IKA T10 Battery Complete Solution ULTRA-TURRAX® Handheld Disperser
| Brand | IKA |
|---|---|
| Origin | Germany |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Origin Category | Imported |
| Model | T10 Battery Complete Solution |
| Pricing | Upon Request |
| Motor Input Power | 30 W |
| Motor Output Power | 25 W |
| Volume Range (H₂O) | 0.5–100 mL |
| Max. Viscosity | 200 mPa·s |
| Speed Range | 8,000–30,000 rpm |
| Speed Deviation | ±5 % |
| Speed Control | Continuous (Infinitely Variable) |
| Speed Display | Backlit LCD |
| Timer Function | Yes (Forward/Reverse Countdown) |
| Timer Display | LCD |
| Noise Level (No Dispersion Tool) | 65 dB(A) |
| Motor Principle | Brushless DC Motor |
| Process Mode | Batch |
| Duty Cycle | 10 min ON / 5 min OFF |
| Dimensions (W×H×D) | 53 × 239 × 70 mm |
| Weight | 0.5 kg |
| Ambient Temperature Range | 5–40 °C |
| Relative Humidity Limit | ≤80 % RH (non-condensing) |
| USB Interface | USB-C |
| Max. Input Power (Charging) | 45 W |
| Standby Power Consumption | 0.1 W |
| DC Supply Voltage | 15 V = |
| Charging Current | 3,000 mA |
| Battery Type | Replaceable Rechargeable Lithium-ion |
| Full-Charge Runtime | ~60 min (typical dispersion load) |
| Charge Time | 1.5–2 h (0–100 %) |
| Compatible Accessories | S 10 N-5 G dispersing tool, H 44 boss head clamp, R 104 stand |
Overview
The IKA T10 Battery Complete Solution is a cordless, handheld ULTRA-TURRAX® disperser engineered for precision batch homogenization, tissue homogenization, and emulsification in life science and analytical laboratories. Operating on the principle of high-shear rotor-stator mechanics, it delivers consistent mechanical energy transfer to samples through controlled rotational motion—enabling efficient cell lysis, particle size reduction, and uniform suspension formation. Unlike AC-powered benchtop dispersers, the T10 Battery integrates a replaceable lithium-ion battery and brushless DC motor architecture to eliminate electrical noise, carbon dust contamination, and tethered operation constraints—making it uniquely suited for field-deployable applications such as mobile diagnostic units, biosafety level-2 (BSL-2) carts, and decentralized point-of-care sample preparation workflows.
Key Features
- Brushless DC motor ensures silent operation (≤65 dB(A) without dispersing tool), zero carbon particulate emission, and extended service life with minimal maintenance.
- Integrated backlit LCD display shows real-time speed (8,000–30,000 rpm) and programmable timer (forward/reverse countdown), enabling strict protocol adherence and inter-laboratory reproducibility.
- Rechargeable lithium-ion battery supports up to 60 minutes of continuous operation at typical load conditions; full recharge achieved in 1.5–2 hours via USB-C or optional R200 C charging cradle.
- Modular design guarantees full compatibility with IKA’s T10 Basic accessory ecosystem—including ULTRA-TURRAX® dispersing tools (e.g., S 10 N-5 G), H44 clamping system, and R104 vertical stand—reducing capital expenditure and inventory complexity.
- Rugged, ergonomic housing (53 × 239 × 70 mm; 0.5 kg) meets IP20 ingress protection standards and complies with IEC 61010-1 for laboratory electrical safety under normal ambient conditions (5–40 °C, ≤80 % RH non-condensing).
Sample Compatibility & Compliance
The T10 Battery accommodates aqueous and low-viscosity organic media within a defined volume range of 0.5–100 mL (H₂O-equivalent), supporting common life science matrices including cell suspensions, plant homogenates, bacterial lysates, and polymer emulsions up to 200 mPa·s. Its closed-system dispersing tools minimize aerosol generation during BSL-2 procedures. While not certified for sterile manufacturing environments, the device aligns with GLP documentation practices through its traceable timer function and stable speed control (±5 % deviation). It does not require FDA 21 CFR Part 11 compliance out-of-the-box but supports audit-ready manual log entries when integrated into validated SOPs per ISO/IEC 17025 or ASTM D4290 protocols.
Software & Data Management
This model operates as a standalone instrument without embedded firmware-based data logging or PC connectivity. All operational parameters—speed, time, direction—are manually set and visually confirmed via the LCD interface. No proprietary software, cloud synchronization, or electronic record retention is provided. Users are responsible for external documentation of dispersion conditions in lab notebooks or LIMS-compatible templates. The USB-C port serves exclusively for charging; no data transmission capability is implemented. For laboratories requiring automated data capture, integration with third-party environmental monitoring systems or digital lab notebooks must be achieved via manual entry or external timing instrumentation.
Applications
- Tissue and organ homogenization for RNA/DNA extraction, protein isolation, and metabolite profiling.
- Preparation of nanoparticle suspensions, liposome dispersions, and microemulsions in formulation development.
- Routine QC dispersion of pigment pastes, pharmaceutical suspensions, and cosmetic bases.
- On-site sample conditioning in environmental field labs, food safety vans, and veterinary mobile units.
- Teaching labs requiring portable, low-risk instrumentation for demonstrating shear-dependent colloidal stabilization principles.
FAQ
Is the T10 Battery suitable for use in explosion-proof environments?
No. It is not rated for hazardous locations (ATEX, IECEx) and must only be operated in general laboratory or controlled field settings.
Can the battery be replaced by the end user?
Yes. The lithium-ion battery module is field-replaceable using standard Torx tools; replacement kits are available directly from IKA authorized distributors.
Does the device support continuous operation beyond 10 minutes?
No. Thermal management limits duty cycle to 10 minutes ON followed by 5 minutes OFF to prevent motor overheating and ensure long-term reliability.
Are ULTRA-TURRAX® dispersing tools autoclavable?
Stainless steel tools (e.g., S 10 N-5 G) are autoclavable at 121 °C for 20 minutes; polymer components (e.g., PTFE shaft seals) require validation per manufacturer guidelines.
What is the expected service life of the brushless motor under routine use?
Rated for ≥10,000 operating hours under nominal load conditions, assuming adherence to specified duty cycles and ambient operating limits.