Anaerobic Incubator
Filter
Showing 31–32 of 32 results
| Brand | Thermo Fisher Scientific |
|---|---|
| Origin | USA |
| Model | 1029 |
| Internal Volume | 77 L |
| Operating Temperature Range | Ambient +4°C to 70°C (fixed at 35°C option available) |
| Temperature Uniformity at 37°C | ±0.4°C |
| Temperature Accuracy | ±0.3°C |
| Anaerobic Gas Requirement | High-purity N₂/H₂/CO₂ (e.g., 85/10/5 or 90/5–7/5 vol%) |
| O₂ Residual Level | <10 ppm |
| Catalyst System | Palladium-based with activated carbon H₂S scavenger |
| Gas Consumption Reduction | ~50% vs. conventional systems |
| Work Area Dimensions (W×H×D) | 445×559×305 mm |
| External Dimensions (W×H×D) | 1539×762×787 mm |
| Glove Port | Elliptical, full-access design |
| Viewing Panel | Flexible anti-fog acrylic (optional rigid polycarbonate) |
| Shelf Configuration | Two adjustable stainless-steel shelves in work area |
| Power Input | 230 V / 50 Hz |
| Compatible Vacuum Pump | Recommended model 901025 |
| Optional Enclosure | Thermo Scientific 1016 Outer Cabinet (1092×762×787 mm) |
| Brand | WIGGENS |
|---|---|
| Origin | Germany |
| Model | WCI-850T |
| Instrument Type | Automated Tri-Gas Incubator |
| Temperature Range | Ambient +5°C to 60°C |
| Temperature Uniformity & Stability | ±0.1°C at 37°C |
| O₂ Concentration Range | 0.6%–85% |
| CO₂ Range | 0%–20% |
| CO₂ Accuracy | ±0.1% at 5% / 37°C |
| O₂ Sensor | Zirconia Electrochemical Sensor |
| CO₂ Sensor | Dual-Beam Infrared Sensor |
| Humidification | Heated Water Pan with Forced-Air Distribution |
| Chamber Material | AISI 304 Stainless Steel |
| Heating System | Six-Side Air-Jacketed Gradient Heating with Independent Calibration per Zone |
| Sterilization Options | Optional UV (4 W), O₃, or Dry-Heat (≤100°C) |
| Communication Interface | RS232 / RS485 |
| Compliance | Designed for ISO 13485-aligned lab environments |