NuAire 8600 CO₂/O₂ Tri-Gas Incubator

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand NuAire
Origin USA
Model 8600
Chamber Volume 160 L
Internal Dimensions (W×D×H) 518 × 524 × 611 mm
External Dimensions 649 × 685 × 958 mm
Control Interface 5″ × 7″ Full-Color Touchscreen with 2-Month Data Logging
Temperature Range Ambient +5°C to 55°C
Temp Uniformity ±0.20°C @ 37°C
Temp Accuracy ±0.1°C
Temp Recovery Rate 0.12°C/min (avg.)
Temp Resolution 0.1°C
CO₂ Sensor Dual-Wavelength NDIR
CO₂ Range 0.1–20%
CO₂ Accuracy ±0.1%
CO₂ Recovery ≤5 min to achieve ±0.5%/+0.2% of setpoint at 5.0%
CO₂ Resolution 0.1%
O₂ Sensor Zirconia Electrochemical
O₂ Range 0.5–21%
O₂ Accuracy ±0.25%
Filtration Integrated Recirculating ISO Class 5 HEPA System (external to chamber)
Interior Stainless Steel, Rounded Corners, Full-View Glass Inner Door
Humidification Standard Full-Size Water Pan
Shelving 4 Standard Slides (max. 16), Autoclavable Rails & Shelves
Filter Access Front-Mounted, Bottom-Positioned HEPA Intake for Visual Inspection & Tool-Free Replacement

Overview

The NuAire 8600 CO₂/O₂ Tri-Gas Incubator is an engineered platform for advanced cell culture environments requiring precise, independent control of carbon dioxide, oxygen, and temperature—critical for hypoxia research, stem cell expansion, tumor biology, and primary cell maintenance. Built on a robust stainless-steel chassis with a fully rounded internal cavity and seamless welds, the 8600 eliminates microbial harborage points while supporting ISO Class 5 air quality via a closed-loop, recirculating HEPA filtration system. Unlike conventional designs that mount HEPA filters inside the chamber—reducing usable volume and introducing contamination risk—the 8600 locates its ISO 5-certified HEPA intake externally beneath the chamber floor, ensuring uninterrupted workspace and eliminating secondary particulate shedding into the culture environment. Temperature regulation employs a microprocessor-controlled PID algorithm with triple-point calibration verification; CO₂ is monitored continuously using dual-wavelength non-dispersive infrared (NDIR) sensing, and O₂ is measured via high-stability zirconia electrochemical technology—both sensors factory-calibrated and field-verifiable.

Key Features

  • 160 L working volume with optimized airflow dynamics for uniform gas distribution and thermal homogeneity (±0.20°C at 37°C)
  • 5″ × 7″ full-color touchscreen interface with embedded data logging—stores up to 60 days of timestamped temperature, CO₂, and O₂ values with configurable export (CSV)
  • Triple-gas capability: independent, real-time control of CO₂ (0.1–20%, ±0.1%), O₂ (0.5–21%, ±0.25%), and temperature (ambient +5°C to 55°C, ±0.1°C accuracy)
  • Autoclavable stainless-steel shelving and slide rails (up to 16 positions); all interior surfaces electropolished for cleanability and corrosion resistance
  • Front-access, bottom-mounted HEPA intake filter—visible without tools, replaceable without opening the chamber or disrupting ongoing experiments
  • Full-view tempered glass inner door with anti-fog coating and integrated gasket seal minimizes condensation and enables non-invasive observation
  • Integrated full-size water pan provides passive humidification; optional active humidity control module available (not included)

Sample Compatibility & Compliance

The NuAire 8600 supports a broad range of biological samples—including adherent and suspension mammalian cells, organoids, co-cultures, and sensitive primary isolates—without compromising sterility or environmental fidelity. Its stainless-steel, corner-free interior meets ASTM E2912-22 standards for cleanroom-compatible equipment design. The external HEPA system complies with ISO 14644-1 Class 5 requirements for airborne particle concentration (<3,520 particles/m³ ≥0.5 µm). All sensor outputs are traceable to NIST standards, and the incubator’s firmware architecture supports audit-ready operation under GLP and GMP frameworks. While not pre-certified for FDA 21 CFR Part 11, the system’s secure user access levels, electronic signature-capable logging, and tamper-evident data storage facilitate compliance readiness when integrated into validated laboratory workflows.

Software & Data Management

Data acquisition and configuration are managed exclusively through the onboard embedded controller—no external PC required. The touchscreen interface supports multi-user profiles with role-based permissions (e.g., operator, technician, administrator), each with configurable access to calibration, alarm thresholds, and log export functions. Logged parameters include chamber temperature, CO₂ %, O₂ %, relative humidity (derived), door-open events, and filter usage time. Exported CSV files retain UTC timestamps and sensor serial identifiers, enabling traceability across instruments and laboratories. Remote monitoring is achievable via optional Ethernet or RS-485 connectivity (hardware add-on); integration with LIMS or building management systems follows Modbus RTU or TCP protocols.

Applications

  • Hypoxia modeling in cancer biology and ischemic disease research (O₂ tunable from 0.5% to 21%)
  • Embryonic and induced pluripotent stem cell (iPSC) culture under physiologically relevant O₂ tension
  • Long-term neuronal or endothelial co-culture studies requiring stable tri-gas equilibration
  • Regulatory-compliant QC testing of biologics and cell therapies per USP and ISO 5725
  • Microbial culture under controlled microaerophilic conditions (e.g., Helicobacter pylori, Campylobacter jejuni)
  • Pre-analytical sample stabilization prior to flow cytometry, RNA-seq, or metabolic profiling

FAQ

Is the HEPA filter certified to ISO Class 5? Why is it mounted outside the chamber?
Yes—the recirculating HEPA system is independently verified to deliver ISO 14644-1 Class 5 air quality within the chamber. Mounting the filter externally preserves full 160 L volume and prevents filter media from becoming a source of particulates or biofilm during extended use.
Can the O₂ and CO₂ sensors be recalibrated in-house?
Both sensors support two-point field calibration using certified gas standards (NIST-traceable 0% and 21% O₂; 0% and 5% CO₂). Calibration procedures are accessible only to administrator-level users and generate auditable logs.
What is the maximum number of shelves supported, and are they autoclavable?
The system accommodates up to 16 shelves on reinforced slides. All shelves and rails are constructed from 316L stainless steel and validated for repeated autoclaving at 121°C for 20 minutes.
Does the incubator meet UL/CSA safety standards?
Yes—the 8600 carries UL 61010-1 and CSA C22.2 No. 61010-1 certification for laboratory equipment, including overtemperature protection, gas leak detection interlocks, and grounded chassis design.
Is remote software update capability available?
Firmware updates are performed via USB drive using signed binaries. Network-based updates are disabled by default to maintain integrity in regulated environments; enterprise deployment requires documented change control per ISO 13485 Annex A.