Konica Minolta
Filter
Showing 1–30 of 35 results
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CA-100Plus |
| Measurement Aperture | Φ21 mm |
| Luminance Range (Standard Detector) | 0.05–1000 cd/m² |
| Luminance Range (High-Luminance Detector) | 0.05–2000 cd/m² |
| Chromaticity Accuracy (White, ≥0.50 cd/m²) | ±0.003 |
| Luminance Accuracy | ±2% ±1 digit |
| Repeatability (Luminance) | 0.2% + 1 digit (2σ) |
| Measurement Speed | Up to 20 readings/sec (xyLv) |
| Interface | USB 1.1 or RS-232C (up to 38400 bps) |
| Storage Channels | 100 |
| Operating Temperature | 10–28°C, RH ≤70% (non-condensing) |
| Power Supply | 100–240 VAC, 50–60 Hz, 50 VA |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CA-210 |
| Photometric Aperture | Φ27 mm |
| Viewing Angle | ±2.5° |
| Luminance Range | 0.01–1000 cd/m² |
| Luminance Accuracy (White) | ±2% + 1 digit |
| Chromaticity Accuracy (White) | ±0.008 (0.10–4.99 cd/m²) to ±0.002 (160 cd/m²) |
| Repeatability (2σ) | 0.001–0.2% + 1 digit |
| Measurement Speed | 5–20 Hz (xyLv) |
| Interface | USB 1.1 / RS-232C (up to 38.4 kbps) |
| Operating Temp. | 10–28°C, RH ≤70% (non-condensing) |
| Dimensions (Main Unit) | 340 × 127 × 216 mm |
| Weight (Main Unit) | 3.58 kg |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CA-2500 |
| Minimum Measurable Luminance | 0.05 cd/m² |
| Luminance Range | 0.05–5,000,000 cd/m² (with lens-dependent dynamic range extension) |
| Sensor Type | XYZ-filtered CCD with CIE 1931 spectral responsivity match |
| Lens Options | Standard, Wide-angle, Telephoto (with macro coupling ring) |
| Synchronization Frequency Range | 4–2,000 Hz |
| Factory Calibration | Per-lens, per-focus-position optical calibration |
| Included Software | CA-S25w (with autofocus assistance, auto-alignment guidance, and automatic ROI detection) |
| SDK Included | Yes, for custom application integration |
| Compliance | Designed for ISO/CIE-compliant photometric and colorimetric evaluation in R&D and production QA environments |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CA-2500 |
| Type | Portable Color Analyzer |
| Light Source | Standard White Reference Illuminant |
| Luminance Repeatability | ±0.5% |
| Chromaticity Repeatability | ±0.001 |
| Measurement Range (with ND filter) | 0.05–100,000 cd/m² |
| Sensor | 2/3-inch Monochrome CCD, 1000 × 1000 Pixels, Integrated XYZ Tristimulus Filter (CIE 1931 Match) |
| Resolution Options | 980 × 980, 490 × 490, or 196 × 196 (via CA-S25w Software) |
| Interface | USB 2.0 |
| Operating Temperature/Humidity | 10–30°C, ≤70% RH (non-condensing) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CA-410 |
| Probe Options | CA-VP427A / CA-VP410A (down to 0.0003 cd/m²) |
| CIE Compliance | CIE 1931 2° Standard Observer, optional CIE 170-2:2015 2° Standard Observer (CA-P427C / CA-P410C probes) |
| Interface | USB (virtual COM port, driver-free), RS-232C |
| Auto-Zero Calibration | Yes |
| Synchronization | External sync input (1.8 V TTL-compatible, from March 2021 production onward) |
| Software Support | CA-SDK2 (backward-compatible with CA-210/CA-310 systems) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CL-200 |
| Product Type | Color Difference Meter |
| Measurement Principle | Tristimulus-based photometric and colorimetric sensing (CIE 1931 XYZ system) |
| Spectral Response | Approximates CIE standard observer functions x̄(λ), ȳ(λ), z̄(λ) with f₁′ ≤ 8% |
| Illuminance Range | 0.1–99,990 lx (4 auto-ranging scales) |
| Chromaticity Range | ≥5 lx (≥0.5 fcd) for color measurement |
| Accuracy (Ev) | ±2% of reading ±1 digit |
| Accuracy (xy) | ±0.002 (at 800 lx, Standard Illuminant A) |
| Repeatability (Ev) | 0.5% ±1 digit (2σ) |
| Repeatability (xy) | ±0.0005 (at 800 lx, Standard Illuminant A) |
