Ion Science
Filter
Showing the single result
| Brand | Ion Science |
|---|---|
| Origin | UK |
| Model | TIGER Select |
| Detection Principle | Photoionization Detection (PID) with 10.0 eV UV Lamp |
| Resolution | 10 ppb (benzene) |
| Range | 0–200 ppm (with benzene pre-filter), 0–20,000 ppm (TAC mode, gas-dependent) |
| Response Time (T90) | <2 s |
| Accuracy | ±10% of reading ±1 digit (benzene) |
| Linearity | ±5% of reading ±1 digit |
| Intrinsically Safe Certification | II 1G Ex ia II C T4 Ga (Ta = –15 °C to +45 °C) |
| Battery Life | Up to 24 h (rechargeable Li-ion), 8.5 h (3×AA alkaline) |
| Data Logging | >120,000 timestamped records |
| Calibration | 2-point and 3-point via certified calibration kit |
| Alarm | Visual (amber/red LED), audible (95 dBA @ 300 mm), vibratory (pre-programmed TWA/STEL) |
| Flow Rate | 220 mL/min (low-flow alarm at 150 cc/min) |
| Operating Temp | –20 °C to +60 °C |
| Humidity | 0–99% RH (non-condensing) |
| IP Rating | IP65 |
| EMC Compliance | EN 61326-1:2006, EN 50270:2006, CFR 47 Part 15 Class A |
| Dimensions (w/o filter) | 465 × 89 × 61 mm |
| Weight (instrument only) | ~0.55 kg |
| packaged weight | 5 kg |
