UIC (USA)
Filter
Showing all 7 results
| Brand | UIC (USA) |
|---|---|
| Origin | USA |
| Model | CM120 |
| Detection Principle | Potentiometric CO₂ Titration |
| Temperature Range | Up to 1100 °C |
| Carbon Quantification Range | 1–10,000 µg absolute carbon |
| Linearity | Full-scale 0–100% CO₂ |
| Resolution | 0.01 µg carbon |
| Sample Types | Solids, liquids, slurries, heterogeneous waste, carbon-rich matrices |
| Compliance | Designed for ASTM D7582, ISO 29541, USP <281>, and GLP-compliant workflows |
| Data Storage | Integrated SD card logging |
| Analysis Time | 5–7 min per sample |
| Interference rejection | Post-combustion scrubbers for SOₓ, NOₓ, H₂O, halides |
| Brand | UIC (USA) |
|---|---|
| Origin | USA |
| Model | CM120 |
| Detection Principle | High-Temperature Catalytic Combustion + Potentiometric CO₂ Detection |
| Measuring Range | 1–10,000 µg C per sample |
| Detection Limit | 0.01 µg C |
| Furnace Max Temperature | 1100 °C |
| Sample Types | Solids (soils, sediments, coal, sulfur, sludge) and Suspended-Particle Liquids |
| Compliance | ASTM D7573, ISO 20236, USP <281>, EPA Method 415.1 compatible |
| Data Storage | Internal memory + 3.5″ floppy disk drive |
| Calibration | Factory-calibrated |
| Brand | UIC (USA) |
|---|---|
| Origin | USA |
| Model | CM120 |
| Detection Principle | High-Temperature Catalytic Combustion + Potentiometric CO₂ Detection |
| Temperature Range | Up to 1100 °C |
| Carbon Range | 1 µg – 10,000 µg (0.1 ppm – 100 wt%) |
| Resolution | 0.01 µg absolute carbon |
| Linearity | Full dynamic range without user calibration |
| Sample Forms | Solids, slurries, particulate-laden liquids, powders, soils, sediments, coals, sulfurs |
| Compliance | ASTM D7573, ISO 20538, USP <281>, EPA Method 9060A (adapted), GLP-ready audit trail support |
| Brand | UIC (USA) |
|---|---|
| Origin | USA |
| Model | CM140 |
| Compliance | ASTM D513 |
| Detection Principle | Potentiometric CO₂ detection after acidification |
| Dynamic Range | 1 µg C to 10,000 µg C (absolute) |
| Resolution | 0.01 µg C |
| Sample Forms | Solid and liquid |
| Acid Digestion Volume Options | 10, 25, 50, or 100 mL |
| Analysis Time | 5–7 min typical |
| Calibration | Factory-calibrated |
| Data Storage | Internal memory + 3.5" floppy disk |
| Software Interface | Standalone instrument control with manual parameter entry |
| Brand | UIC (USA) |
|---|---|
| Origin | USA |
| Model | CM340 |
| Detection Principle | Coulometric Titration (Faraday-based SO₂/H₂S Quantification) |
| Dynamic Range | 1 µg to 10,000 µg total sulfur |
| Resolution | 0.01 µg S |
| Sample Forms | Solid and liquid |
| Acid Digestion Volume Options | 10, 25, 50, or 100 mL |
| Analysis Time | ~5–7 min per sample (temperature- and matrix-dependent) |
| Data Storage | Internal memory for up to 50 samples + 3.5" floppy disk export |
| Compliance | Designed for GLP/GMP environments |
| Brand | UIC (USA) |
|---|---|
| Origin | USA |
| Model | CM350 |
| Instrument Type | Laboratory Benchtop Analyzer |
| Measurement Principle | Coulometric Titration (Faraday-based SO₂/H₂S Detection) |
| Dynamic Range | 1 µg – 10,000 µg S per sample |
| Detection Limit | ~1 µg S (absolute) |
| Linear Range | 0.2% RSD or better for certified standards |
| Sample Forms | Solids (0.5 mg – 10 g) and Liquids |
| Total Analysis Time | 5–7 min (combustion mode) |
| Combustion Temperature | Up to 1100 °C (dual-zone furnace) |
| Acid Digestion | Modified Monier-Williams method with programmable heating & gas purging |
| Data Storage | Internal memory + 3.5″ floppy disk (CM5014S unit) |
| Compliance Support | ASTM D3120, D3246, D5453 |
| Brand | UIC (USA) |
|---|---|
| Origin | USA |
| Model | CM350 |
| Category | Other Elemental Analyzers |
| Compliance | ASTM D3120, D3246, D5453 |
| Detection Principle | Coulometric titration based on Faraday’s law |
| Dynamic Range | 1 µg – 10,000 µg sulfur (equivalent to ~1 ppm – 100% w/w, sample-dependent) |
| Resolution | 0.01 µg S |
| Analysis Time | 5–7 min (combustion mode), ≤15 min (acid digestion mode) |
| Sample Forms | Solids (0.5 mg – 10 g) and liquids |
| Furnace Max Temp | 1100 °C |
| Data Storage | Internal floppy drive (CM5014S), exportable via RS232 |