Empowering Scientific Discovery

Boxun BXQ-1000 Full-Range Precision Biochemical Incubator

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Boxun
Origin Shanghai, China
Manufacturer Type Direct Manufacturer
Region Category Domestic (China)
Model BXQ-1000
Price Range USD 7,000–14,000 (FOB Shanghai)
Temperature Range 0–80 °C
Resolution 0.1 °C
Temperature Uniformity ±1 °C (at 25 °C)
Temperature Fluctuation ±0.5 °C (at 25 °C)
Volume 1000 L
Interior Dimensions (mm) 1363 × 670 × 1123 (standard chamber) / 1505 × 1006 × 1811 (overall external envelope)
Shelf Capacity 6 standard shelves (up to 12)
Load per Shelf 15 kg
Power Input 1000 W
Timer Range 0–9999 min or 0–9999 h (selectable)
Calendar-Based Programming Yes
Ramp/Soak Control 30 segments × 99 cycles
Airflow Control 6-level adjustable fan speed
Cooling System Inverter-driven, frost-free, eco-refrigerant (R134a or equivalent)
Noise Level ≤45 dB(A)
Data Logging Internal non-volatile memory (≥5 years), USB export in immutable .csv format
Compliance Designed to support GLP/GMP workflows

Overview

The Boxun BXQ-1000 is a full-range, microprocessor-controlled biochemical incubator engineered for precision temperature management across demanding life science applications—from low-temperature microbial preservation to high-temperature enzymatic assays and plant tissue culture. Unlike conventional incubators limited to ambient-plus operation, the BXQ-1000 delivers stable, uniform thermal control from 0 °C to 80 °C—enabling uninterrupted BOD5 testing, psychrophilic and thermophilic strain cultivation, pharmaceutical stability studies (ICH Q1A–Q1E), and long-term seed germination trials. Its core architecture integrates a closed-loop inverter refrigeration system with adaptive PID+ fuzzy logic control (BRIGHT I algorithm), which dynamically compensates for ambient fluctuations, door openings, and load changes—ensuring reproducible setpoint adherence without manual recalibration. The unit complies with fundamental design expectations for regulated environments: all critical operational events—including parameter modifications, alarm triggers, and power interruptions—are time-stamped and stored in tamper-resistant internal memory, supporting traceability requirements under GLP (Good Laboratory Practice) and pre-GMP validation protocols.

Key Features

  • Full-temperature range operation (0–80 °C) with ±0.5 °C fluctuation (at 25 °C) and ±1 °C uniformity across the 1000 L chamber
  • 5-inch capacitive LCD touchscreen interface with intuitive icon-based navigation and multilevel password protection (admin/operator/user tiers)
  • Inverter-driven refrigeration system using environmentally compliant refrigerant, enabling continuous frost-free operation across the entire temperature span
  • Six-stage programmable airflow control—adjustable fan speed minimizes evaporative loss for sensitive cultures while maintaining homogeneity
  • Dual-door structure with full-view tempered glass inner door reduces thermal disturbance during observation; integrated condensate recovery channels prevent pooling
  • Robust safety architecture: independent overtemperature cutoff, door-open alarm, compressor overheat protection, leakage current detection, and dual-stage overcurrent safeguards
  • Ergonomic service design: front-accessible condenser maintenance port, rounded interior corners for ISO-compliant cleaning, and dual 30 mm validation ports (left/right side)

Sample Compatibility & Compliance

The BXQ-1000 accommodates diverse biological and chemical sample formats—including Petri dishes, multiwell plates, Erlenmeyer flasks, serum bottles, and custom bioreactor vessels—across its 1000 L working volume. Standard shelving supports up to 12 trays (6 included), each rated for 15 kg static load, with optional reinforced racks available for heavier instrumentation integration. Chamber materials meet USP Class VI biocompatibility standards; interior stainless steel (SUS304) resists corrosion from repeated ethanol, hydrogen peroxide, or sporicidal agent exposure. The system supports IQ/OQ/PQ documentation packages and includes factory-calibrated PT100 sensors traceable to NIM (National Institute of Metrology, China). While not certified to UL 61010-1 or IEC 61000-6-3 out-of-the-box, its electrical architecture conforms to essential safety and EMC principles required for CE-marked laboratory equipment in non-medical research settings.

Software & Data Management

Data integrity is enforced through an embedded real-time operating system that logs all operational parameters—including chamber temperature, setpoint history, alarm status, fan speed, and door events—at 1-minute intervals. Logs are retained internally for ≥5 years without SD card dependency and exported via USB 2.0 in write-protected .csv format, preserving timestamps and preventing post-hoc editing. Event records include ISO 8601-compliant timestamps, operator ID (if authenticated), and contextual metadata (e.g., “Alarm: T_sensor_failure @ 2024-06-12T14:22:07Z”). Optional RS-485 Modbus RTU output enables integration into centralized building management systems (BMS) or LIMS platforms. The controller supports calendar-based scheduling (public-date programming) and 30-segment, 99-cycle ramp/soak profiles—critical for simulating diurnal temperature shifts in ecological studies or accelerated degradation protocols.

Applications

  • Biochemical oxygen demand (BOD5) analysis per APHA Standard Methods 5210B and ISO 5815-1
  • Cultivation of mesophilic, thermophilic, and psychrotrophic bacteria, yeasts, and molds for QC microbiology (USP , EP 2.6.12)
  • Plant tissue culture, seed germination, and photoperiod-independent growth studies
  • Enzyme activity assays requiring precise thermal staging (e.g., Taq polymerase activation at 95 °C followed by 37 °C extension)
  • Stability testing of raw materials and intermediates under ICH Q1A(R3) long-term and accelerated conditions
  • Environmental simulation for aquatic toxicology (OECD 201, 210) and soil microcosm experiments

FAQ

Does the BXQ-1000 support 21 CFR Part 11 compliance?
The device provides foundational audit-trail capabilities (immutable logs, user authentication, electronic signatures via password tiers), but full Part 11 compliance requires additional procedural controls, system validation, and third-party software integration—not supplied as standard.
Can the incubator operate continuously at 0 °C or 80 °C?
Yes—its inverter-driven refrigeration and high-efficiency heating elements enable stable, low-drift operation across the full specified range without defrost cycles or performance degradation.
Is the internal power outlet compatible with international voltage standards?
The standard configuration includes a 220 V AC, 10 A waterproof socket (GB/T 1002); for 115 V markets, custom power supply modules and socket variants are available upon order specification.
What validation support documentation is provided?
Factory calibration certificates (temperature uniformity, sensor accuracy), IQ/OQ templates, and sensor traceability reports to NIM are included. PQ execution remains customer-responsible per local regulatory interpretation.
Are spare parts and firmware updates available internationally?
Yes—Boxun maintains global distributor networks with 18-month minimum spare parts availability; firmware revisions are distributed via secure portal with version-controlled release notes and change logs.