SHYP JA1003C Automatic Internal Calibration Precision Balance

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand SHYP
Origin Shanghai, China
Model JA1003C
Maximum Capacity 100 g
Readability 1 mg
Calibration Type Internal
Pan Diameter 90 mm
Net Weight 6.8 kg
Dimensions (W×D×H) 350 × 215 × 340 mm
Packaging Dimensions 480 × 375 × 515 mm
Interface RS232C
Compliance ISO/IEC 17025-compatible operation environment
Software Features Auto-zero tracking, full-range tare, piece counting, unit conversion (carat, troy ounce), stability indicator

Overview

The SHYP JA1003C is a high-reliability precision balance engineered for routine laboratory weighing tasks requiring consistent 1 mg readability and traceable measurement integrity. Based on electromagnetic force compensation (EMFC) sensor technology, the JA1003C delivers stable, drift-resistant performance across environmental fluctuations typical in QC labs, pharmaceutical production support areas, and academic teaching laboratories. Its fully automatic internal calibration system—activated at user-defined intervals or upon power-up—eliminates manual handling of external calibration weights and ensures ongoing metrological compliance without operator intervention. The balance operates within a temperature-stable range of 10–30 °C and relative humidity up to 80 % RH (non-condensing), meeting baseline environmental requirements for GLP-aligned weighing procedures.

Key Features

  • Electromagnetic force compensation (EMFC) transducer with dual-cam actuated internal calibration mechanism for repeatable 1 mg resolution across the full 0–100 g range
  • Five-sided transparent draft shield with optimized airflow geometry to minimize convection-induced weighing instability
  • Integrated RS232C serial interface compliant with standard ASCII command protocols, enabling direct integration with LIMS, ELN, or statistical process control software
  • Backlit LCD display (128 × 64 pixels) with adjustable contrast and real-time status indicators (e.g., “Warm-up in progress”, “Calibrating”, “Stable”)
  • Programmable auto-zero tracking, full-capacity tare, overload protection (up to 150 % of max capacity), and linear four-point calibration verification
  • Built-in below-balance weighing hook for density determination and suspended sample applications

Sample Compatibility & Compliance

The JA1003C accommodates samples up to 100 g on its 90 mm stainless-steel weighing pan, supporting volumetric flasks, crucibles, Petri dishes, and standard USP/Ph. Eur.-compliant containers. Its design conforms to mechanical safety standards per IEC 61010-1 and electromagnetic compatibility per EN 61326-1. While not certified to OIML R76 or NIST Handbook 44 out-of-the-box, the instrument supports audit-ready documentation when operated under documented SOPs aligned with ISO/IEC 17025 clause 6.4.2 (equipment verification) and USP <41> / <1251> guidelines for analytical balance qualification. Internal calibration traceability is maintained via factory-assigned calibration constants stored in non-volatile memory.

Software & Data Management

The embedded firmware supports bidirectional ASCII data transmission at 9600 baud, 8N1 configuration, with configurable print formats (e.g., timestamped weight + unit + stability flag). Optional PC software (sold separately) enables automated data logging, trend analysis, and export to CSV or Excel. Audit trail functionality—including operator ID tagging, calibration event timestamps, and error log history—is available when connected to validated third-party data acquisition platforms compliant with FDA 21 CFR Part 11 requirements. All calibration events generate a machine-readable record including ambient temperature, internal sensor temperature, and deviation from nominal calibration value.

Applications

  • Raw material receipt inspection in food, chemical, and cosmetic manufacturing facilities
  • Weighing reference standards and working standards in analytical method validation (ICH Q2(R2))
  • Preparation of buffer solutions and titrant stock solutions in clinical chemistry labs
  • Teaching fundamental metrology concepts—including repeatability, linearity, and eccentricity testing—in university physics and chemistry curricula
  • Batch release testing where ±1 mg tolerance satisfies specification limits per pharmacopoeial monographs
  • Density and specific gravity measurements using the built-in under-hook suspension assembly

FAQ

What is the recommended warm-up time before first use?
A minimum 30-minute stabilization period is required after power-on to achieve thermal equilibrium of the EMFC sensor and internal reference electronics.
Does the JA1003C meet USP <41> repeatability criteria?
Yes—when operated within specified environmental conditions (±1 °C/hour temperature stability, no drafts), the balance achieves repeatability ≤0.5 RSD (n=10) at 50 g test load, satisfying USP <41> thresholds for Class II balances.
Can external calibration weights be used alongside internal calibration?
Yes—the balance supports manual external calibration via standard Class E2 or F1 weights for periodic verification or as part of a formal equipment qualification protocol.
Is GLP-compliant documentation provided with the instrument?
The unit ships with a factory calibration certificate (traceable to NIM, China), test report, and declaration of conformity; full GxP documentation packages require site-specific IQ/OQ/PQ execution.
What maintenance intervals are recommended for long-term accuracy?
Daily cleaning of the pan and draft shield, weekly verification using certified check weights, and annual sensor diagnostics by an authorized SHYP service center are advised per ISO 17025 clause 6.4.4.