LAUDA Variocool VC 2000W Process Circulating Chiller for Bioprocessing (–20 °C to +80 °C)
| Brand | LAUDA |
|---|---|
| Origin | Germany |
| Model | VC 2000W |
| Cooling Method | Water-Cooled |
| Temperature Range | –20 °C to +80 °C |
| Refrigeration Capacity | 2.00 kW (at +20 °C output) |
| Temperature Stability | ±0.05 K |
| Circulating Pump Pressure | 0.9 bar |
| Circulating Pump Flow Rate | 28 L/min |
| Reservoir Volume | 15 L |
| Operating Mode | Continuous |
| Refrigerant | R-449A (GWP 1397) |
| Dimensions (W×D×H) | 450 × 550 × 650 mm |
| Weight | 58 kg |
| Power Supply | 230 V, 14 A, 3.3 kW max |
Overview
The LAUDA Variocool VC 2000W is a high-performance, water-cooled process circulating chiller engineered specifically for demanding bioprocess temperature control applications. Designed and manufactured in Germany, it operates across a broad thermal range of –20 °C to +80 °C with exceptional stability (±0.05 K), making it suitable for critical tasks including single-use bioreactor jacket cooling/heating, upstream cell culture temperature management, downstream cold storage conditioning, and environmental simulation in pharmaceutical development labs. Its thermodynamic architecture integrates a high-efficiency vapor-compression refrigeration cycle with R-449A — a low-GWP (1397), non-ozone-depleting refrigerant compliant with EU F-Gas Regulation (EU No 517/2014) and aligned with global sustainability mandates. The unit employs a dual-mode thermal regulation strategy: precise PID-controlled electric heating combined with digitally modulated compressor staging and SmartCool™ adaptive energy management, ensuring minimal thermal overshoot and consistent setpoint adherence under dynamic load conditions typical of fed-batch or perfusion bioreactors.
Key Features
- Compact, integrated design (450 × 550 × 650 mm; 58 kg) optimized for cleanroom-adjacent utility rooms or modular bioprocess skids
- High-flow, pressure-stable circulation system: 28 L/min at 0.9 bar, supporting large-volume jackets and multi-point distribution manifolds
- TFT color display with real-time temperature curve visualization, dual-value readout (setpoint/actual), and intuitive text-based menu navigation in six languages (English, German, French, Spanish, Italian, Russian)
- Electronic continuous controller with auto-tuning PID algorithm and programmable ramp/soak profiles (up to 150 segments)
- SmartCool™ digital cooling management system dynamically adjusts compressor speed and condenser water flow to match thermal load — reducing energy consumption by up to 30% versus fixed-speed chillers
- Non-flammable fluid compatibility: fully operational with deionized water across >90% of its operating range (e.g., stable 5 °C control without glycol), eliminating handling hazards and enabling use in food-grade and GMP-compliant environments
- Comprehensive safety architecture: electronic level monitoring with low-fluid alarm, adjustable pressure bypass, overtemperature cut-off, and refrigerant leak detection readiness
- Standard USB interface for local data logging and firmware updates; expandable via optional interface modules (RS-485, Profibus DP, Ethernet/IP, analog 0–10 V / 4–20 mA, dry-contact I/O)
- Remote operation capability via LAUDA Command Module (cable length up to 50 m), supporting centralized HVAC or SCADA integration in distributed facility layouts
Sample Compatibility & Compliance
The VC 2000W is validated for use with non-flammable heat transfer fluids — primarily purified water and aqueous ethylene glycol solutions — ensuring compatibility with stainless-steel and EPDM/PFA-wetted components common in biopharmaceutical fluid paths. Its design conforms to IEC 61000-6-2 (immunity) and IEC 61000-6-4 (emissions), and meets CE marking requirements under the EU Machinery Directive 2006/42/EC and Low Voltage Directive 2014/35/EU. While not intrinsically rated for hazardous areas, its water-only operation eliminates ignition risk in Class I, Division 2 environments per NEC Article 500. For regulated biomanufacturing, the chiller supports audit-ready documentation packages and can be configured with optional 21 CFR Part 11-compliant software add-ons (via LAUDA’s ProLine+ platform) for electronic signatures, user access control, and immutable audit trails — facilitating alignment with FDA Q5C, ISO 20933, and EMA Annex 15 validation frameworks.
Software & Data Management
Local operation is managed through the embedded controller with onboard program storage and graphical trend logging (temperature, pump status, alarm history). For enterprise-level integration, the optional LAUDA ProLine+ software suite enables remote monitoring, predictive maintenance alerts (based on compressor runtime, delta-T trends, and flow deviation), and export of time-stamped CSV reports compatible with LIMS and MES systems. When paired with RS-485 or Ethernet modules, the VC 2000W supports Modbus TCP/RTU protocols for seamless interoperability with DeltaV, Siemens Desigo, or Rockwell FactoryTalk environments. All configuration changes and alarm events are timestamped and logged with user ID attribution — satisfying GLP/GMP data integrity requirements for traceability and reproducibility.
Applications
- Temperature control of single-use bioreactors (SUBs) and stirred-tank reactors (STRs) during mammalian, microbial, or viral vector production
- Cooling of chromatography columns and ultrafiltration/diafiltration skids during buffer preparation and purification steps
- Environmental chamber and incubator support for cell banking, stability studies, and QC testing per ICH Q5C
- Thermal conditioning of fill-finish equipment, vial washers, and lyophilizer shelves
- Process validation studies requiring documented thermal uniformity, repeatability, and recovery time metrics
- Academic and CRO bioprocess development labs requiring flexible, scalable thermal infrastructure
FAQ
Can the VC 2000W operate reliably at 5 °C using only deionized water?
Yes. Its optimized evaporator design and refrigerant charge allow stable operation down to 0 °C with water alone — eliminating glycol-related viscosity losses, corrosion risks, and disposal costs.
Is remote monitoring supported out-of-the-box?
The base unit includes USB for local diagnostics; remote connectivity requires an optional interface module (e.g., Ethernet-USB or RS-485) and LAUDA ProLine+ software license.
What is the minimum required cooling water supply specification?
A minimum 1 bar differential pressure, 5 L/min flow rate, and inlet temperature ≤15 °C are recommended for optimal condenser performance and sustained 2.0 kW capacity.
Does the chiller meet environmental compliance standards for refrigerant use?
Yes. R-449A has a GWP of 1397 — significantly lower than legacy R-404A (GWP 3922) — and complies with current EU F-Gas phase-down schedules and EPA SNAP program requirements.
How is temperature stability verified during qualification?
Stability is validated per ASTM E2500-13 using NIST-traceable PT100 sensors placed at the outlet port under steady-state load, with data collected at 1-second intervals over ≥30 minutes (±0.05 K maximum deviation).