Model Expert pH
Instrument Type Pen-style
Parameter Type Single-parameter (pH)
Measurement Accuracy ±0.1 pH
pH Range 0.0–14.0 pH
Resolution 0.1 pH
Calibration Standard NIST-traceable buffer solutions
Temperature Range 0.0–50.0 °C
Temp. Resolution 0.1 °C
Temp. Accuracy ±0.5 °C + 1 LSD
Calibration Points 1–3-point auto-recognized
ATC Yes
IP Rating IP67
Auto-off 8.5 min idle
Power 4 × A76/LR44 batteries
Operating Temp. 5–45 °C
Dimensions 170 × 45 × 30 mm (L×W×H)

Overview

The Thermo Scientific™ Eutech™ Expert pH Pen Meter is a compact, field-deployable electrochemical sensor engineered for rapid, reliable pH and temperature measurement in diverse aqueous environments. Based on potentiometric measurement principles using a glass electrode and integrated reference system, the device delivers NIST-traceable accuracy with automatic temperature compensation (ATC) to correct for thermal effects on electrode response. Its pen-form factor—designed for one-handed operation and portability—makes it suitable for routine monitoring where laboratory-grade precision must coexist with ruggedness and ease of use. Unlike benchtop meters requiring dedicated calibration stations, the Expert pH Pen operates autonomously in uncontrolled ambient conditions while maintaining compliance with fundamental metrological requirements for pH measurement in environmental, agricultural, and quality control workflows.

Key Features

  • Simultaneous real-time display of pH and temperature (°C/°F) on a high-contrast, 4-digit dual-line LCD with intuitive status icons for battery life, measurement readiness, and calibration mode
  • Auto-buffer recognition across standard pH 4.01, 7.00, and 10.01 NIST-certified calibration solutions—eliminating manual buffer selection and reducing operator error
  • Integrated ATC via built-in thermistor ensures pH values are corrected to 25 °C per ISO 3696 and ASTM D1293 conventions without external probes or manual input
  • Ergonomic tactile keypad with optimized button spacing and feedback for gloved-hand operation in wet or outdoor settings
  • Multi-functional protective cap serves as both electrode storage vessel and stable sample cup; enables vertical tabletop placement during measurement
  • IP67-rated housing provides full dust ingress protection and immersion resistance up to 1 m for 30 minutes—validated per IEC 60529—and buoyant design allows safe recovery if dropped into water
  • Low-power architecture with auto-shutdown after 8.5 minutes of inactivity extends battery life across ~2,000 measurements per set of four A76/LR44 cells

Sample Compatibility & Compliance

The Expert pH Pen is validated for direct immersion measurement in low- to medium-conductivity aqueous samples including freshwater aquaculture systems, hydroponic nutrient solutions, swimming pool and spa water, soil leachates, irrigation runoff, and general lab-prepared buffers. It is not intended for non-aqueous solvents, highly viscous fluids, or samples containing suspended solids that may coat the junction. The instrument complies with essential electrical safety standards (IEC 61010-1) and supports traceable calibration through NIST-traceable pH standards. While not certified for GLP/GMP-regulated production environments, its documented repeatability (±0.1 pH), defined ATC algorithm, and stable electrode response meet the technical expectations for field verification under ISO 5667-3 (water sampling) and EPA Method 150.1 (pH determination).

Software & Data Management

This pen-style meter operates as a standalone analog-to-digital converter with no onboard data logging or Bluetooth/Wi-Fi connectivity. All measurements are displayed in real time only; no internal memory or export interface is provided. Users intending to archive results must manually record values or integrate the device into external observation protocols. However, its consistent output format, predictable calibration drift behavior (<0.02 pH/day under controlled storage), and compatibility with standard calibration documentation templates facilitate straightforward inclusion in audit-ready QA records. For laboratories subject to FDA 21 CFR Part 11, supplementary procedural controls—including logbook entries, independent verification checks, and scheduled recalibration intervals—are recommended to satisfy electronic record integrity requirements.

Applications

  • Aquaculture & hatchery management: Monitoring pH stability in recirculating aquaculture systems (RAS) and larval rearing tanks
  • Horticulture & greenhouse operations: Verifying nutrient solution pH prior to fertigation in hydroponics and soilless media
  • Municipal & recreational water management: On-site verification of disinfection efficacy via pH-dependent chlorine speciation in pools and spas
  • Environmental field surveys: Rapid assessment of surface water acidification trends in streams, ponds, and wetlands
  • Academic teaching labs: Introducing core electrochemical concepts—including electrode potential, liquid junction potential, and temperature dependence—using a durable, low-cost platform
  • Small-scale food processing: Spot-checking rinse water, brine solutions, and cleaning-in-place (CIP) effluent pH during sanitation validation

FAQ

Does the Expert pH Pen support two-point or three-point calibration?
Yes—it automatically recognizes and applies calibration at one, two, or three points using NIST-traceable buffers (pH 4.01, 7.00, and 10.01).
Is the electrode replaceable or serviceable?
No—the pH sensor is permanently sealed within the pen body and not user-replaceable; the unit is designed as a calibrated disposable system with an expected operational lifespan of 12–18 months under typical field use.
Can it measure temperature independently of pH?
Yes—temperature is measured continuously and displayed alongside pH, and can be read in °C or °F regardless of whether pH measurement is active.
What is the recommended storage method when not in use?
Store upright with the protective cap filled with pH 4.01 or 7.00 storage solution to maintain hydration of the glass membrane and reference junction.
Does it meet any international quality standards for field instrumentation?
It conforms to IEC 60529 (IP67), IEC 61010-1 (safety), and delivers performance aligned with ISO 3696 (grade 3 water) and ASTM D1293 for field pH measurement—but carries no formal ISO/IEC 17025 accreditation.