Empowering Scientific Discovery

GreenPrima PM8202I Fluoride Ion Analyzer

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand GreenPrima
Origin United Kingdom
Manufacturer Type Manufacturer
Origin Category Imported
Model PM8202I
Price Upon Request
Measurement Range 0.00–20,000 ppm F⁻
Resolution 0.0 / 0.1 / 1 ppm (user-selectable)
Accuracy ±0.01 / ±0.1 / ±1 ppm (dependent on range)
Temperature Compensation –10 to 130 °C (manual or automatic
Operating Temperature –10 to 70.0 °C
Storage Temperature –20 to 70.0 °C
Display Backlit large-dot-matrix LCD
Language Support English & Chinese
Data Storage 600,000 records
Power Supply 90–260 VAC, 50/60 Hz or optional 24 VDC
Analog Outputs Two isolated 4–20 mA outputs (max. loop resistance 500 Ω, accuracy 0.1% FS, configurable for fluoride concentration or temperature)
Communication RS485 Modbus RTU
Cleaning Control Output Adjustable cleaning interval (0.1–1000 h) and duration (1–1000 s)
Alarm Outputs Two independent Hi/Lo relay alarms (5 A / 250 VAC / 30 VDC, with hysteresis setting)
Enclosure Rating IP65
Mounting Options Wall-mounted, pipe-mounted, or panel-mounted
Dimensions (W×H×D) 144 × 144 × 108 mm
Panel Cutout 138 × 138 mm
Weight 0.87 kg

Overview

The GreenPrima PM8202I Fluoride Ion Analyzer is a dedicated online electrochemical measurement system engineered for continuous, real-time quantification of fluoride ion (F⁻) concentration in aqueous media. It operates on the principle of potentiometric detection using a solid-state, ion-selective electrode (ISE)—specifically the Bsens710 fluoride-selective electrode—paired with a stable reference electrode. The instrument measures the potential difference across the ISE membrane, which follows the Nernst equation and exhibits logarithmic response to fluoride activity. This method delivers high selectivity against common interferents such as chloride, nitrate, and sulfate—critical for reliable monitoring in complex matrices like municipal drinking water, wastewater effluent, and industrial process streams. Designed for unattended operation in demanding environmental and utility settings, the PM8202I integrates robust signal conditioning, built-in temperature compensation (via NTC30K or PT1000 inputs), and adaptive calibration logic to maintain long-term measurement integrity.

Key Features

  • Triple-parameter capability: Configurable for fluoride (F⁻), chloride (Cl⁻), or ammonium (NH₄⁺) analysis via interchangeable Bsens-series ISEs (Bsens710, Bsens720, Bsens730)
  • IP65-rated sealed enclosure with corrosion-resistant housing—suitable for indoor and semi-outdoor installations in humid or splash-prone environments
  • Backlit high-contrast dot-matrix LCD display showing real-time F⁻ concentration, measured temperature, system time, relay status, and diagnostic indicators
  • Dual isolated 4–20 mA analog outputs with full-scale programmability—supporting simultaneous transmission of fluoride concentration and temperature signals to DCS or SCADA systems
  • RS485 Modbus RTU interface for bidirectional communication, enabling remote configuration, data polling, and firmware updates via standard industrial protocols
  • Configurable cleaning cycle control with adjustable interval (0.1–1000 hours) and duration (1–1000 seconds), compatible with solenoid-valve–driven chemical or ultrapure water rinse systems
  • Two independently programmable relay alarms with user-defined setpoints, hysteresis, and fail-safe behavior—certified for 5 A switching capacity at 250 VAC / 30 VDC
  • Password-protected configuration mode prevents unauthorized parameter changes, supporting operational accountability and audit readiness

Sample Compatibility & Compliance

The PM8202I is validated for use with potable water, secondary-treated wastewater, cooling tower make-up water, and low-TDS industrial process streams. It meets EN 61326-1 (EMC for industrial environments) and conforms to IEC 60529 for IP65 ingress protection. While not certified under ISO 10523 or ASTM D4192 (which govern laboratory-based fluoride titration and colorimetry), its ISE-based methodology aligns with EPA Method 300.0 for fluoride determination in drinking water when deployed within defined matrix limitations (e.g., pH 5.0–8.0, total alkalinity < 200 mg/L as CaCO₃, absence of high Al³⁺ or Fe³⁺). For regulated utilities, the device supports GLP-aligned operational practices through timestamped calibration logs, alarm event records, and full data traceability—facilitating internal QA reviews and regulatory inspections.

Software & Data Management

All measurement, calibration, alarm, and maintenance events are timestamped and stored locally in non-volatile memory (600,000 entries). Data retention includes fluoride concentration (ppm), temperature (°C), electrode mV output, slope (%), offset (mV), calibration date/time, and operator ID (if enabled via external authentication). Export is supported via Modbus register readout for integration into LIMS or historian platforms. The embedded firmware implements write-protected calibration history with version-stamped entries, satisfying basic requirements for data integrity under FDA 21 CFR Part 11 when used in conjunction with procedural controls (e.g., dual-operator calibration verification). No cloud connectivity or proprietary software is required—configuration and diagnostics are fully accessible via front-panel navigation or Modbus commands.

Applications

  • Municipal drinking water treatment: Continuous monitoring of fluoridation dosing systems to ensure compliance with WHO and national guidelines (e.g., 0.5–1.5 ppm F⁻)
  • Wastewater reclamation plants: Detection of fluoride breakthrough in tertiary filtration or ion exchange units prior to discharge or reuse
  • Pharmaceutical manufacturing: In-process verification of fluoride residuals in purified water (PW) and water-for-injection (WFI) distribution loops
  • Electroplating and semiconductor fabs: Monitoring fluoride-containing etchants and rinse waters to prevent equipment corrosion and effluent violations
  • Power plant boiler feedwater systems: Early warning of fluoride-induced stress corrosion cracking risk in stainless steel components

FAQ

What electrode types are compatible with the PM8202I?
The analyzer supports GreenPrima’s Bsens710 (fluoride), Bsens720 (chloride), and Bsens730 (ammonium) ion-selective electrodes—each featuring rugged epoxy-body construction and integrated temperature sensing.
Is automatic temperature compensation mandatory?
No—temperature compensation can be configured as manual (fixed value) or automatic using either NTC30K or PT1000 sensors; auto-compensation is recommended for applications with variable thermal conditions.
How often must the fluoride electrode be calibrated?
Calibration frequency depends on application severity; typical practice is daily for critical potable water dosing, weekly for wastewater monitoring, and per manufacturer guidance (every 7–14 days) for stable industrial streams.
Can the PM8202I operate on DC power?
Yes—24 VDC auxiliary power input is available as an option, enabling deployment in locations without AC infrastructure or where electrical noise mitigation is required.
Does the unit support MODBUS TCP?
No—the communication interface is RS485 Modbus RTU only; TCP/IP gateway integration requires an external protocol converter.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0