Empowering Scientific Discovery

Titan DMC-12K Digital Mini High-Capacity Benchtop Centrifuge

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Titan
Origin Shanghai, China
Manufacturer Type Direct Manufacturer
Instrument Type Benchtop Centrifuge
Speed Range 0–12,000 rpm
Max RCF 9,660 × g
Max Capacity 8 × 50 mL tubes (angle rotor configuration)
Dimensions (W×D×H) 194 × 229 × 120 mm
Regulatory Classification General-purpose low-speed centrifuge for non-hazardous sample processing

Overview

The Titan DMC-12K is a compact, digitally controlled benchtop centrifuge engineered for routine separation tasks in academic laboratories, quality control units, and clinical support environments. It operates on the principle of sedimentation under radial acceleration—applying controlled rotational force to separate components based on density differentials in heterogeneous liquid samples. Designed for reliability and operational simplicity, the DMC-12K delivers consistent performance at speeds up to 12,000 rpm and a maximum relative centrifugal force (RCF) of 9,660 × g. Its miniaturized footprint (194 × 229 × 120 mm) enables deployment on crowded workbenches without compromising mechanical stability or thermal management. The unit complies with IEC 61010-1 safety standards for laboratory electrical equipment and incorporates fail-safe rotor detection, automatic lid interlock, and over-speed protection circuits.

Key Features

  • Digital LCD interface with real-time display of speed (rpm), time (0–99 min 59 s), and remaining run time
  • Precision brushless DC motor ensuring low vibration, extended service life, and minimal heat generation during prolonged operation
  • Integrated angle rotor (standard configuration) supporting eight 50 mL conical tubes—compatible with common polypropylene and polycarbonate centrifuge tubes meeting ISO 8573-1 purity class requirements
  • Programmable memory for up to 10 user-defined protocols, including ramp-up/down profiles and hold-at-speed functions
  • Quiet operation (<55 dB[A] at 12,000 rpm) achieved through balanced rotor assembly and acoustic-dampening housing design
  • Thermal management system with passive airflow channels to maintain ambient rotor chamber temperature within ±3 °C during continuous 30-minute runs

Sample Compatibility & Compliance

The DMC-12K accommodates standard 50 mL conical centrifuge tubes (e.g., Falcon® 352070, Corning® 430829) and is validated for use with aqueous buffers, cell suspensions, serum, plasma, and particulate-laden industrial emulsions. It is not certified for biosafety level 2+ (BSL-2+) containment or high-risk pathogen handling. All rotor assemblies are manufactured to ISO 21501-4 specifications for dimensional tolerance and dynamic balance. The instrument meets CE marking requirements under Directive 2014/30/EU (EMC) and 2014/35/EU (LVD), and conforms to GLP-aligned documentation practices for method traceability in regulated QC workflows.

Software & Data Management

While the DMC-12K operates as a standalone instrument without external PC connectivity, its embedded firmware supports full audit trail logging—including start/stop timestamps, applied parameters, error codes, and cycle count—stored in non-volatile memory (retention >10 years). Data export is enabled via optional RS-232 serial interface (cable and terminal software provided separately), allowing integration into LIMS environments compliant with FDA 21 CFR Part 11 when paired with validated third-party data capture modules. All parameter changes require dual-button confirmation to prevent accidental modification during active runs.

Applications

  • Routine pelleting of mammalian cells (e.g., HEK293, CHO-K1) at 3,000–6,000 × g for downstream lysis or media exchange
  • Clarification of bacterial lysates prior to chromatographic purification
  • Sedimentation of precipitated proteins or nucleic acid complexes in buffer exchange protocols
  • Phase separation in solvent extraction workflows (e.g., phenol-chloroform isolation)
  • Pre-analytical sample conditioning for clinical chemistry analyzers requiring supernatant collection
  • Quality assurance testing of suspension-based pharmaceutical formulations (e.g., oral suspensions, vaccine adjuvant dispersions)

FAQ

Is the DMC-12K suitable for centrifuging infectious samples?
No. This model lacks sealed rotor containment, biocontainment certification, or aerosol-tight lid mechanisms required for BSL-2 or higher applications.
Can I use swing-bucket rotors with this centrifuge?
No. The DMC-12K is mechanically configured exclusively for fixed-angle rotors; no swing-bucket or horizontal rotor adapters are available or supported.
What maintenance intervals are recommended?
Visual inspection of rotor threads and lid seal integrity before each use; full lubrication and dynamic balancing verification every 12 months or after 500 operating hours—whichever occurs first.
Does the unit support timed deceleration control?
Yes. Users may select “soft stop” mode to limit deceleration rate, reducing resuspension risk for delicate pellets.
Is calibration traceable to national standards?
Speed calibration is performed using NIST-traceable tachometric reference instruments; certificate of conformance is supplied with each unit shipment.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0