Empowering Scientific Discovery

Eutech pH450 Portable pH/ION Meter

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Thermo Fisher
Origin Singapore
Manufacturer Eutech Instruments (a Thermo Fisher Scientific brand)
Type Portable Multi-parameter Electrochemical Analyzer
Model pH450
Measurement Range –2.00 to 16.00 pH
Resolution 0.01 pH
Accuracy ±0.01 pH
Calibration Points Up to 5 (USA/NIST), up to 6 (DIN), or up to 5 user-defined points
Buffer Recognition Automatic (USA/NIST/DIN)
Calibration Expiry Alert Configurable
Data Storage 500 datasets with timestamp (RTC)
Connectivity USB and RS232
Power 2 × AA batteries (typical life >2000 hours) or optional AC adapter
Enclosure Rating IP67
Electrode Interface BNC
Temperature Unit °C/°F selectable
GLP Compliance Yes (time-stamped calibration & measurement records)

Overview

The Eutech pH450 is a high-precision, portable multi-parameter electrochemical meter engineered for reliable field and laboratory pH, mV (ORP), and ion-selective electrode (ISE) measurements. Built on a robust potentiometric measurement platform, it employs high-impedance input circuitry (>1012 Ω) to ensure stable readings with standard glass pH electrodes and low-current ISEs. Its core architecture adheres to fundamental electrochemical principles defined in ASTM D1293 (pH of water) and ISO 7888 (water quality — determination of pH), supporting traceable, reproducible quantification across diverse aqueous matrices. Designed for operational continuity in demanding environments, the pH450 integrates IP67-rated ingress protection—fully dust-tight and submersible to 1 m for 30 minutes—making it suitable for wastewater monitoring, environmental sampling, aquaculture, food processing QA, and educational labs where portability and durability are non-negotiable.

Key Features

  • Triple-line LCD display showing real-time pH (or mV/ISE), temperature, and electrode % slope simultaneously—enabling immediate assessment of sensor health and thermal compensation status.
  • Dedicated electrode holder slot allows rapid, vibration-resistant mounting onto beakers, flasks, or process vessels without clamps or stands.
  • Intelligent auto-buffer recognition for USA, NIST, and DIN buffer series eliminates manual buffer selection errors and accelerates calibration workflows.
  • Flexible calibration architecture: supports up to five-point auto-calibration (USA/NIST), six-point (DIN), or five-user-defined points with manual entry—ideal for non-standard buffers or specialized applications such as high-ionic-strength brines or low-conductivity ultrapure water.
  • Configurable calibration expiry timer with visual flagging ensures adherence to internal QA protocols and regulatory expectations for instrument verification intervals.
  • Auto-read stabilization detection with manual lock and auto-hold functions improves repeatability during kinetic or unstable sample measurements.
  • Built-in real-time clock (RTC) stamps all stored data and calibration events with date/time—meeting GLP documentation requirements for audit readiness.

Sample Compatibility & Compliance

The pH450 accepts any standard BNC-compatible pH, ORP, or monovalent/multivalent ISE sensors—including combination electrodes, refillable gel-filled, and solid-state variants—ensuring compatibility with legacy and next-generation probes. It operates across a wide temperature range (0 to 50 °C) with automatic or manual temperature compensation (ATC/MTT). All measurement outputs comply with ISO/IEC 17025 documentation guidelines when used with validated electrodes and certified buffers. The device’s GLP-compliant data logging—including operator ID fields (via optional software), calibration history, and sensor diagnostics—supports traceability in regulated environments governed by FDA 21 CFR Part 11 (when paired with Eutech’s PC-based LogEasy software and audit-trail-enabled configuration).

Software & Data Management

Data export is supported via USB or RS232 interfaces using Eutech’s LogEasy PC software (Windows-compatible), enabling batch download, CSV export, graphical trend analysis, and report generation with customizable templates. Stored datasets include pH/mV/ISE value, temperature, date/time stamp, electrode slope, offset, and calibration buffer IDs. The software enforces user-level access control and maintains immutable audit trails for calibration edits and data exports—aligning with GMP and pharmaceutical QC workflows requiring full data integrity per ALCOA+ principles.

Applications

  • Environmental field testing: surface water, groundwater, effluent, and soil leachate pH profiling under EPA Method 9040C and ISO 10523.
  • Food & beverage QA: titratable acidity verification, fermentation monitoring, and rinse-water validation in HACCP-aligned production lines.
  • Aquaculture & hydroponics: real-time pH and nutrient ion (e.g., NH4+, NO3) tracking to maintain optimal growth conditions.
  • Educational laboratories: teaching core concepts of electrode kinetics, Nernst equation application, and calibration uncertainty propagation.
  • Pharmaceutical water systems: conductivity-coupled pH checks for Purified Water (PW) and Water for Injection (WFI) per USP <791> and EP 2.2.3.

FAQ

Does the pH450 support ISE measurements with concentration units (e.g., ppm, mol/L)?
Yes—it calculates and displays ion concentrations directly when calibrated with at least two standards of known activity, applying the Nernstian slope and selectivity coefficient inputs via the ISE setup menu.
Can calibration data be exported separately from measurement logs?
Yes—LogEasy software permits selective export of calibration reports including slope, offset, buffer IDs, and timestamps, formatted for inclusion in SOP documentation or audit submissions.
Is the IP67 rating maintained when the battery compartment is open?
No—the IP67 rating applies only when both the main housing and battery cover are fully sealed with OEM gaskets intact; routine battery replacement requires resealing verification.
What is the maximum storage duration for logged data before overwrite?
The internal memory retains up to 500 datasets indefinitely until manually cleared or overwritten; no time-based auto-erase function is implemented.
Does the meter perform automatic temperature compensation for ISE measurements?
Yes—ATC is applied to all ISE readings when a compatible ATC probe (e.g., Eutech T-120) is connected, correcting for thermal effects on membrane potential per the extended Nernst equation.