Empowering Scientific Discovery

Labthink C230H Oxygen Permeation Analyzer (Isobaric Method)

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Labthink
Origin Shandong, China
Manufacturer Type OEM Manufacturer
Model C230H
Test Principle Isobaric Method with Coulometric Oxygen Sensor
Standards Compliance ASTM D3985, ASTM F1307, ASTM F1927, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
Measurement Range 0.01–200 cm³/(m²·day) [film], 0.0007–12.9 cc/(100 in²·day), 0.00005–1 cm³/(pkg·day) [package]
Resolution 0.001 cm³/(m²·day)
Repeatability ±0.01 cm³/(m²·day) or ±2%, whichever is greater
Test Temperature 10–55 °C ±0.2 °C
RH Control 0%, 5–90% ±1% (O₂ side), 100% RH option
Carrier Gas Humidity 0%, 5–90% ±2% (optional)
Sample Area 50 cm²
Max Sample Size 108 mm × 108 mm
Max Thickness ≤3 mm
Test Chamber Count 3
Package Testing Capacity Up to 3 L
Carrier Gas ≥99.999% N₂, inlet pressure ≥0.28 MPa (40.6 psi)
Gas Interface 1/8″ stainless steel tubing
Optional Modules DataShield™ Audit Trail, CFR 21 Part 11 Compliance, GMP Computer System Validation Support

Overview

The Labthink C230H Oxygen Permeation Analyzer is an isobaric-method instrumentation system engineered for precise, trace-level quantification of oxygen transmission through polymeric films, rigid sheets, and intact packaging structures. It operates on the fundamental principle of coulometric oxygen detection: oxygen molecules permeating across a conditioned specimen are swept by ultra-high-purity nitrogen carrier gas into a galvanic coulometric sensor, where oxygen is electrochemically reduced at a constant potential. The resulting current is linearly proportional to the molar flow rate of O₂, enabling direct calculation of oxygen transmission rate (OTR) in standardized units (cm³/m²·day or cm³/pkg·day). Unlike differential-pressure methods, the isobaric configuration maintains equal total pressure on both sides of the test specimen—eliminating stress-induced artifacts and ensuring measurement fidelity for low-permeability barrier materials such as metallized PET, SiOₓ-coated PP, PVDC-laminates, and AlOx-coated solar backsheet substrates. Designed in strict alignment with ASTM D3985 (standard test method for oxygen gas transmission rate through plastic film and sheeting using a coulometric sensor), the C230H delivers metrologically traceable data suitable for regulatory submissions, shelf-life modeling, and QC/QA release testing.

Key Features

  • Triple-test-chamber architecture enables parallel evaluation of up to three specimens under identical environmental conditions—improving throughput without compromising inter-chamber reproducibility.
  • Integrated temperature control system maintains chamber setpoints from 10 °C to 55 °C with ±0.2 °C stability, supporting accelerated aging protocols and real-world storage condition simulation.
  • Independent humidity control on the oxygen-exposed side (0%, 5–90% ±1% RH, or 100% RH) allows investigation of moisture-dependent permeation behavior—critical for hygroscopic polymers like EVOH or PVA-based coatings.
  • High-resolution coulometric sensor achieves a detection limit of 0.001 cm³/(m²·day), validated against NIST-traceable O₂ permeation standards.
  • Modular hardware design supports seamless integration of optional compliance packages—including DataShield™ secure audit trail generation and configurable electronic signatures aligned with FDA 21 CFR Part 11 requirements.
  • Dual-mode operation accommodates both flat specimens (max 108 mm × 108 mm, ≤3 mm thick) and 3D packages (up to 3 L volume), with dedicated fixturing for bottles, pouches, trays, and blister cavities.

