimized at 193 nm) |
| Measurement Principle | Dual-PEM Modulated Photoelastic Polarimetry |
| Stokes Parameter Acquisition Rate | >100 Hz |
| Accuracy (180–200 nm) | ±1% |
| Sensitivity | <0.0005 |
| Repeatability (3σ, 180–200 nm) | ≤0.01 |
| Repeatability (3σ, 200–400 nm) | ≤0.0035 |
| PEM Resonant Frequency | 37–100 kHz |
| Compliance | ASTM E284, ISO 15382, USP <857>, GLP/GMP-ready data audit trail |
| Brand | Hinds Instruments |
|---|---|
| Origin | Shanghai, China |
| Model Variants | YP-SD150, YP-SD250, YP-SD500, YP-SD1000, YP-SDP150, YP-SDP250, YP-SDP500, YP-GSP150, YP-GSP250, YP-GSP500 |
| Capacity | 150 L / 250 L / 500 L / 1000 L |
| Temperature Range | 0–65°C (dark), 10–65°C (with illumination) |
| Temperature Uniformity | ±0.5°C |
| Humidity Range | 30–95% RH |
| Humidity Uniformity | ±0.5% RH |
| Illuminance Range | 0–6000 lx (adjustable) |
| Illuminance Accuracy | ≤ ±500 lx |
| Control Resolution | 0.1°C / 0.1% RH |
| Refrigeration System | Hermetic Danfoss compressor (R404A/R23 refrigerant) |
| Cooling Capacity | 90–150 W @ −25°C to +55°C |
| Power Supply | AC 220 V ±10%, 50 Hz |
| Safety Protections | Compressor overheat, overload, water shortage, overtemperature, electrical short-circuit |
| Compliance | Chinese Pharmacopoeia (2010 Edition) Stability Guidelines, GB/T 10586–2018 Environmental Test Equipment General Specifications |
| Construction | Mirror-finish SUS304B interior, arc-shaped cabinet design, double-glazed vacuum-insulated observation window |
| Air Circulation | Optimized ducted convection system for high spatial uniformity |
| Standard Accessories | 2 adjustable stainless-steel shelves, left-side test lead port, PID-tuned Fuji controller with touchscreen interface |