Empowering Scientific Discovery

Tekmar Velocity XPT Purge and Trap System with 72-Position Autosampler

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Origin USA
Manufacturer Type Authorized Distributor
Origin Category Imported
Model Tekmar Velocity XPT
Warranty 6 Months
Instrument Age 4–5 Years
Power Supply 220–240 V, 50/60 Hz, 4 A, 920 W
Operating Temperature 10–30 °C
Relative Humidity 10–90 % RH
Panel Corrosion Resistance pH 1–10
Weight 12.3 kg
Dimensions 21 cm × 44.5 cm × 48 cm
Purge Tube Options 5 mL or 25 mL fritted U-tube, 5 mL or 25 mL needle-type tube
Trap Temperature Range Up to 350 °C (heating), ≥127 °C/min cooling rate
Desorption Rate 800 °C/min
Transfer Line Temp Up to 300 °C (PEEK & stainless steel)
Valve Temp Up to 300 °C (24 V solenoid valves)
DryFlow Moisture Trap Ambient to 350 °C, removes ~90 % water without compromising polar analytes
Flow Control Electronic Mass Flow Controller (MFC), 5–500 mL/min across purge/dry-purge/bake modes
Leak Detection Integrated electronic pressure sensor with automated leak test & diagnostic mode
Foam Monitoring Optional foam sensor and defoaming module
Cooling Time Reduction 25 % faster post-bake trap cooldown
Cycle Time ≤15 min per run (with 11-min purge)
Sample Throughput Up to 100 % increase vs. prior-generation systems
Software Tekmar TO-15 / EPA Method 502.2 / 524.2 / 8260-compliant control suite with audit trail, method validation logging, and GLP/GMP-ready data integrity features

Overview

The Tekmar Velocity XPT Purge and Trap System is a high-performance, EPA-compliant volatile organic compound (VOC) preconcentration platform engineered for laboratories requiring rigorous regulatory adherence, reproducible trace-level analysis, and scalable automation. Based on the fundamental principle of dynamic headspace extraction—where inert gas (N₂ or He) purges VOCs from aqueous or solid-phase samples into a thermally desorbable trap—the Velocity XPT implements advanced fluidic architecture, real-time thermal management, and closed-loop diagnostics to minimize carryover, maximize recovery, and ensure method robustness across US EPA Methods 502.2, 524.2, 8260D, and TO-15. This refurbished unit retains full factory-specification hardware integrity, including its proprietary FFC (Focusing Flow Control) pre-focusing system, which dynamically adjusts carrier gas flow during desorption to sharpen chromatographic peaks and suppress breakthrough—critical for low-concentration environmental and drinking water testing.

Key Features

  • Integrated 72-position autosampler enabling unattended, multi-day batch processing with programmable vial agitation and temperature-controlled sample staging
  • Electronic mass flow controller (MFC) delivering precise, mode-specific gas regulation (5–500 mL/min) across purge, dry-purge, and bake phases—eliminating manual flow calibration drift
  • DryFlow moisture management system operating from ambient to 350 °C, removing ≥90 % of water vapor while preserving recovery of polar VOCs (e.g., vinyl chloride, acrylonitrile)
  • Thermally isolated trap assembly with rapid heating (≤350 °C) and cooling (≥127 °C/min), reducing cycle time to ≤15 minutes (11-min purge standard)
  • Automated leak detection via embedded pressure transducer with pass/fail reporting, diagnostic mode for pinpointing leakage points, and log-traceable validation records
  • Corrosion-resistant front panel (pH 1–10 tolerance) and chemically inert fluidic path (PEEK, electropolished stainless steel, high-temp elastomers) suitable for aggressive matrices including wastewater and landfill leachate
  • Optional foam sensor and mechanical defoamer—prevents column contamination and instrument downtime during high-surfactant sample analysis

Sample Compatibility & Compliance

The Velocity XPT accommodates aqueous samples (drinking water, groundwater, wastewater), soil/sediment slurries (per EPA 5035), and solid-phase microextraction (SPME) eluates. Its trap design conforms to ASTM D3699 and US EPA specifications for trap geometry (0.123″ ±0.002″ OD), adsorbent composition (multi-bed Tenax TA/Carbopack B/Carboxen 1000), and thermal desorption profiles. All operational parameters—including purge time, bake temperature, desorption ramp rate, and transfer line dwell—are fully programmable and locked within validated methods. The system meets 21 CFR Part 11 requirements for electronic records and signatures when paired with Tekmar’s compliant software suite, supporting audit-ready GLP and GMP workflows in contract testing labs and municipal water authorities.

Software & Data Management

Control and data acquisition are managed via Tekmar’s Windows-based TO-15/8260D-compliant software, featuring method templates aligned with EPA SW-846 protocols, user-role access controls, and immutable audit trails. Each run logs instrument status (trap temp, pressure, flow, leak test result), operator ID, timestamp, and sequence position. Raw data files include embedded metadata for LIMS integration, and all method changes trigger version-controlled backups. Software supports IQ/OQ/PQ documentation packages and includes built-in calibration verification tools—ensuring continued compliance after refurbishment and reinstallation.

Applications

  • Regulatory monitoring of VOCs in drinking water per EPA 524.2 and 502.2
  • Groundwater and surface water assessment under CERCLA and RCRA programs
  • Soil gas and landfill leachate screening per EPA Method 5035A
  • Pharmaceutical residual solvent analysis (ICH Q3C) in purified water and excipient matrices
  • Food and beverage safety testing for chlorinated solvents and benzene derivatives
  • Method development and validation in ISO/IEC 17025-accredited laboratories

FAQ

Is this Velocity XPT unit certified for EPA-compliant analysis?
Yes—this refurbished system retains full conformance with US EPA Methods 502.2, 524.2, 8260D, and TO-15 when operated with validated methods and calibrated using NIST-traceable standards.
What does the 6-month warranty cover?
The warranty covers parts and labor for defects in materials or workmanship, including MFC, trap heater, thermal sensors, valve actuators, and autosampler mechanics—excluding consumables (traps, septa, O-rings) and damage from improper operation or non-Tekmar gases.
Can the system be integrated with third-party GC or GC-MS platforms?
Yes—the Velocity XPT uses standard 1/16″ Swagelok fittings and outputs a 0.5–1.0 mL desorption pulse compatible with all major GC inlet types (split/splitless, PTV, cold injection); Tekmar provides interface schematics and timing synchronization protocols.
Does the refurbished unit include original software licenses and method libraries?
Yes—each unit ships with licensed Tekmar software (v4.5+), preloaded EPA method templates, and full documentation including service history, calibration logs, and refurbishment certification.
How is performance verified post-refurbishment?
Each unit undergoes full functional testing: trap desorption efficiency (>95 % recovery for BTEX at 0.5 µg/L), leak integrity (<0.5 sccm at 30 psi), thermal ramp linearity (±1 °C), and autosampler positional accuracy (±0.2 mm)—results documented in the Certificate of Refurbishment.