| Temp. Drift (Ev/xy) | ±3% ±1 digit / ±0.003 |
| Humidity Drift (Ev/xy) | ±3% ±1 digit / ±0.003 |
| Response Time | 0.5 s (continuous mode) |
| Display | 4-digit LCD with auto-backlight |
| Interface | RS-232C (bidirectional) |
| Remote Sensing | Up to 100 m via LAN cable (with T-A20/T-A21 adapters, optional) |
| Multi-sensor Support | 2–30 detectors per host |
| Power | 2 × AA alkaline batteries (72 h typical) or AC adapter (AC-A10/A10N, optional) |
| Dimensions | 69 × 174 × 35 mm |
| Weight | 215 g (without batteries) |
| Operating Environment | −10–40 °C, ≤85% RH (non-condensing) |
| Storage | −20–55 °C, ≤85% RH (non-condensing) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CM-2300d |
| Instrument Type | Portable |
| Optical Geometry | 0/45° integrating sphere |
| Light Source | Dual pulsed xenon lamps |
| Spectral Range | 360 nm to 740 nm |
| Wavelength Interval | 10 nm |
| Spectral Bandwidth | ~10 nm |
| Reflectance Range | 0–175% |
| Display Resolution | 0.01% |
| Measurement Time | ~1.5 s (SCI & SCE simultaneously) |
| Repeatability | ΔE*ab ≤ 0.08 (after white calibration, 30 measurements at 10 s intervals) |
| Inter-instrument Agreement | ΔE*ab ≤ 0.4 (SCI, based on BCRA Series II standards) |
| Measuring/Aperture Area | Ø8 mm / Ø11 mm |
| Weight | 670 g (without batteries) |
| Data Storage Capacity | 1,700 sets (each set includes paired SCI/SCE data) |
| Compliance | CIE No. 15, ISO 7724-1, ASTM E1164, DIN 5033 Teil 7, JIS Z8722 Condition C |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CM-2300d |
| Optical Geometry | d/8° (diffuse illumination / 8° viewing angle) with simultaneous SCI & SCE measurement |
| Wavelength Range | 360–740 nm |
| Interval | 10 nm |
| Spectral Bandwidth | ~10 nm |
| Measurement Area | Ø8 mm / Illumination Area: Ø11 mm |
| Repeatability | ΔE*ab ≤ 0.08 (after white calibration, 30 measurements at 10-sec intervals) |
| Inter-instrument Agreement | ΔE*ab ≤ 0.4 (SCI, BCRA Series II avg.) |
| Data Storage | 1,700 sets (each with paired SCI/SCE spectra) |
| Measurement Time | ~1.5 sec |
| Minimum Interval | 3 sec |
| Power | 4 × AA alkaline batteries or AC adapter |
| Weight | 670 g (without batteries) |
| Compliance | CIE No. 15, ISO 7724-1, ASTM E1164, DIN 5033 Teil 7, JIS Z8722 Condition C |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CM-2500c / CM-2600d |
| Instrument Type | Portable Spectrophotometer |
| Optical Geometry | 45°/0° (CM-2500c) / d/8° integrating sphere with SCI & SCE simultaneous measurement (CM-2600d) |
| Light Source | Pulsed Xenon Lamp(s) |
| Spectral Range | 360–740 nm |
| Spectral Interval | 10 nm |
| FWHM | ~10 nm |
| Repeatability | ΔE*ab ≤ 0.04 (CM-2600d, MAV/SCI, 30 measurements) or ≤ 0.05 (CM-2500c) |
| Reflectance Range | 0–175% |
| Resolution | 0.01% |
| Measurement Time | ~1.5 s (standard), ~2 s (fluorescence mode) |
| Aperture Options | MAV Φ8 mm / Φ11 mm |
| Data Storage | 1700 measurements (CM-2600d), 700 (CM-2500c) |
| Compliance | CIE No. 15, ISO 7724-1, ASTM E1164, DIN 5033-7, JIS Z8722 Condition C |
| UV Adjustment | Digital real-time UV correction (CM-2600d only) |
| Interface | RS-232C |
| Power | 4 × AA batteries or AC adapter |
| Dimensions (W×H×D) | 69 × 96 × 193 mm |
| Weight | ~670 g (battery-free) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CM-2500c |
| Illumination/Reception Geometry | 45°/0° (45° annular illumination, vertical reception) |
| Detector | Dual 40-element silicon photodiode arrays |
| Spectral Separation | Planar diffraction grating |
| Wavelength Range | 360–740 nm |
| Wavelength Interval | 10 nm |
| FWHM | Approx. 10 nm |
| Reflectance Range | 0–175% (resolution: 0.01%) |
| Light Source | Dual pulsed xenon lamps |
