Empowering Scientific Discovery

Jiada Xiaotianshe GDYQ-901SC2 Food Nitrite Rapid Analyzer

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Jiada Xiaotianshe
Origin Jilin, China
Manufacturer Type Direct Manufacturer
Regional Category Domestic (China)
Model GDYQ-901SC2
Function Type Single-Analyte Detection
Sample Versatility Multi-Sample Compatible
Optical Wavelength 538 nm
Detection Channels 1
Detection Limit 2.0 mg/kg
Measurement Range 0.0–200.0 mg/kg
Accuracy ±2%
Regulatory Method Modified GB/T 5009.33–2003

Overview

The Jiada Xiaotianshe GDYQ-901SC2 Food Nitrite Rapid Analyzer is a dedicated photometric instrument engineered for the quantitative determination of nitrite (NO₂⁻) in food matrices using the Griess reaction principle. It operates at a fixed optical wavelength of 538 nm—optimized for maximum absorbance sensitivity of the stable azo dye formed when nitrite reacts with sulfanilamide and N-(1-naphthyl)ethylenediamine dihydrochloride under acidic conditions. Designed specifically for routine food safety screening in QC laboratories, regulatory inspection units, and on-site monitoring environments, the GDYQ-901SC2 delivers reliable, standardized results within 10 minutes per sample—without requiring HPLC or spectrophotometer infrastructure. Its architecture aligns with the methodological framework of GB/T 5009.33–2003 (National Standard of the People’s Republic of China), incorporating procedural refinements to enhance repeatability and reduce inter-operator variability.

Key Features

  • Fixed-wavelength photometry at 538 nm ensures consistent spectral alignment with the nitrite-diazotization chromophore, minimizing interference from matrix-absorbing compounds.
  • Integrated microcontroller unit enables automated timing control for reaction incubation, signal acquisition, and concentration calculation—eliminating manual stopwatch dependency.
  • Backlit LCD display with Chinese-language interface supports intuitive navigation through measurement, calibration, data recall, system setup, and built-in help menus—designed for operators with varying technical backgrounds.
  • Dedicated sample preparation module—including centrifuge-compatible extraction tubes, filtration devices, and pre-measured reagent kits—standardizes pretreatment across diverse food types (e.g., cured meats, pickled vegetables, dried fruits, beverages).
  • Single-channel optical path with calibrated LED light source and silicon photodiode detector provides high signal-to-noise ratio and long-term photometric stability.
  • Built-in memory stores up to 200 test records with timestamp, sample ID, and result value—exportable via RS-232 serial interface for traceability in GLP-compliant workflows.

Sample Compatibility & Compliance

The GDYQ-901SC2 is validated for use with solid, semi-solid, and liquid food samples including but not limited to: cured pork and poultry products, canned meat preparations, fresh and processed fish, fermented and salted vegetables (e.g., kimchi, sauerkraut), table salt, carbonated and non-carbonated beverages, and roasted or sun-dried nuts and seeds. Sample homogenization, aqueous extraction, and clarification steps follow harmonized protocols derived from GB/T 5009.33–2003, ensuring method equivalence for regulatory reporting. While not certified to ISO/IEC 17025 or accredited under CNAS, the instrument’s performance characteristics—including detection limit (2.0 mg/kg), linear dynamic range (0.0–200.0 mg/kg), and accuracy (±2% relative error against reference standards)—meet internal quality control thresholds defined in enterprise food safety management systems aligned with China’s GB 2760–2024 and GB 2726–2016 requirements.

Software & Data Management

The embedded firmware supports real-time calibration curve generation using at least three standard concentrations (0, 10, 50, and 100 mg/kg NaNO₂), with automatic polynomial regression (linear or quadratic) selection based on R² thresholding (>0.995). All measurement events—including calibration history, blank subtraction logs, and outlier flags—are time-stamped and stored locally. Data export via RS-232 allows integration into LIMS platforms or Excel-based audit trails. Though lacking native 21 CFR Part 11 compliance features (e.g., electronic signatures, role-based access), the device supports manual documentation practices required under basic GMP and HACCP verification schemes.

Applications

  • Routine batch release testing of processed meats prior to distribution.
  • On-factory surveillance of nitrite levels during curing process validation.
  • Third-party inspection labs conducting statutory food safety audits per SAMR directives.
  • University teaching laboratories demonstrating colorimetric quantification principles in food chemistry curricula.
  • Import/export quarantine units performing rapid triage of high-risk consignments prior to confirmatory LC-MS/MS analysis.

FAQ

What sample types require dilution prior to analysis?
High-salt or pigment-rich matrices (e.g., soy sauce, red wine vinegar, beetroot puree) may necessitate 1:2 or 1:5 dilution with deionized water to remain within the instrument’s linear range and avoid turbidity-induced scattering errors.
Is the reagent kit compatible with other nitrite analyzers?
The proprietary single-use reagent vials are optimized for GDYQ-901SC2’s reaction kinetics and volume delivery; cross-platform use is not validated and may compromise accuracy.
How often must the optical path be cleaned?
The cuvette holder and LED window should be wiped with lint-free tissue and 70% ethanol after every 20 measurements—or immediately following any visible residue—to maintain photometric fidelity.
Can the instrument store calibration curves across power cycles?
Yes—calibration parameters are retained in non-volatile memory and persist through shutdown, eliminating daily recalibration unless reagent lot changes occur.
Does the device meet international export documentation requirements?
While the GDYQ-901SC2 conforms to Chinese national standards and supports traceable measurement records, CE marking, FDA listing, or IEC 61010-1 certification are not applicable; users intending export deployment must verify local metrological acceptance criteria independently.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0