Sample Compatibility & Compliance

The C230H accepts a broad spectrum of industrial materials: flexible films (OPP, BOPP, PET, PE, PA, CPP), coated substrates (K-coated paperboard, SiOx-PET, AlOx-PET), laminates (aluminum foil composites, metallized CPP), rigid sheets (PP, PVC, PVDC, aluminum foil, silicone rubber), and finished packages (vacuum-sealed pouches, HDPE bottles, aluminum collapsible tubes, thermoformed trays). Its mechanical clamping system ensures uniform gasket compression and leak-tight sealing across variable geometries. Regulatory alignment includes full support for ASTM F1307 (packaging), ASTM F1927 (high-barrier films), GB/T 31354 (Chinese food-contact packaging), and YBB 00082003-2015 (pharmaceutical packaging). All firmware and software modules undergo periodic verification per ISO/IEC 17025:2017 clause 5.9.1, and raw data files retain immutable timestamps, operator IDs, and environmental logs required for GLP and GMP audits.

Software & Data Management

Control and analysis are executed via Labthink’s proprietary OX-TEST v4.x platform—a Windows-based application compliant with IEC 62304 Class B software safety requirements. The interface provides real-time OTR curve visualization, automatic endpoint detection based on signal stabilization criteria (±0.5% over 30 min), and batch report generation in PDF/XLSX formats with embedded metadata (test ID, calibration certificate number, environmental log, operator signature). DataShield™ module enforces ALCOA+ principles: attributable, legible, contemporaneous, original, accurate, complete, consistent, enduring, and available. Electronic records include SHA-256 hash validation, role-based access control (RBAC), and configurable retention policies. Optional CFR 21 Part 11 add-on delivers electronic signature workflows, audit trail review tools, and system validation documentation packages for pharmaceutical and medical device manufacturers.

Applications

  • Food packaging R&D: Quantifying OTR of snack pouches, coffee bags, and modified-atmosphere packaging (MAP) films to predict oxidative rancidity onset.
  • Pharmaceutical primary packaging: Validating OTR of cold-form blisters, child-resistant closures, and desiccant-containing bottles per USP and Ph. Eur. 5.3.
  • PV module encapsulation: Measuring oxygen ingress through ethylene-vinyl acetate (EVA) and POE encapsulants to assess potential PID (potential-induced degradation) risk.
  • Aerospace & automotive: Evaluating fuel tank liner permeability (e.g., nylon-6/PA6 blends) per SAE J1737 and ISO 11600.
  • Battery enclosure qualification: Assessing O₂ diffusion through lithium-ion battery cell housings to mitigate electrolyte oxidation and SEI layer instability.
  • Regulatory dossier preparation: Generating test reports admissible for FDA 510(k), CE Marking (MDR Annex II), and CFDA registration dossiers.

FAQ

What is the minimum detectable OTR value for film testing?
The instrument achieves a resolution of 0.001 cm³/(m²·day) under standard conditions (23 °C, 0% RH on N₂ side), verified using certified low-permeability reference films.
Can the C230H test non-planar packages without custom fixtures?
Yes—the system includes universal package adapters accommodating cylindrical, rectangular, and irregularly shaped containers up to 3 L volume; no machining or bespoke tooling is required for standard configurations.
Is carrier gas purity critical for measurement accuracy?
Absolutely. Use of ≥99.999% N₂ is mandatory to avoid background O₂ interference; impurities >1 ppm will elevate baseline signal and compromise repeatability below 1 cm³/(m²·day).
How does the C230H handle humidity-controlled testing when evaluating hydrophilic barrier layers?
The O₂-side RH chamber employs chilled-mirror dew point control with closed-loop feedback, maintaining setpoints within ±1% RH tolerance across the 5–90% range—essential for characterizing EVOH or PVOH performance decay under humid conditions.
Does the system support IQ/OQ/PQ documentation for GMP environments?
Yes—Labthink provides vendor-supplied qualification protocols, factory-verified calibration certificates, and on-site installation/commissioning support aligned with Annex 15 and ASTM E2500.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0