Cambridge Sensotec
Filter
Showing all 2 results
| Brand | Cambridge Sensotec |
|---|---|
| Origin | United Kingdom |
| Model | 1100 |
| Measurement Range | 1 ppm to 100% O₂ |
| Response Time (T₉₀) | < 5 s (at 1 L/min flow) |
| Accuracy | ±2% of actual O₂ concentration |
| Precision | ±0.5% of reading |
| Operating Temperature | 5–35 °C |
| Power Supply | 110/220 V AC, 50/60 Hz |
| Dimensions | 205 × 160 × 150 mm |
| Weight | 4 kg |
| Display | 16 × 2-character backlit LCD (9 mm height) |
| Warm-up Time | 3–4 min at 20 °C |
| Pump Type | AC-driven linear piston pump |
| Max Free Air Displacement | 4 L/min (0.28 bar typical) |
| Noise Level | ≤40 dB(A) at 1 m |
| Max Inlet Gas Temperature | 55 °C |
| Sensor Lifetime | >35,000 h |
| Analog Outputs | 0–5 V DC and 4–20 mA (user-programmable) |
| Digital Interface | RS232, data stream at 10 Hz |
| Calibration | Two-point, user-selectable gases (e.g., ambient air + zero or span gas) |
| Internal Pressure Compensation | Yes |
| Built-in Diaphragm Pump | Yes |
| Auto-Cleaning Function | Yes |
| Brand | Cambridge Sensotec |
|---|---|
| Origin | United Kingdom |
| Model | Rapidox 6100 |
| Gas Measurement Range | SF₆ 0–100% (±0.5% FS), H₂O −60 to +20 °C dew point (±2 °C), SO₂ 0–100 ppm (±2%), HF 0–25 ppm (±1 ppm), CF₄ 0–80% (±1%), O₂ 0–100% (±1%) |
| Analysis Time | 3–5 minutes per sample |
| Enclosure | IP66-rated Peli™ carry case |
| Weight | 17 kg |
| Internal Compressor Output | 10 bar |
| Battery Life | Up to 8 hours (Li-ion) |
| Display | 7-inch full-color touchscreen |
| Data Export | USB flash drive (Excel-compatible .csv format) |
| Onboard Thermal Printer | Yes |
| Language Options | Multi-language UI with password-protected access |