YMC
Filter
Showing all 5 results
| Brand | YMC |
|---|---|
| Origin | Japan |
| Model | DAU-50 |
| Maximum Operating Pressure | 10 MPa |
| Internal Diameter | 50 mm |
| Column Length | 700 mm |
| Bed Length Range | 100–400 mm |
| Axial Compression Mechanism | Hydraulic/Pneumatic Piston with Continuous Dynamic Load |
| Optional Slurry Delivery System Capacity | 2 L |
| Instrument Type | Medium-Low Pressure Preparative Liquid Chromatography System |
| Application Level | Laboratory-Scale Purification |
| Brand | YMC |
|---|---|
| Origin | Japan |
| Model | K-Prep EX |
| Application Level | Industrial-Scale Preparation |
| Instrument Type | Medium-to-Low Pressure Preparative LC System |
| Flow Rate Range | 30–3000 mL/min (EX03KG) |
| Flow Accuracy | ±2% (at ≥300 mL/min for EX03KG |
| Flow Precision | 0.5% RSD |
| Maximum Operating Pressure | 10 MPa |
| Wavelength Range | 195–600 nm |
| Wavelength Repeatability | ±1 nm |
| UV/VIS Detector | Fixed-wavelength or scanning, 1 nm resolution |
| Fraction Collection | 5-channel valve-switched system with drain line |
| Solvent Selection | 4-channel (2 × 2 configuration) |
| Control Interface | FA PC-based control (non-hazardous area), PLC-integrated architecture |
| Software Compliance | Developed per MHLW Japan’s “Guideline on Validation of Computerized Systems (CSV)” |
| Electrical Supply | 3-phase AC 200 V, 50/60 Hz (4.0–23 kW depending on model) |
| Construction Materials | High-purity SUS316 stainless steel tubing (ID: 7.5 mm to 22.1 mm), explosion-proof rated enclosures (ATEX / IECEx / Class I Div 1 compliant) |
| Brand | YMC |
|---|---|
| Origin | Japan |
| Model | K-Prep FC |
| Application Level | Industrial Preparative |
| Instrument Type | Medium-Low Pressure Preparative LC |
| Flow Rate Range | 1–750 mL/min |
| Flow Rate Reproducibility | ≤±2% (≥75 mL/min) |
| Maximum Pressure | 10 MPa |
| Wavelength Range | 195–600 nm |
| Gradient Capability | Isocratic & Gradient (FC750G / FC1500G variants) |
| Detection | UV/VIS |
| Pump Technology | Dual-plunger linear cam, pulse-free delivery |
| Column ID Compatibility | φ50–φ150 mm |
| Software Compliance | CSV-aligned per MHLW Japan “Guideline on Computerized Systems in GxP Environments” |
| Regulatory Alignment | Supports IQ/OQ protocols |
| Brand | YMC |
|---|---|
| Origin | Japan |
| Model | K-Prep LAB |
| Application Level | Laboratory-scale |
| Instrument Type | Medium-to-Low Pressure Preparative LC System |
| Flow Rate Range | 1–100 mL/min (LAB-100S/G) / 1–300 mL/min (LAB-300S/G) |
| Flow Accuracy | ±2% (≥10 mL/min for LAB-100 |
| Maximum Operating Pressure | 15 MPa (LAB-100 series) / 10 MPa (LAB-300 series) |
| UV Detection Wavelength Range | 195–370 nm |
| Gradient Capability | Dual-pump high-precision gradient formation |
| Sample Loop Capacity | 50 mL or 100 mL (configurable) |
| Injection Volume | Up to 25 mL via syringe (expandable with pump-assisted loading) |
| System Control | PC-based PLC-integrated software with audit trail |
| Compliance | Designed for GLP/GMP environments per FDA 21 CFR Part 11 and ISO/IEC 17025-aligned validation protocols |
| Column ID Compatibility | 20–200 mm |
| Packing Particle Size Support | 5–10 µm |
| Dead Volume | Minimized via motor-driven ball valves and low-ID stainless steel/FEP fluidic pathways |
| Power Supply | AC 100–240 V, 50/60 Hz |
| Power Consumption | 0.6 kW (single-pump) / 1.1 kW (dual-pump) |
| Brand | YMC |
|---|---|
| Origin | Japan |
| Model | LC-Forte/R |
| Application Level | Laboratory-scale |
| Instrument Type | Medium- to Low-Pressure Preparative LC System |
| Flow Rate Range | 0.1–50.00 mL/min |
| Flow Accuracy | ±2% (at ≥5 mL/min) |
| Flow Precision | ≤0.5% RSD |
| Maximum Pressure | 30 MPa |
| Wavelength Range | 190–400 nm (programmable, 1 nm step) |
| Wavelength Repeatability | ±0.5 nm |
| Baseline Noise | ≤1.3 × 10⁻⁴ AU (254 nm, 1 s response) |
| Data Acquisition Rate | 0.5 s per point |
| Detection Modes | UV-Vis (triple-wavelength simultaneous), optional RI detection |