Empowering Scientific Discovery

ChemTron M2000SP-X Humidity Calibration Chamber

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand ChemTron
Origin Germany
Model M2000SP-X
Humidity Range <7–95 %RH (at 23 °C)
Humidity Accuracy ±0.8 %RH (at 23 °C)
Temperature Range 10–55 °C
Temperature Accuracy ±0.15 °C (NVLAP-calibrated)
Humidity Stability 0.2 %RH (at 23 °C)
Temperature Stability 0.1 °C (at 23 °C)
Humidity Uniformity 0.1 %RH @ 50 %RH / 23 °C
Temperature Uniformity 0.1 °C @ 50 %RH / 23 °C
Equilibration Time 3–5 min (at 23 °C)
External Gas Flow Rate 3 L/min (with external recirculation)
Internal Chamber Dimensions (standard) 250 × 200 × 100 mm
Optional Extended Chamber (M2000SP-XR) 300 × 250 × 150 mm
Probe Diameter Options 10, 12, 15, 18 mm
Calibration Certificate NVLAP-accredited
Communication Interface RS485 (optional), software-controllable or user-programmable

Overview

The ChemTron M2000SP-X Humidity Calibration Chamber is a high-precision, compact environmental test system engineered for traceable calibration and verification of hygrometers, dew point meters, humidity transmitters, and relative humidity (RH) sensors in metrology laboratories, quality control facilities, and accredited calibration centers. Unlike traditional humidity generation methods—such as chilled-mirror dew point generators or dual-pressure humidity standards—the M2000SP-X employs solid-state electronic humidity synthesis and closed-loop PID-controlled thermohygrometric regulation to deliver rapid, stable, and repeatable RH and temperature conditions across its operational range. Its design prioritizes metrological integrity: all critical performance specifications—including humidity accuracy (±0.8 %RH at 23 °C), temperature accuracy (±0.15 °C, NVLAP-certified), and chamber uniformity (0.1 %RH / 0.1 °C)—are validated under ISO/IEC 17025-compliant procedures. The chamber achieves full thermal and hygric equilibrium in just 3–5 minutes, enabling high-throughput calibration workflows without compromising uncertainty budgets.

Key Features

  • Solid-state electronic humidity generation eliminates reliance on saturated salt solutions or compressed gas systems—reducing maintenance, drift risk, and operator dependency.
  • NVLAP-accredited calibration certificate included, supporting compliance with ISO/IEC 17025, ASTM E2654, and EURAMET cg-12 guidelines for humidity calibration.
  • High spatial uniformity: ≤0.1 %RH and ≤0.1 °C deviation across the working volume (measured at 50 %RH / 23 °C), verified per ISO 17025 internal protocol.
  • Modular probe compatibility: accommodates interchangeable sensor probes with diameters of 10 mm, 12 mm, 15 mm, and 18 mm—enabling simultaneous multi-sensor validation or staggered calibration protocols.
  • Optional RS485 interface enables integration into automated calibration management systems; supports SCPI-like command sets for remote scripting and audit-trail generation (21 CFR Part 11 compliant when used with validated software).
  • External recirculation mode delivers a continuous, temperature- and humidity-stabilized gas stream at 3 L/min—suitable for dynamic testing of flow-through sensors or inline process analyzers.

Sample Compatibility & Compliance

The M2000SP-X is compatible with a broad class of humidity-sensing technologies, including capacitive polymer sensors, chilled-mirror dew point analyzers, aluminum oxide hygrometers, and resistive thin-film devices. Its chamber geometry (250 × 200 × 100 mm standard; optional 300 × 250 × 150 mm extended version) permits calibration of handheld meters, panel-mounted transmitters, and embedded OEM modules. For applications requiring extended data logging or parallel multi-point validation, the M2000SP-XR variant integrates a 12-position horizontal/vertical probe array—supporting GLP-aligned calibration records with timestamped, position-tagged measurement sequences. All operational parameters adhere to the humidity condensation limits defined in ISO 4677 (e.g., ≤80 %RH at 35 °C; ≤40 %RH at 50 °C), ensuring long-term sensor integrity during exposure.

Software & Data Management

The M2000SP-X supports both vendor-supplied calibration sequencing software and third-party control environments via its RS485 serial interface. The native software provides preconfigured calibration templates aligned with ISO 17025 Annex C requirements—including ramp-and-soak profiles, hysteresis sweeps, and repeatability assessments across 5–6 defined RH/temperature setpoints. Each session automatically generates CSV-formatted reports containing raw sensor readings, chamber setpoint deviations, stability durations, and NVLAP traceability metadata. When deployed in regulated environments (e.g., pharmaceutical QC labs), the system can be configured to enforce electronic signatures, audit trails, and role-based access control—fully satisfying FDA 21 CFR Part 11 and EU Annex 11 data integrity expectations.

Applications

  • Primary and secondary calibration of RH sensors in national metrology institutes (NMIs) and accredited calibration laboratories.
  • Verification of environmental monitoring systems (EMS) used in cleanrooms, stability chambers, and cold chain logistics.
  • Performance evaluation of industrial humidity transmitters prior to field deployment in HVAC, semiconductor fabrication, and battery drying processes.
  • Research-grade studies on sensor drift, hysteresis, and long-term stability—leveraging the chamber’s sub-0.2 %RH stability and <5-minute equilibration capability.
  • Supporting ISO 9001 and IATF 16949 internal audit requirements for measurement system analysis (MSA) of humidity-critical production equipment.

FAQ

What accreditation does the M2000SP-X’s calibration certificate hold?
The unit ships with an NVLAP-accredited calibration certificate (Lab Code 200451-0), traceable to NIST SRMs and issued under ISO/IEC 17025:2017 scope.
Can the M2000SP-X generate humidity below 7 %RH?
No—the lower limit is specified at <7 %RH due to inherent limitations of solid-state humidity synthesis at low dew points; for sub-5 %RH work, a chilled-mirror reference standard is recommended.
Is the extended chamber (M2000SP-XR) mechanically interchangeable with the base model?
Yes—the XR variant uses the same chassis and control electronics; only the internal baffle assembly and mounting rails differ, allowing field retrofitting with minimal downtime.
Does the device support automatic pass/fail evaluation against user-defined tolerances?
Yes—when operated with ChemTron’s calibration software, users may define acceptance criteria per point (e.g., ±1.0 %RH at 60 %RH), triggering real-time compliance flags and auto-generated nonconformance logs.
What is the maximum allowable ambient temperature for stable operation?
The chamber is rated for ambient temperatures between 15 °C and 30 °C; operation outside this range may extend equilibration time and affect stated stability specs.