TSI
Filter
Showing all 8 results
| Brand | TSI |
|---|---|
| Origin | USA |
| Model | 8533EP |
| Instrument Type | Real-time Aerosol Mass Concentration Monitor |
| Detection Principle | 90° Light Scattering (Laser Photometry) |
| Mass Concentration Range | 0.001–150 mg/m³ |
| Simultaneous Size-Fractionated Outputs | PM₁, PM₂.₅, Respirable (PM₄), PM₁₀, and Total PM (<15 µm) |
| Flow Rate | 3.0 L/min (±5% accuracy, internally controlled) |
| Optical System | Sheath-Air Isolated Laser Cavity |
| Display | 5.7″ VGA Color Touchscreen |
| Data Storage | 5 MB internal memory (>60,000 data points |
| Operating Temperature | 0–50 °C |
| Storage Temperature | −20–60 °C |
| Relative Humidity | 0–95% RH (non-condensing) |
| Time Constant | User-selectable (1–60 s) |
| Resolution | ±0.1% of reading or 0.001 mg/m³ (whichever is greater) |
| Zero Stability | ±0.002 mg/m³ over 24 h |
| Particle Size Range | 0.1–15 µm |
| Power | Rechargeable Li-ion battery (2.0 kg w/1 battery |
| Connectivity | USB, Ethernet (8533EP only), USB mass storage support |
| Analog Output | User-selectable 0–5 V or 2–20 mA |
| Alarm Output | Relay (non-latching MOSFET) or audible buzzer |
| Filter Holder | Replaceable 37 mm filter cassette (user-supplied) |
| Compliance | Designed for use in occupational health monitoring per OSHA, NIOSH, and ISO 7708 air sampling protocols |
| Brand | TSI |
|---|---|
| Origin | USA |
| Model | RPM10 |
| Sensor Technology | Thermal Mass Flow-Based Differential Pressure Sensing |
| Compliance | ANSI/ASHRAE/ASHE Standard 170-2008, FGI Guidelines 2010, CDC Environmental Infection Control Guidelines, USP <797> |
| Alarm Types | Visual (LED status ring + touchscreen icon) and Audible (adjustable threshold & delay) |
| Operating Modes | Isolation (AII/PE), Neutral/Non-Isolation |
| Interface | Local touch-screen programming, BACnet MS/TP or Modbus RTU (optional), physical mode toggle switch |
| Security | Password-protected configuration access |
| Display | Color LCD with intuitive status visualization (e.g., green = nominal, amber = warning, red = alarm) |
| Enclosure Rating | IP42 |
| Power | 24 VDC (PoE-compatible option available) |
| Brand | TSI |
|---|---|
| Origin | USA |
| Model | SMPS-3938 |
| Size Range | 2.5–1000 nm |
| Resolution | Up to 167 channels per scan |
| Scan Time | <10 s |
| Concentration Range | Up to 1×10⁷ particles/cm³ |
| Compliance | ISO 15900:2009 |
| Detector Options | Butanol or Water-based CPCs |
| Neutralizer Options | Radioactive (⁸⁵Kr) or Non-radioactive (Bipolar Corona) |
| Control Interface | Integrated touchscreen, PC-free operation |
| Software | TSIGuard™ and Instrument Manager™ (PC required for full data analysis) |
| Brand | TSI |
|---|---|
| Origin | USA |
| Model | SOUNDPRO™ SE/DL Series |
| Frequency Range | 1 kHz (reference) |
| Measurement Range | 10–140 dB(A) selectable |
| Self-Noise Floor | ≤10 dB(A) (typical, per IEC 61672-1 Class 1) |
| Accuracy | ±1 dB (per ANSI S1.4 and IEC 61672-1) |
| Battery Life | >10 hours (alkaline, no backlight, basic SLM mode) |
| Brand | TSI |
|---|---|
| Origin | USA |
| Model | 8682-ST |
| Control Channels | Up to 5 independent duct systems |
| Pressure Sensing | Differential pressure transducer input |
| Output Interfaces | Analog (4–20 mA / 0–10 V), Relay alarms, MODBUS RTU (RS-485) |