| Measurement Time | ~1.5 s |
| Minimum Measurement Interval | 4 s |
| Battery Life | ~1,000 measurements (alkaline AA, 23°C) |
| Aperture Size | 7 mm measuring / 11 mm illumination |
| Repeatability | ΔE*ab ≤ 0.05 (white calibration tile, 30 measurements at 10 s intervals) |
| Instrument Error | ΔE*ab ≤ 0.25 (BCRA Series II tiles, avg., vs. master unit, 23°C) |
| Measurement Modes | Single / auto-averaged (3/5/8 shots) |
| Interface | RS-232C |
| Observer Angles | 2° and 10° (CIE 1931 & CIE 1964) |
| Illuminants | A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12 (dual-illuminant simultaneous evaluation) |
| Color Spaces | L*a*b*, L*C*h, CMC(1:1), CMC(2:1), CIE94, CIEDE2000, Yxy, XYZ, Hunter Lab, MI, WI(ASTM E313), WI(CIE), YI(ASTM E313/D1925), ISO Brightness (ISO 2470), Tint (CIE), A/T concentration |
| Data Storage | 700 measurements |
| Pass/Fail Criteria | Box or ellipse tolerance definition |
| Power Supply | 4 × AA batteries or AC adapter |
| Dimensions (W×H×D) | 69 × 96 × 193 mm |
| Weight | ~670 g (battery-free) |
| Operating Environment | 5–40°C, ≤80% RH (non-condensing at ≤35°C) |
| Storage Conditions | 0–45°C, ≤80% RH (non-condensing at ≤35°C) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CM-3500d |
| Illumination/Reception Geometry (Reflectance) | d/8° (diffuse illumination, 8° viewing), switchable between SCI (Specular Component Included) and SCE (Specular Component Excluded) |
| Illumination/Reception Geometry (Transmittance) | d/0° (diffuse illumination, 0° viewing) |
| Detector | Dual 18-element silicon photodiode array with wedge-type continuous interference filter |
| Wavelength Range | 400–700 nm |
| Wavelength Interval | 20 nm |
| FWHM | ~20 nm (average) |
| Photometric Range | 0–175% (resolution: 0.01%) |
| Light Source | Pulsed xenon arc lamp |
| Measurement Time | Approx. 2.5 s (to data output) |
| Minimum Measurement Interval | 3 s |
| Aperture Options (Reflectance) | LAV (Φ30 mm measurement / Φ36 mm illumination), MAV (Φ8 mm measurement / Φ11 mm illumination) |
| Aperture (Transmittance) | ~Φ22 mm |
| Repeatability (after white calibration, 30 measurements at 10 s intervals on white calibration tile) | Spectral reflectance SD ≤ 0.20%, ΔE*ad ≤ 0.05 |
| Instrumental Error | ΔE*ad ≤ 0.15 (typical, BCRA Series II 12-tile average) |
| Temperature Drift | Spectral reflectance ±0.30%/°C, ΔE*ab ≤ 0.05/°C |
| Interface | RS-232C |
| Power Supply | AC 100/120/230 V, 50/60 Hz (with included AC adapter) |
| Operating Environment | 0–40°C, ≤85% RH (non-condensing at ≤35°C) |
| Storage Environment | −20–45°C, ≤85% RH (non-condensing at ≤35°C) |
| Dimensions (W × H × D) | 413 × 190 × 240 mm |
| Weight | 10 kg |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CM-3600d |
| Illumination/Viewing Geometry | d/8° (diffuse illumination, 8° viewing) for reflectance |
| Compliance | CIE No. 15, ISO 7724-1, ASTM E1164, DIN 5033-7, JIS Z8722 Condition C |
| Spectral Range | 360–740 nm |
| Spectral Interval | 10 nm |
| FWHM | ~10 nm |
| Detector | Dual 40-element silicon photodiode arrays |
| Grating Type | Planar diffraction grating |
| Integrating Sphere Diameter | Φ152 mm |
| Light Source | Four pulsed xenon lamps |
| Measurement Time | ~1.5 s (to data output) |
| Apertures | LAV (Φ25.4 mm), MAV (Φ8 mm), SAV (Φ4 mm) for reflectance |
| Repeatability | ΔE*ab ≤ 0.02 (standard deviation), spectral reflectance SD ≤ 0.1% |
| Instrumental Error | ΔE*ab ≤ 0.15 (LAV/SCI, BCRA Series II tile average) |
| UV Control | Digital UV adjustment with selectable 400 nm / 420 nm cutoff filters |
| Transmittance Chamber | 133 mm (W) × ~50 mm (D), ~17 mm aperture |
| Interface | RS-232C (D-sub 9-pin female) |
| Power | AC 120 V / 230 V, 50 / 60 Hz (via included adapter) |
