BrandANJIE
OriginShanghai, China
ModelAJ-3700
Wavelength Accuracy±0.2 nm
Wavelength Repeatability0.1 nm
Baseline Noise±0.0002 Abs
Baseline Drift±0.0005 Abs (30 min)
Linearity (r)>0.9995
Detection LimitsNH₃-N: 0.005 mg/L
Total N0.02 mg/L
S²⁻0.005 mg/L
NO₂⁻-N0.003 mg/L
NO₃⁻-N0.006 mg/L
Kjeldahl N0.02 mg/L
Quantification RangesNH₃-N: 0.02–80 mg/L
Total N0.1–160 mg/L
S²⁻0.02–200 mg/L
NO₂⁻-N0.012–200 mg/L
NO₃⁻-N0.03–160 mg/L
Kjeldahl N0.08–80 mg/L
Precision0.5–1.0% RSD (at specified concentrations)
Measurement Time100–280 s (including online digestion for total nitrogen)