- All
- Favorite
- Popular
- Most rated
| Brand | Aliben |
|---|---|
| Origin | Sichuan, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Category | Domestic |
| Model | UVP-TOF MS 2000 LITE |
| Detection Principle | Vacuum Ultraviolet Photoionization (VUV-PI) coupled with Time-of-Flight Mass Spectrometry |
| Instrument Type | Online Analyzer |
| Configuration | Multi-Component VOC & Inorganic Gas Monitor |
| Detection Range | 0–1 ppbv (extendable to % v/v) |
| Mass Resolution (FWHM) | >2000 |
| Accuracy | RSD <3% |
| Response Time | <1 s |
| Brand | Aliben |
|---|---|
| Origin | Sichuan, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Origin Category | Domestic |
| Model | UVP-TOF MS 2000 PLUS |
| Detection Principle | Vacuum Ultraviolet Photoionization + Time-of-Flight Mass Spectrometry |
| Instrument Type | Online Analyzer |
| Configuration | Multi-Component VOC Monitor |
| Sampling Method | Direct Gas Inlet |
| Detection Range | 100 pptv – 1 ppmv |
| Mass Resolution (FWHM) | >2500 |
| Accuracy (RSD) | <3% |
| Response Time | <0.5 s |
| Brand | Aliben |
|---|---|
| Origin | Sichuan, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Category | Domestic Instrument |
| Model | PTR-TOF MS 6000 |
| Detection Principle | Proton Transfer Reaction (H₃O⁺-based soft ionization) |
| Instrument Type | Online Real-time Analyzer |
| Configuration | Integrated VOCs Monitoring System |
| Mass Resolution (m/Δm) | ≥ 6,000 (FWHM at m/z 100) |
| Mass Range | m/z 10–500 |
| Detection Limit | < 10 pptv (for common VOCs, 1-s integration) |
| Response Time | < 100 ms |
| Vacuum System | Multi-stage differential pumping with turbomolecular pumps |
| Ion Source | Radiofrequency-guided H₃O⁺ primary ion source |
| Analyzer | Orthogonal acceleration reflectron TOF mass analyzer |
| Data Acquisition Rate | Up to 10 spectra per second |
| Brand | Aliben |
|---|---|
| Origin | Sichuan, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Category | Domestic Instrument |
| Model | PTR-TOF MS 8000 |
| Detection Principle | Proton Transfer Reaction (H₃O⁺-based soft ionization) |
| Instrument Type | Online Real-Time Analyzer |
| Configuration | Integrated VOCs Monitoring System |
| Mass Resolution (m/Δm) | ≥ 8,000 (FWHM at m/z 100) |
| Mass Range | m/z 10–500 |
| Detection Limit | < 10 pptv (for common VOCs, 1-s integration) |
| Data Acquisition Rate | Up to 10 Hz full-spectrum acquisition |
| Vacuum System | Multi-stage differential pumping with turbomolecular pumps |
| Ion Source | RF-driven hollow cathode H₃O⁺ generator |
| Mass Analyzer | Orthogonal acceleration reflectron TOF |
| Software Platform | ALIBEN Acquire v3.2 with real-time spectral deconvolution and compound identification engine |
| Brand | Aliben |
|---|---|
| Origin | Sichuan, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Category | Domestic Instrument |
| Model | PTR-TOF MS 2500 |
| Detection Principle | Proton Transfer Reaction (H₃O⁺-based soft ionization) |
| Instrument Type | Online Real-Time Analyzer |
| Configuration | Integrated VOC Monitoring System |
| Mass Resolution (m/Δm) | ≥ 6,000 (FWHM at m/z 100) |
| Mass Range | m/z 10–300 |
| Detection Limit | < 10 pptv (for formaldehyde, 1 s integration) |
| Data Acquisition Rate | Up to 10 Hz full-spectrum |
| Vacuum System | Multi-stage differential pumping with turbomolecular pumps |
| Ion Source | RF-driven hollow cathode H₃O⁺ generator |
| Mass Analyzer | Orthogonal acceleration reflectron TOF |
| Software Platform | ALIBEN Control Suite v4.2 (GLP-compliant audit trail, 21 CFR Part 11 ready) |