NCS (China Iron and Steel Research Institute Group)
Filter
Showing all 11 results
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Model | NC-GET Series |
| Origin | Beijing, China |
| Manufacturer | NCS |
| Country of Origin | China |
| Pricing | Available Upon Request |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | OEM Manufacturer |
| Country of Origin | China |
| Model | NC-GUFT Series |
| Pricing | Available Upon Request |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Origin | Beijing, China |
| Manufacturer | Yes |
| Country of Origin | China |
| Model | NC-PUT Series |
| Pricing | Upon Request |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Country of Origin | China |
| Model | NC-RGET Series |
| Pricing | Available Upon Request |
| Inspection Mode | Automated Inspection |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Origin | Beijing, China |
| Manufacturer | Yes |
| Country of Origin | China |
| Model | NC-C-SC-AN Series for Bar Inspection |
| Pricing | Upon Request |
| Probe Configuration | 128-element array, max 32 elements fired per group |
| Static Sensitivity | Φ0.8 mm flat-bottom hole at depths of ½, ¼, and 1/8 bar diameter (≥5 mm min depth), SNR > 12 dB |
| Dynamic Sensitivity | Φ0.8 mm FBH at ≤0.5 m/s |
| End Blind Zone | <30 mm |
| Near-Surface Blind Zone | None |
| False Negative Rate | 0% |
| Inspection Speed | ≥60 m/min |
| Scan Modes | Linear Scan, Sector Scan, Dynamic Depth Focusing (DDF) |
| Applicable Standards | GB/T 4162, ISO 18563-1:2015, ISO 18563-2:2015, ASTM E2700 |
| Material Coverage | Carbon steel, alloy steel, bearing steel, spring steel, cold-heading steel |
| Bar Diameter Range | 16–65 mm (customizable) |
| Bar Length Range | 6–12 m |
| Surface Roughness Requirement | ≤25 µm Ra |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Origin | Beijing, China |
| Manufacturer | Yes |
| Country of Origin | China |
| Model | NC-C-SC-AN Series for Steel Plate Inspection |
| Ultrasound Channels | 16 (expandable to 64 via cascading) |
| Probe Configuration | Up to 4 probes, 128 total elements |
| Phased Array Beamforming Capacity | 1024 delay laws |
| Data Throughput | 40 MB/s via Gigabit Ethernet |
| Network Interface | Industrial Power + Dual-Mode Fiber Optic |
| Encoder Input | 2-channel quadrature (A/B phase) |
| Trigger Mode | Internal or External (software-selectable) |
| Power Supply | AC 220 V ±10%, 50/60 Hz |
| TCG (Time-Corrected Gain) | Integrated |
| Minimum Detectable Flaw | Ø3 mm flat-bottom hole (FBH) at full material thickness |
| Scanning Coverage | Adjustable multi-probe array with 15 mm overlap per 120 mm probe |
| Mechanical Architecture | Immersion-type water bath with dual staggered probe carriage assemblies |
| Compliance Standards | ISO 3183, GB/T 2970, ASTM A435, ASTM A578, ISO 10893-11 |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Model | NC-C-SCAN Series |
| Origin | Beijing, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Regional Classification | Domestic (PRC) |
| Pricing | Available Upon Request |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Origin | Beijing, China |
| Manufacturer | Yes |
| Country of Origin | China |
| Model | NX-300FA |
| Pricing | Upon Request |
| Detection Principle | Energy-Dispersive X-ray Fluorescence (EDXRF) |
| Sample Throughput | Up to 84 samples unattended |
| Analysis Time | 3 min (Cd screening), 10 min (Cd quantification), 6 min (multi-element screening), 20 min (multi-element quantification) |
| Detection Limits | Cd ≤ 0.038 mg/kg, Se ≤ 0.048 mg/kg |
| Linear Range | Cd 0.038–2.6 mg/kg, Se 0.048–2.0 mg/kg |
| Precision (RSD) | Cd ≤ 5%, Se ≤ 10% |
| Reproducibility | Absolute difference between two independent measurements ≤ 20% of mean |
| Radiation Safety | Compliant with GBZ 115–2002 (Chinese national standard for radiation protection of XRF instruments) |
| Sample Form | Direct analysis of powdered or granular cereal grains and processed food matrices |
| Elements Supported | Cd, Pb, As, Cr, Hg, Se |
| Regulatory Alignment | Designed for compliance with GB 2762–2022 (Chinese National Standard for Contaminants in Foods), aligned with ISO 13253:2012 (EDXRF for metals in food), and supports audit-ready data handling per GLP/GMP principles |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Model | OPA-500 |
| Optical System | Czerny-Turner spectrometer with 500 mm focal length |
| Grating | 2700 lines/mm |
| Linear dispersion | 0.7407 nm/mm |
| Spectral resolution | ≤0.01 nm |
| Wavelength range | 160–650 nm |
| Excitation source | High-stability continuous spark source, no pre-burn required |
| Data acquisition rate | 100 kHz |
| Scanning platform | Siemens CNC-controlled 3-axis system (X: 360 mm, Y: 247 mm, Z: 350 mm) |
| Positioning accuracy (bidirectional) | X-axis ±0.016/±0.010 mm, Y/Z-axis ±0.012/±0.010 mm |
| Repeatability (bidirectional) | ±0.008/±0.006 mm (X/Y/Z) |
| Maximum scan area | 300 × 200 mm |
| Sample weight capacity (uniform load) | 200 kg |
| Compressed air requirement | 0.6 MPa |
| Power supply | 220 VAC ±10%, 50 Hz |
| Dimensions (W×D×H) | 1929 × 1428 × 2016 mm |
| Scan patterns | Rectangular, circular, cross (“rice”), sector, and custom-defined contours |
| Data interface | Ethernet |
| Software compliance | Supports audit trail, user access control, and electronic signature per FDA 21 CFR Part 11 requirements |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Origin | China |
| Manufacturer Type | Authorized Distributor |
| Model | Pure Iron Flux |
| Pricing | Available Upon Request |
| Brand | NCS (China Iron and Steel Research Institute Group) |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Benchtop |
| Excitation Method | Spark Discharge |
| Detector Type | Charge-Coupled Device (CCD) |
| Focal Length | 500 mm |
| Grating Groove Density | 2700 lines/mm |
| Wavelength Range | 130–800 nm (enabling detection of N, Li, Na, K, and other elements) |
| Spectral Coverage | Full-spectrum acquisition |
| Pixel Array | 3648 + 46 pixels |
| Pixel Size | 8 µm |
| Optical Architecture | Paschen-Runge mounting with holographic concave grating |
| Temperature Stability | ±0.1 °C via integrated thermostatic shielding |
| Argon Consumption | 60 mL/min in standby mode |
| MTBF | >3000 hours |
| Dimensions (W×D×H) | 470 × 872 × 435 mm |
| Weight | 100 kg |
| Power Requirement | 220 V ±10%, single-phase, 10 A, 2.5 kVA |