| Operating Environment | 13–33 °C, ≤80% RH (non-condensing) |
| Dimensions (W×H×D) | 244 × 208 × 378 mm |
| Weight | 12 kg |
| Standard Accessories | White calibration tile, aperture masks (Φ4/Φ8/Φ25.4 mm), zero calibration box, AC adapter, dust cover, accessory case, driver software, RS-232C cable (2 m) |
| Brand | Konica Minolta |
|---|---|
| Origin | Shanghai, China |
| Model | CM-3700d |
| Illumination/Viewing Geometry | d/8° (diffuse illumination / 8° viewing), switchable SCI (Specular Component Included) / SCE (Specular Component Excluded) |
| Transmission Mode | d/0° |
| Wavelength Range | 360–740 nm |
| Wavelength Interval | 10 nm |
| FWHM | ~14 nm (average) |
| Photometric Range | 0–200% |
| Resolution | 0.001% |
| Light Source | Pulsed Xenon Arc Lamp |
| Measurement Time | 0.6–0.8 s (to data output) |
| Apertures (Reflectance) | LAV (Φ28 mm illum. / Φ25.4 mm meas.), MAV (Φ11 mm / Φ8 mm), SAV (5×7 mm / 3×5 mm) |
| Aperture (Transmittance) | ~Φ20 mm |
| Repeatability (White Tile, 30 measurements @ 10 s intervals) | ΔE*ad ≤ 0.005 |
| Repeatability (Black Tile, BCRA Series II, 1% R) | ΔE*ad ≤ 0.05 |
| Instrument Error (LAV, BCRA Series II 12-tile avg.) | ΔE*ad ≤ 0.08 (typical) |
| Temperature Drift | Spectral Reflectance ±0.10%/°C |
| UV Adjustment | PC-controlled, continuously variable (1000-step) |
| Sample Thickness Limit (Transmittance) | < 50 mm |
| Interface | RS-232C (1200–9600 bps) |
| Power Supply | AC 100/120/230 V, 50/60 Hz |
| Operating Environment | 13–33 °C, ≤80% RH (non-condensing) |
| Storage Environment | 0–40 °C, ≤80% RH (non-condensing) |
| Dimensions (W×H×D) | 271×259×500 mm |
| Weight | 18 kg |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CM-512m3 |
| Illumination/Reception Geometry | 25°/0°, 45°/0°, 75°/0° (annular multi-angle) |
| Wavelength Range | 400–700 nm |
| Wavelength Interval | 20 nm |
| Reflectance Range | 25°: 0–300% |
| 45°/75° | 0–200% |
| Photometric Resolution | 0.01% |
| Light Source | Pulsed Xenon Arc Lamp |
| Minimum Measurement Interval | 7 s (with white calibration tile, at 23°C) |
| Repeatability | ΔE*ab ≤ 0.05 (n=30, 10-s interval), Spectral Reflectance SD ≤ 0.30% |
| Aperture | Measuring Ø12 mm / Illuminating Ø20 mm |
| Temperature Sensing | Thermopile detector (8–13 µm), Range: −10 to +80°C (±2.5°C accuracy at 0–50°C, ε=0.93) |
| Display | 128×56-dot reflective LCD (21 chars × 7 lines) |
| Interface | RS-232C (D-sub 9-pin female) |
| Power | 4× AA alkaline batteries or optional AC adapter |
| Dimensions (W×H×D) | 115×257×164 mm |
| Weight | ≈1.4 kg (battery-free) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CM-700d / CM-600d |
| Illumination/Viewing Geometry | di:8°, de:8° (Diffuse illumination / 8° viewing) |
| Integrating Sphere Diameter | Φ40 mm |
| Detector | Silicon photodiode array (dual 36-element) |
| Spectral Dispersion | Diffraction grating |
| Wavelength Range | 400–700 nm |
| Measurement Interval | 10 nm |
| FWHM | ~10 nm |
| Light Source | Pulsed xenon lamp with UV-cut filter |
| Measurement Time | Approx. 1 sec |
| Repeatability | Spectral reflectance SD ≤0.1% |
| Instrument Error | ΔE*ab ≤0.2 (MAV/SCI, vs. BCRA Series II 12-color tile set, 23°C) |
| Apertures | MAV: Φ8 mm / Φ11 mm |
| SAV | Φ3 mm / Φ6 mm (CM-700d only) |
| Display | 2.36″ TFT color LCD |
| Interfaces | USB 1.1, Bluetooth® v1.2 (Class 1) |
| Color Spaces | CIE L*a*b*, L*C*h, Hunter Lab, Yxy, XYZ, Munsell (ΔE only for non-Munsell spaces) |
| Color Difference Formulas | ΔE*ab (CIE 1976), ΔE94 (CIE 1994), ΔE00 (CIE 2000), CMC (l:c) |
| Whiteness/Yellowness Indices | WI (ASTM E313), YI (ASTM E313-73 / D1925), ISO Brightness |
| Gloss | 8° gloss measurement capability |
| Data Storage | 4,000 sample measurements + 1,000 standard references |
