SCIEX
Filter
Showing 31–35 of 35 results
| Brand | SCIEX |
|---|---|
| Origin | USA |
| Model | X500R QTOF |
| Instrument Type | Quadrupole-Time-of-Flight (QTOF) |
| Configuration | LC-MS/MS Platform |
| Ion Source | Turbo V™ Ion Source (N-type ion path architecture) |
| Mass Accuracy | < 2 ppm (internal calibration), < 3 ppm (external calibration, over 24 h) |
| Resolution | ≥ 35,000 FWHM (at m/z 622) |
| Mass Range | m/z 5–2,000 |
| Dynamic Range | > 5 orders of magnitude (linear) |
| Acquisition Modes | IDA, MRMHR, SWATH® Acquisition, MSᴱ, Full Scan + MS/MS |
| Software | SCIEX OS v2.4 or later |
| Compliance | Designed for GLP/GMP environments |
| Brand | SCIEX |
|---|---|
| Origin | USA |
| Model | X500R QTOF System |
| Application Domain | General-purpose LC-MS/MS Analysis |
| Compliance | Not registered as medical device in China |
| Brand | SCIEX |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | SCIEX ZenoTOF 7600+ System |
| Instrument Type | Hybrid Quadrupole-Time-of-Flight (Q-TOF) Mass Spectrometer |
| Application Scope | Broad-spectrum Proteomics, Biomarker Discovery, Translational Research, Clinical Assay Development |
| Acquisition Mode | ZT Scan DIA (Data-Independent Acquisition with Quadrupole Dimensionality), DDA, Targeted MS/MS |
| Scan Speed | Up to 640 Hz |
| Fragmentation Technology | Electron-Activated Dissociation (EAD) |
| Mass Accuracy | < 2 ppm (internal calibration), < 3 ppm (external calibration) |
| Mass Resolution | > 35,000 FWHM (at m/z 200) |
| Dynamic Range | > 4 orders of magnitude |
| Brand | SCIEX |
|---|---|
| Origin | USA |
| Model | SCIEX ZenoTOF 8600 System |
| Instrument Type | Quadrupole-Time-of-Flight (Q-TOF) Mass Spectrometer |
| Application Scope | Universal (Small Molecules to Intact Proteins) |
| Ion Source | OptiFlow Pro Turbo Ionization Source |
| Mass Analysis Technology | Zeno™ Trap-Enhanced TOF with High-Frequency Pulsed Injection |
| Max MS/MS Scan Rate | 858 Hz (dependent on mass range and acquisition mode) |
| Quadrupole Scanning Speed | up to 4000 Da/s |
| Fragmentation Modes | CID, EAD (Electron-Activated Dissociation) up to 50 eV in both positive and negative ion modes |
| Vacuum System | Oil-Free Dry Pumping |
| Software Platform | SCIEX OS 4.0 with Real-Time Performance Monitoring, Health Diagnostics & Auto-Tuning |
| Brand | SCIEX |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | SCIEX ZenoTOF® 7600 System |
| Instrument Type | Quadrupole-Time-of-Flight (Q-TOF) Mass Spectrometer |
| Application Scope | Universal |
| Mass Range | >40,000 Da |
| Sensitivity Enhancement | 5–20× vs. prior-generation Q-TOF |
| Resolution | >42,000 (FWHM at m/z 1,222) |
| Mass Accuracy | <1 ppm (external calibration) |
| Pressure Range | 0–1024 bar (HPLC/UHPLC compatible) |