Brand METTLER TOLEDO
Origin Switzerland
Model Easy D30, Easy D40
Instrument Type Benchtop Digital Density Meter
Sample Type Solid and Liquid
Density Range 0–3 g/cm³
Temperature Range 15–25 °C (Easy D30), 15–50 °C (Easy D40)
Density Resolution 0.001 g/cm³ (Easy D30), 0.0001 g/cm³ (Easy D40)
Density Accuracy ±0.001 g/cm³ (Easy D30), ±0.0005 g/cm³ (Easy D40)
Temperature Accuracy ±0.2 °C (Easy D30), ±0.1 °C (Easy D40)
Minimum Sample Volume 1.2 mL
Measurement Units Density, Specific Gravity, Alcohol % v/v, Brix, API Gravity, Acidity, Urea Concentration, Custom Concentration
Built-in Methods 16
Bubble Detection BubbleCheck™ (Easy D40 only)
Viscosity Correction Automatic (both models)
Cleaning Check Yes (both models)
Dimensions (W × H × D) 170 × 156 × 273 mm

Overview

The METTLER TOLEDO EasyPlus Series digital density meters — Easy D30 and Easy D40 — are precision benchtop instruments engineered for high-reproducibility density and specific gravity measurements of liquids and solids across research, quality control, and production environments. Based on the oscillating U-tube principle (ASTM D4052, ISO 12185, USP ), these instruments determine density by measuring the resonant frequency of a precisely manufactured borosilicate glass U-tube filled with sample. The frequency shift correlates directly to mass per unit volume, enabling traceable, SI-based density values with minimal operator intervention. Designed in Switzerland and built to industrial-grade tolerances, the EasyPlus platform integrates thermal stabilization, vibration-dampening mechanical architecture, and real-time optical monitoring of the measurement cell to ensure robustness under variable lab or near-line conditions.

Key Features

  • High-stability oscillating U-tube sensor with temperature-controlled Peltier element for rapid thermal equilibration (15–25 °C on Easy D30; extended 15–50 °C range on Easy D40)
  • Dual-resolution capability: 0.001 g/cm³ (Easy D30) and 0.0001 g/cm³ (Easy D40), supporting both routine QC and high-precision R&D applications
  • Automatic viscosity correction algorithm compliant with ISO 12185 Annex A, compensating for damping effects in non-Newtonian or high-viscosity samples without manual input
  • BubbleCheck™ real-time optical detection (Easy D40 only) identifies entrapped air in the U-tube prior to measurement, preventing erroneous low-density readings
  • Integrated cleaning check function validates post-measurement rinse efficacy using reference water density verification
  • Intuitive LongClick™ interface enables one-touch start of measurement sequences; contextual on-screen guidance reduces training time and minimizes procedural errors
  • Color-coded pass/fail display (green/red LED indicators) linked to user-defined specification limits — supports immediate go/no-go decisions per ASTM E29 or ISO/IEC 17025 compliance workflows
  • Compact footprint (170 × 156 × 273 mm) optimized for fume hoods, QC benches, or production-floor deployment

Sample Compatibility & Compliance

The Easy D30 and Easy D40 accommodate a broad range of homogeneous liquid and solid samples — including solvents, fuels, pharmaceutical syrups, food emulsions, battery electrolytes, and polymer solutions — provided minimum volume (≥1.2 mL) and viscosity constraints are met. Both models comply with international standards for density metrology: ASTM D4052 (petroleum products), ISO 12185 (crude oil and liquid petroleum products), USP (pharmaceutical density testing), and EN ISO 279 (detergents). The measurement cell design meets GLP/GMP data integrity requirements when paired with optional audit-trail-enabled software. All internal calibration routines are traceable to NIST-certified reference materials, and temperature control accuracy is validated per ISO/IEC 17025 clause 6.4.10.

Software & Data Management

EasyDirect™ PC software provides secure, audit-ready data handling in accordance with FDA 21 CFR Part 11 requirements (when configured with electronic signatures and role-based access control). It enables automated report generation — including time-series density plots, statistical summaries (mean, SD, RSD), and trend analysis against specification limits. Raw measurement data (including frequency, temperature, viscosity-corrected density, and BubbleCheck™ status flags) are stored in timestamped, non-editable .csv or .xlsx formats. Optional integration with LIMS or MES systems is supported via OPC UA or TCP/IP protocols. The instrument’s embedded memory retains up to 1,000 measurements with full metadata (operator ID, method name, timestamp, environmental conditions).

Applications

  • Pharmaceutical: Excipient density verification, syrup concentration control (Brix, % w/w), alcohol content in tinctures (USP )
  • Food & Beverage: Sugar content in soft drinks and juices (Brix), alcohol % in wines and spirits (OIV methods), density grading of dairy products
  • Petrochemical: API gravity and distillation fraction density (ASTM D1298), fuel blend verification (gasoline, biodiesel)
  • Chemical Manufacturing: Catalyst slurry density, acid concentration (H₂SO₄, HCl), urea solution monitoring in SCR systems
  • Academic Research: Polymer solution characterization, ionic liquid density–temperature profiling, nanofluid stability assessment

FAQ

What is the difference between Easy D30 and Easy D40 in terms of temperature control?
The Easy D30 operates within a fixed 15–25 °C range, while the Easy D40 extends this to 15–50 °C, enabling measurements of higher-boiling-point solvents or temperature-dependent studies.
Can the EasyPlus density meters measure suspended solids or heterogeneous mixtures?
No — the oscillating U-tube principle requires homogeneous, particle-free samples. Suspensions must be centrifuged or filtered prior to analysis to avoid clogging or signal noise.
Is external calibration required for routine use?
No. Daily verification using certified water (at 20 °C: 0.998203 g/cm³) and air (0.0012 g/cm³) is sufficient. Full calibration with two-point NIST-traceable standards is recommended annually or after maintenance.
How does automatic viscosity correction work?
The instrument measures damping of the U-tube oscillation amplitude and applies a physics-based correction model derived from ISO 12185 Annex A, eliminating need for manual viscosity input or lookup tables.
Does the system support multi-user operation with individual method libraries?
Yes — EasyDirect™ software allows creation of user-specific method folders, password-protected method editing, and assignment of method permissions per operator role.