| Power | 4 × AA alkaline or Ni-MH rechargeable (2300 mAh) |
| Dimensions | 73 (W) × 211.5 (H) × 107 (D) mm |
| Weight | ~550 g (excl. white calibration tile and batteries) |
| Operating Environment | 5–35°C, RH ≤80% (non-condensing) |
| Storage Environment | 0–45°C, RH ≤80% (non-condensing) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CR-10 |
| Illumination/Viewing Geometry | 8°/diffuse (d) |
| Measurement Area | Ø8 mm |
| Display Modes | Δ(L*a*b*) & ΔE*ab, or Δ(L*C*H*) & ΔE*ab |
| Standard Observer | CIE 10° |
| Illuminant | CIE D65 |
| Repeatability | σ(ΔE*ab) ≤ 0.1 (measured on white calibration tile, average of multiple readings) |
| Measurement Interval | Approx. 1 second |
| Power Supply | 4 × AA batteries or optional AC adapter AC-A12 |
| Operating Temperature | 0–40 °C, RH < 85% (non-condensing) |
| Standard Accessories | Soft case CR-A68, protective cap CR-A72, wrist strap CR-A73, 4 × AA batteries |
| Optional Accessories | AC adapter AC-A12, printer cable CR-A75 |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CR-20 |
| Instrument Type | Portable |
| Optical Geometry | d/8° integrating sphere (with specular component included) |
| Light Source | Pulsed Xenon Lamp |
| Spectral Range | 360–750 nm |
| Repeatability | ΔE*ab ≤ 0.1 (std. dev., 30 measurements on white calibration tile at 10-sec intervals) |
| Aperture | Ø8 mm |
| Observer Angle | 10° |
| Illuminant | D65 |
| Display | Backlit LCD |
| Data Storage | Up to 1,000 measurement sets (including reference) |
| Interface | USB 2.0 |
| Power | 4 × AA alkaline/Ni-MH batteries, USB bus power, or AC adapter (optional) |
| Dimensions | 66 × 159.5 × 85 mm |
| Weight | 420 g (without batteries) |
| Compliance | DIN 5033-7, JIS Z 8722 Condition d, ISO 7724-1, CIE No. 15, ASTM E1164 |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CR-400 |
| Type | Portable Color Difference Meter |
| Optical Geometry | d/8° integrating sphere (diffuse illumination, 8° viewing, including specular component) |
| Light Source | Pulsed Xenon Lamp |
| Spectral Range | Y Reflectance: 0.01–160.00% |
| Repeatability | ΔE*ab ≤ 0.07 (std. dev., white calibration, 30 measurements at 10 s intervals) |
| Instrument Error | ΔE*ab ≤ 0.6 (BCRA Series II 12-color average, CIE D65 illuminant, 2° standard observer) |
| Measurement Aperture | 8 mm (measuring), 11 mm (illumination) |
| Measurement Time | 1 s |
| Minimum Interval | 3 s |
| Data Storage | 1,000 measurements (instrument only) |
| Display | Backlit dot-matrix LCD (15×9 characters + icon line) |
| Power | 4× AAA alkaline/Ni-MH batteries or AC adapter AC-A17 (120 V / 230 V, 50–60 Hz) |
| Dimensions | 102 × 244 × 63 mm |
| Weight | ~550 g (incl. batteries, excl. cable) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CR-400 / CR-410 |
| Illumination/Reception Geometry | d/0 (diffuse illumination, 0° viewing |
| Detector | Silicon photodiodes (6 channels) |
| Light Source | Pulsed xenon lamp |
| Measurement Time | 1 s |
| Minimum Interval Between Measurements | 3 s |
| Repeatability | ΔE*ab ≤ 0.07 (after white calibration, 30 measurements at 10 s intervals) |
| Instrument Error | ΔE*ab ≤ 0.6 (CR-400), ≤ 0.8 (CR-410) against BCRA Series II 12-color standard set |
| Aperture Size | CR-400: Ø8 mm measurement / Ø11 mm illumination |
| CR-410 | Ø50 mm measurement / Ø53 mm illumination |
| Reflectance Range | Y = 0.01–160.00% |
| Observer Angle | CIE 2° standard observer |
| Illuminants | C, D65 |
| Color Spaces | XYZ, Yxy, L*a*b*, L*C*h, CIELAB, CIELUV, CMC(l:c), CIE 1994, CIE 2000, Hunter Lab, CIE Whiteness (WI) & Yellowness (YI) per ASTM E313/D1925 (C or D65 dependent) |
| Display | Backlit dot-matrix LCD (15 × 9 characters + icon line) |
