Empowering Scientific Discovery

Junyi JY600HE Electrophoresis Power Supply

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Junyi
Origin Beijing, China
Model JY600HE
Output Modes Constant Voltage / Constant Current / Constant Power
Voltage Range 5–600 V (1 V resolution)
Current Range 1–1000 mA (1 mA resolution)
Power Range 1–300 W (1 W resolution)
Timing Range 1 min – 99 h 59 min
Volt-Hour Range 1–99999 V·h (1 V·h increment)
Display Backlit LCD
Output Ports 4 independent pairs
Dimensions (L×W×H) 263 × 275 × 125 mm
Net Weight 4.4 kg
Protection Features Overload, Open-Circuit, Leakage Detection
Memory Capacity 100 user-defined protocols
Stepwise Programming Up to 10 sequential programs, each with up to 10 steps
Pause & Real-Time Adjustment Supported without compromising output stability

Overview

The Junyi JY600HE Electrophoresis Power Supply is a microprocessor-controlled, multi-mode DC power source engineered for precision and reproducibility in nucleic acid and protein electrophoresis applications. It operates on fundamental principles of Ohm’s Law-based current regulation and closed-loop feedback control to deliver stable, programmable voltage, current, or power outputs—enabling consistent migration kinetics across diverse gel matrices including agarose, polyacrylamide (PAGE), SDS-PAGE, native PAGE, IEF gels, and cellulose acetate membranes. Designed for benchtop integration in academic, clinical, and QC laboratories, the JY600HE supports routine analytical separations as well as preparative-scale electrophoresis workflows requiring extended run times and high reproducibility.

Key Features

  • Triple-output mode selection: constant voltage (CV), constant current (CC), or constant power (CP), each adjustable with 1 V, 1 mA, and 1 W resolution respectively
  • Dual-core microprocessor architecture with real-time monitoring and adaptive regulation to maintain setpoint accuracy under variable load conditions
  • Backlit LCD interface displaying both preset and real-time output values simultaneously, enabling immediate verification of operational parameters
  • 100-method internal memory with non-volatile storage—retains protocols after power cycling and supports rapid recall during high-throughput workflows
  • Stepwise program capability: users can define up to 10 sequential stages per method, with each stage specifying independent voltage/current/power and timing parameters—ideal for gradient gels, multi-phase transfers, or staged blotting protocols
  • Intelligent pause function with live parameter adjustment: operators may temporarily halt execution and fine-tune settings without resetting timers or compromising output stability upon resume
  • Comprehensive safety architecture including automatic shutdown upon open-circuit detection, overcurrent limiting, thermal overload protection, and earth leakage monitoring compliant with IEC 61010-1 safety standards

Sample Compatibility & Compliance

The JY600HE is compatible with standard horizontal electrophoresis tanks (e.g., for agarose DNA analysis), mini- and midi-vertical slab units (including multi-gel comb configurations), capillary electrophoresis modules (when interfaced via external adapters), and electroblotting systems used in Western/Northern/Southern transfer workflows. Its 600 V maximum output accommodates high-resolution native PAGE and isoelectric focusing (IEF) applications requiring elevated field strengths. The instrument meets CE marking requirements for laboratory electrical equipment and conforms to ISO/IEC 17025-relevant environmental and operational robustness criteria when deployed in GLP-compliant environments. While not FDA 21 CFR Part 11–certified out-of-the-box, its audit-trail-capable protocol logging (via optional USB export) supports alignment with GMP documentation practices where traceability of electrophoretic conditions is required.

Software & Data Management

The JY600HE operates as a standalone instrument with no mandatory PC dependency; however, it supports protocol export via USB interface (optional cable) for offline archiving and cross-laboratory method harmonization. All stored methods include timestamped metadata (creation date, last modification, operator ID field). The device logs runtime statistics—including cumulative volt-hours per method—which are critical for equipment utilization tracking and preventive maintenance scheduling. Exported files follow plain-text CSV format, ensuring compatibility with LIMS platforms and spreadsheet-based QA review workflows. No proprietary drivers or cloud services are required; data integrity is preserved through EEPROM-backed memory with >100,000 write-cycle endurance.

Applications

  • Routine DNA fragment analysis using agarose and low-percentage polyacrylamide gels
  • High-resolution protein separation via SDS-PAGE, native PAGE, and two-dimensional electrophoresis (2-DE) first-dimension IEF strips
  • Seed purity testing and varietal identification in agricultural biotechnology labs
  • Electroelution and electrotransfer (semi-dry and wet blotting) of proteins and nucleic acids onto PVDF or nitrocellulose membranes
  • Capillary zone electrophoresis (CZE) support when paired with compatible electrode chambers and buffer reservoirs
  • Quality control of recombinant proteins, monoclonal antibodies, and oligonucleotide therapeutics per USP and EP 2.2.47 guidelines

FAQ

What electrophoresis tank configurations are supported by the JY600HE?
It interfaces with all standard commercial horizontal and vertical electrophoresis systems that accept 4-mm banana plug inputs—including Bio-Rad, Thermo Fisher, C.B.S. Scientific, and locally manufactured units meeting IEC 60130-10 specifications.
Does the JY600HE support timed ramping of voltage or current?
No—ramping is not implemented. However, stepwise programming enables discrete transitions between up to 10 predefined output levels with independent timing, effectively simulating controlled ramp profiles for applications such as gradient gel polymerization or progressive transfer conditions.
Can the device be integrated into automated lab workflows?
Yes—via TTL-compatible trigger input (optional accessory), allowing synchronization with external devices such as UV transilluminators, gel documentation systems, or robotic liquid handlers for unattended batch processing.
Is firmware upgrade capability available?
Firmware updates are performed via USB using manufacturer-provided utilities; version history and release notes are published on Junyi’s global technical support portal for registered institutional users.
What is the recommended calibration interval?
Annual calibration against NIST-traceable reference meters is advised for laboratories operating under ISO/IEC 17025 or CAP accreditation; factory calibration certificates include uncertainty budgets at key points across the 5–600 V and 1–1000 mA ranges.