| Brand | Acmer |
|---|---|
| Model | MSP30 |
| Type | Industrial Syringe Pump |
| Stroke Accuracy | ±0.5% |
| Repeatability | ±0.05% |
| Flow Rate Calculation | Linear Speed × Cross-sectional Area of Syringe Barrel |
| Syringe Compatibility | 50 µL to 5 mL |
| Max Pressure Rating | 0.5 MPa |
| Linear Speed Range | 0.025–25 mm/s |
| Step Resolution | 0.01 mm (standard mode) or 0.625 µm (microstepping mode) |
| Valve Switching Time | <280 ms |
| Drive Force | ≥68 N |
| Power Supply | DC 24 V / 1.5 A |
| Dimensions (L×W×H) | 110 × 44.4 × 127 mm |
| Weight | 0.8 kg |
| Operating Temperature | 15–40 °C |
| Relative Humidity | <80% |
| Communication | RS485/USB (auto-detect), 9600/38400 bps |
| Addressing | 4-bit BCD switch (15 addresses) |
| I/O | 2 TTL inputs (resume control), 3 TTL outputs (status indication) |
| Valve | 3-port 120° ceramic valve (PCTFE option available) |
| Fluidic Path Materials | Ceramic, borosilicate glass |
| Port Interface | 1/4"-28 internal thread |
| Brand | Acmer |
|---|---|
| Model | DSC-B01/W70B |
| Type | Micro-syringe Pump |
| Flow Rate Range | 1 pL/min to 44.11 mL/min |
| Accuracy | ±0.5% |
| Repeatability | ±0.05% |
| Display | 5-inch HD Touchscreen |
| Control Interface | RS485 / USB (configurable address & baud rate) |
| External Trigger | TTL-compatible digital input (5–24 V or dry contact) |
| Status Outputs | Run/Stop, Direction signals |
| Syringe Compatibility | 1–10 mL plastic syringes and µL-range glass syringes |
| Force Adjustment | Programmable push force based on syringe barrel diameter and material |
| Brand | Acmer |
|---|---|
| Origin | Hebei, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic |
| Model | SP75-1030 |
| Pump Type | Industrial Syringe Pump |
| Flow Rate | 23.1 µL/s to 9.25 mL/s |
| Accuracy | ±1% |
| Repeatability | ±0.5% |
| Syringe Compatibility | 30 mL disposable medical syringes (max. 28.5 mL usable volume) |
| Rated Stroke | 75 mm |
| Linear Speed Range | 0.0625–25 mm/s |
| Position Resolution | 0.025 mm/step (1.56 µm/step) |
| Max Piston Force | ≥500 N |
| Max Fluid Pressure Rating | 0.2 MPa |
| Power Supply | DC 24 V / 5 A |
| Dimensions (L×W×H) | 508 × 245 × 130 mm |
| Weight | 5.92 kg |
| Operating Temperature | 15–40 °C |
| Relative Humidity | <80% RH |
| Communication Interfaces | RS485 & USB (auto-baud detection at 9600/38400 bps) |
| Address Setting | 4-digit BCD DIP switch (15 addresses) |
| Microstepping Modes | 8x or 16x subdivision (48,000 microsteps per full stroke) |
| Programmable Functions | Ramp profiles, target speed, piston velocity, gap compensation, instruction sequencing, delay, loop, and motion termination |
| Brand | Acmer |
|---|---|
| Origin | Hebei, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic |
| Model | LSP H |
| Instrument Type | Micro-syringe Pump |
| Flow Rate Range | 0.0007 nL/min to 109.136 mL/min |
| Accuracy | ±0.5% |
| Repeatability | ±0.05% |
| Brand | Acmer |
|---|---|
| Origin | Hebei, China |
| Manufacturer Type | Authorized Distributor |
| Product Origin | Domestic (China) |
| Model | DSC |
| Instrument Type | Micro-volume Syringe Pump |
| Flow Rate Range | 0.96 pL/min to 2035.01 µL/min |
| Accuracy | ±0.5% |
| Repeatability | ±0.05% |
| Brand | Acmer |
|---|---|
| Origin | Hebei, China |
| Manufacturer Type | Authorized Distributor |
| Product Origin | Domestic (China) |
| Model | LSP A/B |
| Instrument Type | Micro-syringe Pump |
| Flow Rate Range | 0.0007 nL/min to 109.136 mL/min |
| Accuracy | ±0.5% |
| Repeatability | ±0.05% |
| Brand | Acmer |
|---|---|
| Model | LSP-U |
| Pump Type | Microfluidic Syringe Pump |
| Flow Rate Range | 0.0001 nL/min to 21.8273 mL/min |
| Accuracy | ±0.5% |
| Repeatability | ±0.05% |
| Display | 5-inch HD capacitive touchscreen |
| Control Interface | RS485 or USB (configurable address & baud rate) |
| External Trigger | TTL-compatible digital input (5–24 V logic level or dry contact) |
| Status Outputs | Real-time run/stop and direction signals |
| Syringe Calibration | Built-in calibration with user-defined profile storage |
| Power Recovery | Auto-resume operation with preserved runtime parameters after power interruption |
| Brand | Acmer |
|---|---|
| Model | LSP-B01H2 |
| Type | Micro-syringe Pump |
| Flow Rate Range | 1.426 nL/min to 30.41 mL/min |
| Accuracy | ±0.5% |
| Repeatability | ±0.05% |
| Display | 5-inch HD Touchscreen |
| Calibration | User-performed syringe calibration with parameter storage |
| Control Interface | RS485 or USB (configurable address & baud rate) |
| External Control | TTL-compatible digital input (5–24 V DC or dry contact) |
| Status Outputs | Real-time run/stop and direction signals |
| Power Recovery | Auto-resume from last operational state after power interruption |
| Brand | Acmer |
|---|---|
| Origin | Hebei, China |
| Model | EW110B |
| Pump Type | Micro-syringe Pump |
| Flow Rate Range | 1 pL/min – 173.25 mL/min |
| Accuracy | ±0.1% |
| Repeatability | ±0.05% |
| Maximum Linear Stroke | 110 mm |
| Plunger Force | >120 N |
| Linear Speed Range | 0.127 µm/min – 254 mm/min |
| Control Resolution | 0.198 µm per microstep |
| RS485 Address Range | 1–32 |
| Baud Rate | 9600 bps |
| Power Supply | DC 24 V |
| Operating Temperature | 0–40 °C |
| Relative Humidity | 20–80% RH (non-condensing) |
| IP Rating | IP31 |
| Dimensions (W×H×D) | 230 × 38 × 72 mm |
| Weight | 1.1 kg |
| Syringe Compatibility | 0.5 µL – 60 mL standard Luer-lock syringes |
| Brand | Acmer |
|---|---|
| Model | LSP-D02 |
| Type | Microfluidic Syringe Pump |
| Flow Rate Range | 0.0001 nL/min to 109.136 mL/min |
| Accuracy | ±0.5% |
| Repeatability | ±0.05% |
| Display | 5-inch HD Color Touchscreen |
| Control Interface | RS485 / USB (configurable address & baud rate) |
| External Trigger | TTL-compatible 5–24 V logic level or dry contact |
| Force Adjustment | Per-channel programmable push force |
| Compliance | GLP-supporting operation log, audit-ready parameter storage |