| Memory Capacity | 1000 measurements (head-only mode) |
| Interface | RS-232C (default baud rate: 9600 bps) |
| Power | 4 × AAA alkaline or Ni-MH batteries |
| Dimensions (W × H × D) | 102 × 244 × 63 mm |
| Weight | ~550 g (CR-400), ~570 g (CR-410), incl. batteries |
| Operating Environment | 0–40 °C, ≤85% RH (non-condensing) |
| Storage Environment | −20–40 °C, ≤85% RH (non-condensing) |
| Brand | KONICA MINOLTA |
|---|---|
| Origin | Japan |
| Model | CS-100A |
| Type | SLR-type tristimulus spot chroma meter for luminance and colorimetric measurement |
| Luminance Range (Fast) | 0.01–299,000 cd/m² |
| Luminance Range (Slow) | 0.01–49,900 cd/m² |
| Luminance Accuracy | ±2% ±1 digit |
| Chromaticity Accuracy (x,y) | ±0.004 (CIE Illuminant A, 18–28°C) |
| Repeatability (Y) | ±0.2% ±1 digit |
| Repeatability (x,y) | ±0.001 (Y ≥100 cd/m², fast mode) |
| Measurement Angle | 1° |
| Field of View | 9° with 1° measurement area indicator |
| Optical System | 85 mm f/2.8 lens |
| Focus Distance | 1014 mm to infinity |
| Sensor | Three filtered silicon photodiodes approximating CIE 1931 color matching functions |
| Response Time | Fast (sampling: 0.1 s, display: 0.8–1.0 s) |
| Slow (sampling | 0.4 s, display: 1.4–1.6 s) |
| Display | External 3-digit LCD (Y, x, y) |
| Interface | RS-232C (4800 bps) |
| Power | 9 V alkaline battery (included) |
| Dimensions (W×H×D) | 79 × 208 × 154 mm |
| Weight | 890 g (without battery) |
| Operating Environment | 0–40°C, ≤85% RH (non-condensing at ≤35°C) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CS-150 / CS-160 |
| Measurement Angle | 1° (CS-150) / 1/3° (CS-160) |
| Luminance Range | 0.01–999,900 cd/m² (CS-150) / 0.1–9,999,000 cd/m² (CS-160) |
| Colorimetric Compliance | CIE 1931 x(λ), y(λ), z(λ) tristimulus matching |
| Spectral Response Match | Optimized to CIE V(λ) photopic luminosity function |
| Minimum Measured Area | 14.4 mm Ø (CS-150, 1014 mm distance) / 4.5 mm Ø (CS-160, 213 mm) |
| Repeatability | ±0.2% +1 digit (luminance), ±0.001 (chromaticity at ≥10 cd/m²) |
| Accuracy | ±2% ±1 digit (luminance), ±0.004 (chromaticity at ≥5 cd/m²) |
| Data Storage | 1,000 measurements |
| Interface | USB 2.0 (data transfer & bus power) |
| Software | CS-S20w (included) |
| Calibration Channels | 10 user-definable reference sources |
| Operating Temp. | 0–40°C, ≤85% RH (non-condensing) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CS-200 |
| Measurement Angles | 1° / 0.2° / 0.1° |
| Luminance Range | 0.01–20,000,000 cd/m² (at 0.1°) |
| Accuracy (Lv) | ±2% ±1 digit (≥10 cd/m², Source A, slow mode) |
| Accuracy (xy) | ±0.003 (≥10 cd/m², Source A, slow mode) |
| Repeatability (Lv) | ±0.1% ±1 digit (2–200,000 cd/m², 2σ, auto mode) |
| Spectral Method | Grating + Linear Photodiode Array |
| Minimum Measured Area | 0.1 mm (with close-up lens) |
| Interface | USB 1.1 |
| Power | AC adapter or 4× AA batteries (~3 h continuous operation) |
| Dimensions | 95 × 127 × 330 mm |
| Weight | 1.8 kg (body only) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CS-200 |
| Type | Color and Luminance Meter |
| Measurement Principle | Spectral Matching Method (CIE 1931 2° Standard Observer) |
| Photodetector | Silicon Photodiode |
| Luminance Range | 0.01 to 20,000,000 cd/m² (at 0.1° field of view) |
| Illuminance Range | 0.1 to 99,990 lx |
| Chromaticity Accuracy (xy) | ±0.002 (Std. Illuminant A, 800 lx) |
| Repeatability (xy) | ±0.0005 |
| Temporal Stability | 2 measurements/sec (continuous) |
| Output Interface | RS-232C |
| Display | 4-digit LCD with backlight |
| Power | Four AA alkaline batteries (≈72 h continuous use) or AC adapter AC-A10 |
| Dimensions | 69 × 174 × 35 mm |
| Weight | 215 g (body only) |
