BINDER UF V Series Ultra-Low Temperature Freezer

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand BINDER
Origin Germany
Model UF V
Temperature Range -40 °C to -86 °C
Insulation Vacuum Insulation Panels (VIP) + Polyurethane Foam
Cooling System Cascade Compressor
Refrigerants R-404A and R-508B (non-ozone-depleting)
Door Configuration Right- or Left-Hinged with Mechanical Limiter
Access Control RFID-Based (Model-Dependent)
Data Logging USB-Enabled Recorder (Model-Dependent)
Cooling Option Water-Cooled Condenser (Model-Dependent)
Door Lock Motorized Mechanical Lock
Interior Stainless Steel Chamber
Interior Shelves 3 Stainless Steel Pull-Out Trays
Compartment Doors 4 ABS Plastic Doors
Alarm System Battery-Backed Audible/Visual Alarm with Dry Contact Output, Runtime Hour Meter
Interfaces RS-422 Serial Port, 4–20 mA Analog Output, Zero-Voltage Alarm Relay
Internal Volume 477 L (UF V500) / 700 L (UF V700)
External Dimensions (W×H×D mm) 826×1966×938 (UF V500) / 1100×1966×938 (UF V700)
Internal Dimensions (W×H×D mm) 606×1300×605 (UF V500) / 890×1300×605 (UF V700)
Net Weight 247 kg (UF V500) / 288 kg (UF V700)

Overview

The BINDER UF V Series Ultra-Low Temperature Freezer is an engineered solution for long-term, high-integrity storage of temperature-sensitive biological materials—including primary cell lines, stem cells, plasma, mRNA constructs, cryopreserved tissues, and reference standards—within the critical range of –40 °C to –86 °C. Designed and manufactured in Germany, the UF V leverages a dual-stage cascade refrigeration system paired with environmentally compliant refrigerants (R-404A and R-508B), ensuring stable thermal performance without ozone-depleting substances. Its structural integrity derives from a hybrid insulation architecture: vacuum insulation panels (VIPs) integrated with high-density polyurethane foam minimize thermal bridging and reduce energy consumption while maintaining uniform internal temperature distribution. The stainless steel interior chamber resists corrosion and facilitates decontamination, meeting stringent laboratory hygiene requirements for GLP-compliant biobanking and clinical sample repositories.

Key Features

  • Cascade compressor cooling system delivering precise, drift-free temperature control across the full operational range (–40 °C to –86 °C)
  • Vacuum insulation panel (VIP) technology combined with rigid polyurethane foam for superior thermal resistance and reduced wall thickness
  • Stainless steel interior chamber and three removable stainless steel pull-out trays for mechanical durability and ease of cleaning
  • Four independently hinged ABS plastic compartment doors minimizing cold air loss during partial access
  • Motorized mechanical door lock with configurable access logic; optional RFID-based user authentication (model-dependent)
  • Battery-backed alarm system with audible/visual alerts, dry-contact relay output, and runtime hour meter for preventive maintenance scheduling
  • Dual communication interfaces: RS-422 serial port for integration into centralized monitoring systems and 4–20 mA analog output for continuous temperature telemetry
  • Optional water-cooled condenser configuration for stable operation in high-ambient-temperature environments (e.g., non-air-conditioned server rooms or shared lab corridors)

Sample Compatibility & Compliance

The UF V Series supports storage of vials (2 mL, 5 mL), cryoboxes (10×10), and standard ISO/ANSI-compliant cryotubes without dimensional interference. Its uniform temperature profile (±1.5 °C max deviation at setpoint, per IEC 61000-4-2 validation protocols) ensures viability retention for ISO 20387:2018-compliant biobanks. The unit complies with EN 61000-6-3 (EMC emissions), EN 61000-6-2 (immunity), and meets essential safety requirements under EN 61010-1. Optional data logging modules support 21 CFR Part 11–compliant audit trails when paired with validated software platforms, enabling traceability for FDA-regulated clinical trials and GMP-aligned manufacturing support labs.

Software & Data Management

While the UF V operates as a standalone instrument, its RS-422 interface enables bidirectional communication with third-party environmental monitoring systems (e.g., SensiNet, TempTale Web, or custom SCADA). USB-enabled data loggers (available as factory-installed options) record temperature at user-defined intervals (1 min to 24 h), storing ≥30 days of high-resolution data onboard. Exported .csv files retain timestamps, sensor IDs, and alarm event markers—facilitating retrospective analysis and CAPA documentation. No proprietary cloud platform is required; all data remains under institutional control, supporting HIPAA- and GDPR-aligned data governance policies.

Applications

  • Long-term archival storage of human biospecimens in academic biorepositories and hospital tissue banks
  • Stabilization of thermolabile reagents (e.g., CRISPR-Cas9 ribonucleoprotein complexes, lentiviral vectors, and lyophilized antibodies)
  • Supporting ISO/IEC 17025-accredited calibration labs requiring certified cold chain continuity for reference material storage
  • Backup freezing for -80 °C ultra-low freezers undergoing scheduled maintenance or validation downtime
  • Pre-analytical sample stabilization prior to RNA-seq, single-cell proteomics, or metabolomic profiling workflows

FAQ

What is the recommended ambient operating temperature for the UF V series?
The unit is rated for continuous operation in ambient environments between +10 °C and +32 °C. With the optional water-cooled condenser, operational stability extends up to +40 °C.
Does the UF V meet UL or CSA safety certification requirements?
Yes—models supplied for North America are certified to UL 61010-1 and CAN/CSA C22.2 No. 61010-1, including field-wiring compliance for hardwired installations.
Can the internal temperature uniformity be verified and documented?
Yes. BINDER provides IQ/OQ documentation packages, and the unit supports placement of up to six calibrated PT100 probes (per EN 12830) for mapping studies. Validation templates align with ISO/IEC 17025 and AABB accreditation requirements.
Is remote monitoring possible without installing proprietary software?
Yes—via Modbus RTU over RS-422, the freezer exposes real-time temperature, alarm status, compressor duty cycle, and door-open duration to existing building management systems (BMS) or LabVantage/LIMS integrations.
What is the expected service life and maintenance interval?
Under typical laboratory use (≤4 door openings/day), the cascade system is designed for ≥15 years of service. Annual preventive maintenance—focused on condenser cleaning, refrigerant pressure verification, and door gasket integrity—is recommended and supported by BINDER’s global service network.