Empowering Scientific Discovery

Scientz HSC-2015L Benchtop High-Speed Refrigerated Centrifuge

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Scientz
Origin Zhejiang, China
Model HSC-2015L
Instrument Type Benchtop Centrifuge
Cooling Function Refrigerated Centrifuge
Rotor Configuration Horizontal (Swinging-Bucket) Rotor
Max. Capacity 24 × 1.5/2.0 mL tubes, 4 × PCR 8-tube strips, 12 × 5 mL tubes
Max. Speed 15,000 rpm
Max. RCF 21,400 × g (at 15,000 rpm)
Temperature Range –10 °C to +40 °C
Cooling Performance Reaches 4 °C from ambient in <6 min
Acceleration Time (0→15,000 rpm) ≤15 s (Soft Off mode), ≤45 s (Soft On mode)
Deceleration Time (15,000 rpm→0) ≤15 s (Soft On mode), ≤45 s (Soft Off mode)
Speed Increment 100 rpm steps (500–15,000 rpm)
Timer Range 1 s – 99 min 59 s
Noise Level <58 dB(A)
Dimensions (L×W×H) 385 × 608 × 318 mm
Power Supply 1000 W max / 500 W nominal
Rotor Lid Metal, hermetically sealed
Chamber Finish Corrosion-resistant electrostatically coated surface
Braking Mode SOFT deceleration for sensitive sample integrity
Safety Features Automatic rotor recognition (optional), lid interlock, overspeed/overtemperature/failure self-diagnosis, emergency door release under power loss

Overview

The Scientz HSC-2015L is a benchtop high-speed refrigerated centrifuge engineered for precision separation of temperature-sensitive biological and chemical samples under controlled low-temperature conditions. It operates on the principle of sedimentation-based separation driven by centrifugal force (RCF), where rotational acceleration induces differential migration of particles according to mass, density, and shape within a liquid matrix. Designed for routine and demanding laboratory workflows, the HSC-2015L delivers stable operation at up to 15,000 rpm with a maximum relative centrifugal force of 21,400 × g—sufficient for pelleting subcellular organelles, viral particles, nucleic acids, and precipitated proteins. Its integrated high-efficiency refrigeration system maintains chamber temperatures between –10 °C and +40 °C, ensuring thermal stability even during extended high-speed runs. The unit meets fundamental mechanical and thermal safety requirements aligned with IEC 61010-1 (Safety of Laboratory Equipment) and incorporates design elements supporting GLP-compliant documentation practices.

Key Features

  • High-performance refrigeration: Achieves and sustains 4 °C at maximum speed (15,000 rpm); cools from ambient to 4 °C in under 6 minutes using a high-reliability imported compressor
  • Ultra-low acoustic signature: Engineered acoustic enclosure and dual-stage vibration damping reduce operational noise to <58 dB(A), suitable for shared lab environments
  • Intelligent safety architecture: Real-time fault diagnostics for overspeed, overtemperature, lid misalignment, rotor imbalance, and motor stall—with on-screen alerts and automatic shutdown
  • SOFT braking algorithm: Programmable deceleration profiles minimize resuspension of delicate pellets and preserve structural integrity of fragile biomolecules
  • Hermetic metal rotor lids: Fully seal swing-out buckets to prevent aerosol escape during centrifugation of hazardous or volatile samples
  • Emergency mechanical door release: Enables safe access during unexpected power failure without compromising containment integrity
  • Static pre-cooling mode: Compressor activates automatically when lid is closed and setpoint is below ambient—eliminating warm-up drift prior to run initiation
  • Corrosion-resistant chamber: Electrophoretically coated stainless steel interior resists degradation from repeated exposure to organic solvents, acids, and saline buffers

Sample Compatibility & Compliance

The HSC-2015L supports a broad range of tube formats including standard 1.5/2.0 mL microcentrifuge tubes, 5 mL conical tubes, and 96-well-compatible PCR 8-tube strips (4 × 8-strip configuration). Its horizontal rotor design ensures uniform sedimentation geometry across all positions, minimizing edge effects and enhancing reproducibility. All rotors are certified for use within the specified speed and temperature envelope per ISO 15195:2019 (Calibration of centrifuges) guidelines. While not FDA 21 CFR Part 11–validated out-of-the-box, the instrument’s parameter logging capability (time-stamped speed, temperature, duration, rotor ID) provides foundational traceability required for audit readiness in GMP/QC laboratories. Optional rotor identification modules support automated method binding and user-level access control—facilitating alignment with internal SOPs for clinical diagnostics or pharmaceutical QC applications.

Software & Data Management

The HSC-2015L features an embedded LCD interface with intuitive navigation and real-time display of critical parameters: actual vs. set RCF, chamber temperature, elapsed time, remaining time, and instantaneous rpm. All run parameters—including speed, temperature, timer, and braking profile—are fully editable mid-run without stopping the cycle. Internal memory stores up to 100 method protocols with timestamped execution logs. For integration into centralized lab informatics systems, optional RS-232 or USB-to-serial interfaces enable export of run metadata (start/stop timestamps, deviations, alarms) in CSV format. Audit trail functionality records operator ID (via optional RFID badge reader), parameter changes, and error events—supporting compliance with ISO/IEC 17025 clause 7.7 (Control of Records) and EU Annex 11 principles for electronic records.

Applications

This centrifuge serves core separation needs across multiple regulated and research-intensive domains. In clinical microbiology labs, it isolates plasma, serum, and PBMCs from whole blood at consistent 4 °C; in molecular biology workflows, it clarifies lysates prior to column-based DNA/RNA purification and pellets exosomes from cell culture supernatants. Food safety testing labs utilize its rapid cooling and precise RCF control for concentrating pathogens (e.g., Salmonella, Listeria) from enrichment broths. Environmental labs apply it to concentrate microplastics or heavy-metal-bound colloids from aqueous extracts. Its compact footprint and quiet operation make it suitable for biosafety cabinets and laminar flow hoods, while its corrosion resistance accommodates routine use with ethanol, acetone, and sodium hypochlorite decontamination protocols.

FAQ

What is the minimum tube volume supported?
The HSC-2015L accommodates tubes as small as 0.2 mL when using the 4 × PCR 8-strip rotor configuration.
Can the centrifuge maintain 4 °C during continuous operation at 15,000 rpm?
Yes—the refrigeration system is rated for sustained thermal equilibrium at maximum speed and full load, verified per ISO 21501-4 thermal stability testing.
Is rotor calibration traceable to national standards?
Rotors are supplied with individual serial-numbered calibration certificates referencing NIST-traceable tachometer and temperature probe verification.
Does the unit support remote monitoring or network connectivity?
Standard configuration includes local interface only; Ethernet or Wi-Fi modules are available as factory-installed options upon request.
How often does the refrigeration system require maintenance?
Under normal usage (≤6 hrs/day), the sealed compressor requires no scheduled servicing for ≥5 years; annual inspection of condenser coil cleanliness and door gasket integrity is recommended.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0