| Compliance | JIS C 1609-1993 Class AA (fs ≤ 8%), ISO/CIE traceable calibration protocols |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | CS-2000 |
| Measurement Wavelength Range | 380–780 nm |
| Wavelength Resolution | 0.9 nm/pixel |
| Spectral Bandwidth (FWHM) | <5 nm |
| Measurement Angles | 1°, 0.2°, 0.1° (interchangeable) |
| Luminance Range (A-source) | 0.003–5,000 cd/m² (1°), 0.075–125,000 cd/m² (0.2°), 0.3–500,000 cd/m² (0.1°) |
| Minimum Displayed Luminance | 0.00002 cd/m² |
| Luminance Accuracy | ±2% (A-source, ≥0.003 cd/m²) |
| Chromaticity Accuracy | x,y ±0.003 to ±0.001 (dependent on luminance level) |
| Repeatability (2σ) | Luminance 0.15–0.4%, Chromaticity 0.0004–0.002 |
| Polarization Error | ≤2% (1°), ≤3% (0.1°/0.2°) |
| Integration Time | Fast mode 0.005–16 s, Normal mode 0.005–120 s |
| Measurement Time | 2–243 s |
| Operating Environment | 5–35°C, RH ≤80% (non-condensing) |
| Dimensions | 158 × 200 × 300 mm (main unit), Ø70 × 95 mm (lens) |
| Weight | 6.2 kg |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | LS-100 |
| Measurement Angle | 1° |
| Optical System | Single-lens reflex (SLR) with f=85 mm, F2.8 lens |
| Field of View (Finder) | 9° with diopter adjustment |
| Minimum Measurement Distance | ∞ to 1,014 mm (∞ to φ14.4 mm) |
| Close-up Lens Options | No.153 (φ18.7–φ8.0 mm), No.135 (φ8.7–φ5.2 mm), No.122 (φ4.3–φ3.2 mm), No.110 (φ1.5–φ1.3 mm) |
| Detector | Silicon photodiode |
| Spectral Match Error (fₛ) vs. CIE V(λ) | ≤8% |
| Response Modes | FAST (display time: 0.8–1.0 s |
| sampling | 0.1 s), SLOW (display time: 1.4–1.6 s |
| sampling | 0.4 s) |
| Luminance Range | FAST: 0.001–299,900 cd/m² |
| SLOW | 0.001–49,990 cd/m² |
| Accuracy | ±2% ±1 digit (>1,000 cd/m²) |
| Repeatability | ±0.2% ±1 digit (>1,000 cd/m²) |
| Temperature Drift | ±3% ±1 digit (0–40°C, referenced to 20°C) |
| Calibration Modes | Konica Minolta standard or user-defined reference (1 channel) |
| Luminance Ratio Mode | 1 channel, manual or measured reference input |
| Measurement Functions | Instant luminance, peak luminance, luminance ratio (%) |
| Display | 4-digit LCD (external & eyepiece-mounted with LED backlight) |
| Digital Interface | RS-232C compliant (4,800 bps), bidirectional control and data logging |
| Power Supply | One 9 V alkaline battery (JIS 6F22) or optional DP-10 data printer |
| Power Consumption | ~16 mA (finder illumination ON, measurement active) |
| Operating Environment | 0–40°C, ≤85% RH (non-condensing), Pollution Degree 2, Overvoltage Category II |
| Storage | −20–55°C, ≤85% RH (non-condensing) |
| Dimensions & Weight | 79 × 208 × 150 mm (W × H × D), 850 g (body only, no battery) |
| Standard Accessories | Lens cap, eyepiece ND filter, eyepiece cap, 9 V battery, carrying case |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | LS-150 / LS-160 |
| Luminance Range | 0.001–999,900 cd/m² (LS-150), 0.01–9,999,000 cd/m² (LS-160) |
| Measurement Angle | 1° (LS-150), 1/3° (LS-160) |
| Optical System | Single-Lens Reflex (SLR), f=85 mm, F2.8 |
| Spectral Response | Matched to CIE V(λ) Photopic Luminosity Function |
| Accuracy | ±2% ±2 digits (≤1 cd/m²), ±2% ±1 digit (>1 cd/m²) |
| Repeatability | 0.2% + 1 digit |
| Minimum Measured Area | Ø14.4 mm (1°), Ø4.5 mm (1/3°) |
| with optional macro lens | Ø1.3 mm (LS-150), Ø0.4 mm (LS-160) |
| Shortest Measurement Distance | 1012 mm (1°), 213 mm (1/3°) |
| Stray Light Suppression | ≤1% |
| Polarization Error | ≤1% |
| Temperature Drift | ≤0.2%/K |
| Compliance | DIN 5032-7 Class B (LS-150) |
| Interface | USB 2.0 |
| Power | USB bus power or AC adapter (optional) |
| Dimensions | 71 × 214 × 154 mm |
| Weight | 850 g (body only) |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | Multi Gloss 268 |
