microLAN
Filter
Showing all 4 results
| Brand | microLAN |
|---|---|
| Origin | Netherlands |
| Model | ALGcontrol |
| Measurement Parameters | Chlorophyll a (Chl-a), Dissolved Organic Matter (DOM), Turbidity |
| Wavelengths | 365, 450, 525, 570, 590, 615, 710 nm |
| Chl-a Range | 0–200 µg/L |
| Chl-a Resolution | 0.2 µg/L |
| Turbidity Range | 0–400 NTU |
| Detector | DTGS with 24-bit ADC sampling |
| Interface | LAN, CAN-Bus, RS232, RS485, Modbus TCP/RTU |
| Analog Outputs | 2 × 4–20 mA |
| Power Supply | 24 VDC |
| Power Consumption | 45 W |
| Enclosure Material | Aluminum |
| IP Rating | IP54 (optional IP65) |
| Dimensions (H×W×D) | 450 × 450 × 260 mm |
| Operating Sample Temp. | 10–35 °C |
| Ambient Temp. | 15–30 °C |
| Sample Flow Rate | 2–10 L/h |
| Cleaning System | Automated NaOCl-based (≤0.05% active chlorine) |
| OS | Embedded Linux |
| Software | SQL-based PC software with audit-ready data logging |
| Brand | microLAN |
|---|---|
| Country of Origin | Netherlands |
| Model | BACTcontrol |
| Measurement Principle | Enzymatic Fluorescence Assay |
| Target Analytes | β-Galactosidase (coliforms), β-Glucuronidase (E. coli), Alkaline Phosphatase (total viable biomass) |
| Measurement Cycle | 1–2 h |
| Detection Method | Substrate-specific fluorogenic enzymatic reaction |
| Sample Flow Rate | 3 L/h |
| Sample Temperature Range | 10–35 °C |
| Ambient Operating Temperature | 15–30 °C |
| Power Supply | 220 V / 50 Hz or 110 V / 60 Hz |
| Power Consumption | <100 W |
| Enclosure Rating | IP54 (optional IP65) |
| Dimensions (H×W×D) | 460 × 450 × 321 mm |
| Weight | 25 kg |
| Housing Material | Aluminum |
| Max. Sample Pressure | 0.05 bar |
| Interface | 2× USB 2.0 (Type A), 2× Gigabit Ethernet (RJ45), Modbus TCP/RTU, 2× digital inputs, 2× relay outputs, 1× 4–20 mA analog output |
| Optional | UMTS/ISDN modem slot, dual-sample inlet with auxiliary cleaning module, climate control unit (>30 °C ambient), automated NaOCl-based cleaning system (≤0.05% active chlorine) |
| Brand | microLAN |
|---|---|
| Origin | Netherlands |
| Model | iTOXcontrol |
| Instrument Type | Online |
| Biological Indicator | *Aliivibrio fischeri* (formerly *Vibrio fischeri*) |
| Operating Temperature | 5–40 °C |
| Reagent Temperature Range | 15–30 °C |
| Repeatability | Relative Error ≤5% |
| Sample Throughput | 10 mL/h (4.5 mL per test, 2 tests/h) |
| Measurement Duration | 40 min |
| Enclosure Rating | IP31 |
| Power Supply | 220 VAC, 50 Hz or 110 VAC, 60 Hz |
| Dimensions (D×W×H) | 50×50×110 cm (plus 73 cm with trolley) |
| Weight | 90 kg (excl. optional air conditioning unit: +20 kg) |
| Brand | microLAN |
|---|---|
| Origin | Netherlands |
| Model | TOX mini |
| Instrument Type | Portable |
| Biological Indicator | Aliivibrio fischeri (formerly Vibrio fischeri) |
| Operating Temperature | 15–30 °C |
| Reagent Storage & Use Temperature | 15–30 °C |
| Detection Principle | Bioluminescence Inhibition Assay |
| Compliance | EN ISO 11348-3 |
| Detector | Ultrafast Single-Photon Counting Photomultiplier Tube (PMT) |
| Wavelength Range | 380–630 nm |
| Data Storage Capacity | 2000 measurements |
| Power Supply | Rechargeable 3×Ni-Cd battery or external AC adapter (230 V/50 Hz or 115 V/60 Hz) |
| Dimensions (H×W×D) | 170×150×280 mm |
| Weight | 2.0 kg (with battery) |
| Sample Vessel | Standard 15 mm diameter × 47–75 mm height test tubes |
| Humidity Range | 10–90 % RH (non-condensing) |
| Onboard Templates | 6 preconfigured toxicity test protocols |