| Brand | Shanghai Jinsen / Insment |
|---|---|
| Origin | Shanghai, China |
| Model | KS-84CPA |
| Compliance Standard | CPAI-84 (1995 Edition) |
| Test Method | Vertical Flame Test |
| Sample Mounting | Stainless Steel Clip Fixture |
| Flame Source | Methane Gas Burner |
| Flame Height | 38 ± 2 mm (adjustable) |
| Ignition Duration | 12 s (pre-set, adjustable) |
| Afterflame Time Measurement Range | 1.0–999.9 s (digital display, manual hold function) |
| Specimen Dimensions | Bottom material: 230 × 230 mm (±3 mm), Wall & roof materials: 90 × 300 mm (±3 mm), four specimens per orientation |
| Burner Dimensions | 320 × 320 × 760 mm (W×D×H) |
| Gas Inlet Pressure | 2.5 ± 0.25 psi |
| Methane Purity | ≥97% |
| Observation Window | Transparent tempered glass |
| Air Ventilation | Integrated passive airflow ports |
| Control System | Electromechanical interface with manual flame adjustment valve |
| Reference Flame Gauge | 1.6 mm diameter pilot tube, positioned 3 mm from burner edge, flame length 3 mm |
| Specimen Holder Alignment | Centered vertical positioning via fixed clamping block |