| Measurement Angles | 20°, 60°, 85° |
| Measurement Spot Size | 20°: 10 × 10 mm |
| 60° | 9 × 15 mm |
| 85° | 5 × 38 mm |
| Standards Compliance | ISO 2813, ISO 7668, ASTM D523, ASTM D2457, DIN 67530, JIS Z 8741, BS 3900, BS 6161-12 |
| Measurement Range | 20°: 0.0–2000 GU |
| 60° | 0.0–1000 GU |
| 85° | 0.0–160 GU |
| Resolution | 0.1 GU (0.0–99.9 GU), 1 GU (100–2000 GU) |
| Repeatability | ±0.2 GU (0.0–99.9 GU), ±0.2% of reading (100–2000 GU) |
| Instrument Error | ±0.5 GU (0.0–99.9 GU), ±0.5% of reading (100–2000 GU) |
| Data Storage | 999 measurements with timestamp |
| Standard Memory | 50 reference standards |
| Measurement Speed | 0.5 s per reading |
| Power Source | 1.5 V AA (R6) alkaline batteries (approx. 10,000 measurements) |
| Auto Power-Off | Adjustable 10–99 s |
| Operating Environment | 15–40 °C, ≤85% RH (non-condensing at ≤35 °C) |
| Storage Conditions | −10–60 °C, ≤85% RH (non-condensing at ≤35 °C) |
| Dimensions | 155 × 73 × 48 mm |
| Weight | 400 g |
| Interface | RS-232C |
| Supported Languages | English, Spanish, German, French, Italian, Japanese |
| Software | GlossLink data exchange software |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | ProMetric I/Y Series |
| Sensor Resolution Options | 2 MP (I2/Y2), 8 MP (I8), 16 MP (I16/Y16), 29 MP (I29/Y29) |
| Sensor Type | Scientific-grade CCD, 12-bit, interline transfer |
| Cooling | Thermoelectric (TE) stabilization |
| Dynamic Range | High-dynamic-range (HDR) multi-exposure mode |
| Optical Interface | Interchangeable C-mount lenses with variable focal length and f-number |
| Spectral Matching | CIE 1931 photopic luminosity function (V(λ)) and CIE 1931/1964 color matching functions |
| Filter System | Precision CIE-matched color filters + calibrated neutral density (ND) filters |
| Software | ProMetric Control & Analysis Suite (v5.x), TrueTest™, TrueMura™, PM-KB Keyboard Analysis Module |
| Compliance | Designed for ISO/CIE-compliant photometric and colorimetric measurements |
| Connectivity | Gigabit Ethernet, USB 3.0 |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | T-10 / T-10M |
| Illuminance Range | 0.01–299,900 lx |
| Spectral Match | f₁′ ≤ 8% (vs. CIE V(λ)) |
| Cosine Error | ±1% at 10°, ±7% at 60°, ±25% at 80° |
| Accuracy | ±2% ±1 digit (Konica Minolta standard) |
| Output | RS-232C digital, 1 mV/digit analog |
| Display | Backlit LCD, 3–4 digits |
| Operating Temp | –10 to 40°C, RH ≤ 85% (non-condensing) |
| Power | 2× AA batteries (≥72 h continuous) |
| Dimensions (main unit) | 69 × 161.5 × 30 mm |
| Weight | 200 g (T-10), 205 g (T-10M) |
| Compliance | Designed for ISO/CIE-compliant photometric measurement in industrial, architectural, and environmental applications |
| Brand | Konica Minolta |
|---|---|
| Origin | Japan |
| Model | T-10 |
| Compliance | JIS C 1609-1993 Class AA (T-10/T-10M), JIS Precision Grade (T-10P) |
| Photodetector | Silicon photodiode |
| Spectral mismatch (fₛ) | ≤4% (T-10P), ≤8% (T-10/T-10M) |
| Cosine error | ±1% at 10°, ±2% at 30°, ±6% at 50°, ±7% at 60°, ±25% at 80° |
| Illuminance range | 0.01–299,900 lx |
| Integrated illuminance range | 0.01–999,900×10³ lx·h |
| Measurement functions | NORM (illuminance), Δ (difference/ratio), Σ (integrated illuminance, duration, hourly average) |
| Accuracy | ±2% ±1 digit |
| Temp./humidity drift | ±3% ±1 digit |
| Output | RS-232C digital, analog (1 mV/digit, max 3 V, 10 kΩ output impedance) |
| Display | 3-digit (some 4-digit) LCD with auto-backlight |
| Power | Two AA alkaline batteries (≥72 h continuous use) or optional AC adapter |
| Dimensions | 69 × 174 × 35 mm |
| Weight (excl. batteries) | 200 g |
| Operating environment | −10–40 °C, ≤85% RH (non-condensing), Pollution Degree 2, Equipment Class II |