SARAD
Filter
Showing all 2 results
| Brand | SARAD |
|---|---|
| Origin | Germany |
| Model | RTM 1688-2 |
| Sampling Method | Active sampling |
| Measurement Range | 0–10 MBq/m³ |
| Detection Limit | Not specified (contextually implied as sub-10 Bq/m³ in slow mode) |
| Response Time | 95% in 12 min (fast mode) |
| Flow Rate | 0.30 L/min |
| Detector Type | Integrated semiconductor alpha spectrometer with electrostatic collection chamber |
| Sensitivity | 3 cpm/(Bq/m³) (fast mode) / 7 cpm/(Bq/m³) (slow mode) |
| Chamber Volume | ~250 mL |
| Humidity Independence | Yes, no desiccant required |
| Data Storage | 2047 measurement records |
| Compliance | Designed for ISO 11665 series, ICRP recommendations, and EPA/WHO radon assessment protocols |
| Brand | SARAD |
|---|---|
| Origin | Germany |
| Model | RTM2200 |
| Sampling Method | Active |
| Measurement Range | 0–10,000,000 Bq/m³ |
| Minimum Detectable Concentration | 1 Bq/m³ |
| Response Time | 12 min (Po-214 excluded) / 120 min (Po-214 included) |
| Detector Type | Silicon semiconductor alpha spectrometer with deep ion implantation |
| Sensitivity | 3.2 cpm/(kBq/m³) (Po-214 excluded) / 7 cpm/(kBq/m³) (Po-214 included) |
| Chamber Volume | ~135 mL |
| Built-in Pump Flow Rate | 300–800 mL/min |
| Operating Interface | 240 × 128 pixel graphical touchscreen |
| Data Storage | SD card with GPS tagging (optional) |
| Communication Interfaces | USB, RS232 (supports GSM modem, ZigBee, or Wi-Fi adapter upgrade) |
| Dimensions | 235 × 140 × 255 mm |
| Environmental Sensors (standard) | Air temperature, relative humidity, flow rate, barometric pressure, battery status |