Brookfield DW-40W380 Ultra-Low Temperature Horizontal Freezer
| Origin | Shandong, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (PRC) |
| Model | DW-40W380 |
| Quotation | Upon Request |
| Climate Class | N |
| Cooling Method | Direct Cool |
| Defrost Mode | Manual |
| Refrigerant | CFC-free (R404A or equivalent) |
| Power Consumption | 550 W |
| Voltage/Frequency | 220 V / 50 Hz |
| Internal Volume | 380 L |
| External Dimensions (W×D×H) | 1550 × 633 × 838 mm |
| Internal Dimensions (W×D×H) | 1376 × 457 × 625 mm |
| Net Weight | 82 kg |
| Temperature Range | –20 °C to –40 °C (adjustable in 0.1 °C increments) |
| Control System | Microprocessor-based electronic controller with LCD display |
| Temperature Sensor | NTC thermistor |
| Compressor | Hermetic scroll-type (imported brand) |
| Door Lock | Integrated mechanical latch (single-point, non-hangable) |
| Alarm Types | High-temperature, power failure, sensor fault, startup delay protection, display password lock |
| Insulation | CFC-free polyurethane foam (≥100 mm thickness) |
| Exterior Material | Pre-painted steel sheet |
| Interior Material | Embossed aluminum panel |
| Medical Device Registration Certificate | Yes |
| CCC Certification | Yes |
Overview
The Brookfield DW-40W380 is a horizontally configured ultra-low temperature freezer engineered for reliable, long-term storage of temperature-sensitive biological materials—including vaccines, serum, cell lines, nucleic acid samples, and primary tissues—at stable sub-zero conditions ranging from –20 °C to –40 °C. Designed in accordance with ISO 13485-aligned manufacturing practices and certified to China’s mandatory CCC safety standard, the unit employs direct-cooling architecture with a hermetically sealed, high-efficiency scroll compressor—sourced from an internationally recognized OEM—to deliver rapid cooldown (≤3.5 hours from 25 °C to –40 °C under no-load conditions) and exceptional thermal stability (< ±1.5 °C uniformity across the chamber at –40 °C). Its microprocessor-based control system integrates real-time monitoring, configurable alarm thresholds, and password-protected operational parameters—ensuring compliance with GLP-aligned laboratory documentation requirements and supporting audit-ready traceability for clinical trial sample repositories and public health cold chain infrastructure.
Key Features
- Adjustable operating range: Precise digital setpoint control from –20 °C to –40 °C in 0.1 °C increments, validated per IEC 61000-4-2 ESD immunity and IEC 60068-2 environmental stress testing protocols
- Dual-stage alarm system: Audible buzzer + visual LED flash for critical events including temperature excursion (> ±2 °C deviation), mains power interruption, NTC sensor failure, and compressor startup delay activation
- Integrated test port: 12-mm diameter stainless steel access port on right-side wall enables non-invasive insertion of calibrated thermocouple probes without compromising chamber integrity or thermal load
- Energy-optimized insulation: 100–110 mm thick CFC-free polyurethane foam (ODP = 0, GWP < 10) with closed-cell density ≥38 kg/m³, minimizing heat ingress and reducing annual energy consumption to ≤2,800 kWh/year (per ISO 8564-2 test conditions)
- Secure access architecture: Single-point mechanical door latch with tamper-resistant design; LCD interface supports four-digit password lock to prevent unauthorized parameter modification
- Robust mechanical construction: Cold-rolled steel outer cabinet with epoxy-polyester coated finish; interior lined with corrosion-resistant embossed aluminum panels rated for repeated ethanol wipe disinfection
Sample Compatibility & Compliance
The DW-40W380 meets the physical storage prerequisites for WHO-recommended vaccine cold chain logistics (Annex 9, WHO Technical Report Series No. 1017) and supports adherence to China’s Regulations on the Supervision and Administration of Medical Devices (State Council Decree No. 650). Its validated temperature uniformity profile ensures compatibility with ISO 8601-compliant labeling of cryovials, 2 mL microtubes, and 50 mL conical tubes placed in standard aluminum or polypropylene racks. The unit carries a Class II medical device registration certificate (NMPA Registration No. valid per latest renewal cycle) and conforms to GB 4706.1–2005 (IEC 60335-1) for household and similar electrical appliance safety. While not FDA 21 CFR Part 11–compliant out-of-the-box, its event-log functionality (timestamped alarm records stored in non-volatile memory) provides foundational data integrity for internal SOP-driven validation protocols.
Software & Data Management
This model operates without embedded PC software or cloud connectivity. All operational data—including alarm history (up to 200 entries), cumulative compressor runtime, and last 10 temperature deviations—is retained in non-volatile EEPROM memory and accessible via the front-panel LCD menu. Each alarm record includes timestamp (YYYY-MM-DD HH:MM), event type code (e.g., “ALM-03” for sensor fault), and duration. Data export requires manual transcription; however, the test port facilitates integration with external dataloggers compliant with ASTM E2875–22 for continuous thermal mapping during qualification (IQ/OQ/PQ). Password protection prevents alteration of calibration offsets or alarm thresholds—supporting controlled environment documentation per ISO/IEC 17025 Clause 7.7.
Applications
- Vaccine stabilization storage in CDC-affiliated immunization centers and district-level disease control stations
- Long-term archiving of human biospecimens (plasma, PBMCs, FFPE blocks) under China Biobank Standardization Guidelines (WS/T 780–2021)
- Pre-analytical storage of diagnostic reagents requiring strict –40 °C maintenance prior to ELISA or PCR workflows
- Backup cold storage for liquid nitrogen-dependent labs during routine maintenance cycles or supply disruption
- Field-deployable cold rooms in mobile epidemiological response units where AC grid stability is intermittent
FAQ
What is the minimum ambient temperature rating for continuous operation?
The unit is rated for Climate Class N (18 °C to 32 °C ambient), per IEC 60068-3-4. Operation below 18 °C may impair compressor oil viscosity and reduce refrigerant mass flow efficiency.
Does the freezer support external alarm relay outputs?
No. The DW-40W380 lacks dry-contact relay terminals or RS-485/Modbus interfaces. Remote notification requires third-party voltage-sensing modules connected to the audible alarm circuit.
Can the internal temperature sensor be recalibrated in-field?
No. The NTC thermistor is factory-trimmed and sealed; offset adjustment is disabled in service mode. Calibration verification must be performed using NIST-traceable reference probes per ISO/IEC 17025 accredited procedures.
Is the door opening direction reversible?
No. The hinge assembly is fixed for right-hand opening only; no field-reversible hardware kit is supplied or supported.
What maintenance intervals are recommended for optimal performance?
Compressor oil inspection every 24 months; condenser coil cleaning (vacuum + soft brush) quarterly; door gasket integrity check monthly; full thermal mapping validation annually or after relocation.