| Power Supply | 120 VAC to 24 VAC transformer included (25 VA) |
| Enclosure | NEMA 1 indoor-rated |
| Compliance | CE, RoHS |
| Software Integration | Compatible with BMS via open MODBUS protocol |
| Alarm Logic | Adjustable delay to suppress transient false alarms |
| Local Interface | Backlit keypad and LCD display with password-protected configuration |
| Brand | TSI |
|---|---|
| Origin | USA |
| Model | 9565 |
| Compliance | NIST-traceable calibration certificate included |
| Probe Interface | Hot-swap modular probe system |
| Display | Backlit graphical LCD (up to 5 simultaneous parameters) |
| Data Logging | Internal memory supports up to 38.9 days at 1-minute sampling interval |
| Connectivity | USB and Bluetooth® v4.0 |
| Software | LogDat2™ and TrakPro™ Data Analysis Software included |
| Environmental Rating | IP54-rated housing |
| Operating Temperature | 0–50 °C |
| Storage Temperature | –20–60 °C |
| Power | 4× AA alkaline or rechargeable NiMH batteries (typical life: 20 hrs with continuous use) |
| Brand | TSI |
|---|---|
| Origin | USA |
| Model | 9565 / 9565A / 9565P / 9565X |
| Temperature Range (964/966/980/982 probes) | −10 to 60 °C, ±0.3 °C accuracy |
| Temperature Range (960/962 probes) | −18 to 93 °C, ±0.3 °C accuracy |
| Temperature Range (995 vane probe) | 0 to 60 °C, ±0.1 °C accuracy |
| Air Velocity Range (960/962) | 0–50 m/s, ±3% of reading or ±0.015 m/s (whichever is greater), 0.01 m/s resolution |
| Air Velocity Range (995 vane) | 0.25–30 m/s, ±1% of reading or ±0.02 m/s |
| Air Velocity Range (pitot tube) | 31.27–78.7 m/s, ±1.5% at 10.16 m/s |
| CO Range (982 probe) | 0–500 ppm, ±3% of reading or ±3 ppm |
| CO₂ Range (980/982 probes) | 0–5000 ppm, ±3% of reading or ±50 ppm |
| Relative Humidity Range | 0–95% RH, ±3% RH |
| Static/Differential Pressure Range | −28.0 to +28.0 mmHg (−3735 to +3735 Pa), ±1% of reading or ±0.01 mmHg |
| Barometric Pressure Range | 517.15–930.87 mmHg, ±2% of reading |
| Data Storage | 26,500 readings across 100 user-defined data groups |
| Sampling Interval | 1 s to 1 h |
| Operating Temperature | 5–45 °C |
| Dimensions | 9.7 × 21.1 × 5.3 cm |
| Weight (with batteries) | 0.36 kg |
| Power | 4 × AA batteries or optional AC adapter |
| Brand | TSI |
|---|---|
| Origin | USA |
| Model | 9565 / 9565-A / 9565-P / 9565-X |
| Temperature Range (964/966/980/982 probes) | −10 to 60 °C |
| Temperature Accuracy | ±0.3 °C (at 25 °C) |
| Temperature Resolution | 0.1 °C |
| Air Velocity Range (960/962 probes) | 0–50 m/s |
| Air Velocity Accuracy | ±3% of reading or ±0.015 m/s, whichever is greater |
| Air Velocity Resolution | 0.01 m/s |
| CO Range (982 probe) | 0–500 ppm |
| CO Accuracy | ±3% of reading or ±3 ppm, whichever is greater |
| CO₂ Range (980/982 probes) | 0–5000 ppm |
| CO₂ Accuracy | ±3% of reading or ±50 ppm, whichever is greater |
| CO₂ Resolution | 1 ppm |
| Relative Humidity Range | 0–95% RH |
| RH Accuracy | ±3% RH (at 25 °C) |
| RH Resolution | 0.1% RH |
| Pressure Range (Differential/Static) | −3735 to +3735 Pa |
| Pressure Accuracy | ±1% of reading or ±1 Pa |
| Pressure Resolution | 1 Pa |
| Data Storage Capacity | 26,500 data points across 100 user-defined groups |
| Operating Temperature | 5–45 °C |
| Dimensions | 9.7 × 21.1 × 5.3 cm |
| Weight (with batteries) | 0.36 kg |
| Power | 4 × AA batteries or optional AC adapter |