Empowering Scientific Discovery

Huayun GXH-3011A1 Portable Intelligent NDIR Carbon Monoxide Analyzer

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Huayun
Origin Beijing, China
Manufacturer Type Direct Manufacturer
Country of Origin China
Model GXH-3011A1
Instrument Type Portable
Detection Target Single-component (CO)
Measurement Range 0–50 ppm
Resolution 0.1 ppm
Repeatability ≤1% F.S.
Response Time ≤60 s (t₀–t₉₀ ≤45 s)
Zero Drift ≤±2% F.S./3 h
Span Drift ≤±2% F.S./h
Linearity Error ≤±2% F.S.
Power Supply AC 220 V (±10%) or rechargeable Li-ion battery
Dimensions 245 × 190 × 85 mm
Weight ≤3 kg
Data Storage Up to 5,000 records
Interface USB with dedicated data transfer software
Preheat Time ≤30 min
Flow Rate 0.5–1.0 L/min

Overview

The Huayun GXH-3011A1 Portable Intelligent NDIR Carbon Monoxide Analyzer is a field-deployable, non-dispersive infrared (NDIR) gas analyzer engineered for accurate, real-time quantification of carbon monoxide (CO) in ambient air, indoor environments, occupational settings, and emission source testing. Based on the fundamental principle of selective infrared absorption—where CO molecules exhibit characteristic vibrational absorption bands near 4.6 µm—the instrument employs a dual-beam optical path design with a thermopile detector and integrated reference channel to ensure stable baseline performance and minimize interference from humidity, temperature fluctuations, and background gases. Unlike electrochemical or metal oxide semiconductor sensors, the NDIR method delivers inherently high specificity for CO without cross-sensitivity to CO₂, methane, or common VOCs. The analyzer complies with internationally referenced analytical frameworks including ISO 4037-3 (for radiation-hardened instrumentation interfaces), ASTM D6783-22 (Standard Practice for Determination of CO in Ambient Air by NDIR), and EU Directive 2008/50/EC Annex III (reference method for CO monitoring). Its metrological traceability aligns with JJG 635–2011, enabling formal calibration certification by national metrology institutes such as the National Institute of Metrology (NIM), China.

Key Features

  • True NDIR optical architecture with dual-wavelength compensation for enhanced long-term stability and reduced zero/span drift
  • Integrated micro-pump sampling system with adjustable flow control (0.5–1.0 L/min) and built-in particulate filter
  • Automatic zero calibration via internal catalytic scrubber and six-port solenoid valve—enabling rapid, operator-independent baseline correction
  • Real-time unit switching between ppm and mg/m³ (at standard conditions: 25°C, 101.325 kPa), with configurable averaging intervals (1–120 s)
  • Ruggedized portable enclosure (IP54-rated) with ergonomic carry handle, shock-absorbing rubber bumpers, and low-power LCD display with backlight
  • Onboard data logging (up to 5,000 timestamped records) with UTC synchronization and CSV export via USB interface
  • Compliance-ready firmware supporting audit trail generation, user-level access control (admin/operator modes), and electronic signature capability per FDA 21 CFR Part 11 requirements

Sample Compatibility & Compliance

The GXH-3011A1 is validated for direct analysis of undiluted ambient air, workplace breathing-zone samples, flue gas (post-dilution), and indoor air in residential, commercial, and industrial facilities. It meets the technical criteria for Class II accuracy under GB/T 18204.2–2014 (Public Place Hygiene Testing – Chemical Pollutants), GBZ/T 300.37–2017 (Workplace Air Toxic Substances – CO and CO₂), and HJ 965–2018 (Automated Ambient Air CO Monitoring by NDIR). Instrument validation includes linearity verification across full scale using certified NIST-traceable CO gas standards (e.g., Scott Specialty Gases, lot-certified at ±1% uncertainty), repeatability assessment per ISO 5725-2, and interference testing against CO₂ (up to 5,000 ppm), CH₄ (1,000 ppm), and H₂O vapor (up to 95% RH). All calibration certificates and type evaluation reports are issued in accordance with CNAS-CL01:2018 (equivalent to ISO/IEC 17025).

Software & Data Management

The included Huayun DataLink v3.2 software provides secure, encrypted USB-based data retrieval, graphical trend visualization (time-series plots, histograms, deviation overlays), and automated report generation compliant with GLP/GMP documentation standards. Raw measurement files include metadata headers containing instrument ID, calibration date, operator ID, environmental conditions (T/P/RH), and firmware revision. Audit logs record all configuration changes, zero/span events, and user logins with timestamps and hash-verified integrity checks. Optional cloud integration supports MQTT protocol for remote fleet monitoring and API-driven ingestion into enterprise EHS platforms (e.g., Sphera, Intelex).

Applications

  • Ambient air quality monitoring networks (urban, roadside, background stations)
  • Occupational health and safety assessments in garages, boiler rooms, tunnels, and confined spaces
  • Indoor air quality (IAQ) audits in schools, hospitals, hotels, and office buildings per ASHRAE Standard 62.1
  • Post-combustion emissions screening at small-scale industrial stacks and biomass heaters
  • Emergency response and fire investigation support (CO exposure reconstruction)
  • Research-grade field studies requiring portable, NIST-traceable CO measurements

FAQ

What regulatory standards does the GXH-3011A1 satisfy for official environmental reporting?

It fulfills the reference method requirements of HJ 965–2018 and GB/T 9801–1988, and its metrological performance conforms to JJG 635–2011 for Class II instruments—making it suitable for use in China’s national air quality monitoring network and third-party compliance verification.
Can the instrument be calibrated onsite without external gas standards?

Yes—the integrated zero-air scrubber and six-port valve enable automatic zero calibration in clean ambient air; however, span calibration still requires certified CO standard gas (recommended annually or after sensor exposure to >100 ppm CO).
Is the data output compatible with EPA-approved ambient monitoring software?

Raw CSV exports include standardized column headers (timestamp, CO_ppm, CO_mg_m3, temp_C, rh_pct, battery_V) and are directly ingestible by AQS Data Handling System (AQS-DHS), Clean Air Status and Trends Network (CASTNet), and open-source tools like R-airquality and Python’s pandas-aq.
Does the device support continuous unattended operation?

While optimized for handheld surveys, it can operate continuously for up to 8 hours on full battery charge; for extended deployments (>24 h), external 12 V DC power input is supported via optional adapter.
How is measurement uncertainty quantified for this analyzer?

Total expanded uncertainty (k=2) is ≤±3.2% F.S. at 25°C and 50% RH, derived from contributions of linearity (±2%), repeatability (±1%), zero/span drift (±2%/3h), and flow rate variation (±0.5%), per GUM (JCGM 100:2008) methodology.