Truelab X2E-290 Desiccator Cabinet
| Brand | Truelab |
|---|---|
| Model | X2E-290 |
| Capacity | 202 L |
| Internal Dimensions (W×D×H) | 598 × 645 × 524 mm |
| External Dimensions (W×D×H) | 600 × 695 × 642 mm |
| Shelf Quantity | 2 pcs |
| Shelf Size | 555 × 580 × 20 mm |
| Humidity Control Range | ≤5% RH |
| Humidity Accuracy | ±3% RH |
| Temperature Accuracy | ±1°C |
| Control System | Humidity Manager V2 (fully automatic, with digital temperature/humidity curve logging) |
| Power Supply | 220 V |
| Avg. Power Consumption | 55 W |
| Max. Power Consumption | 155 W |
Overview
The Truelab X2E-290 Desiccator Cabinet is an engineered solution for long-term storage of moisture-sensitive materials in laboratory, pharmaceutical, and electronics manufacturing environments. Designed on the principle of active desiccation via integrated Peltier-based dehumidification and sealed chamber architecture, it maintains a stable internal relative humidity (RH) of ≤5% RH—critical for preserving hygroscopic reagents, reference standards, calibration weights, lithium battery components, and precision optical elements. Unlike passive silica-gel cabinets, the X2E-290 employs closed-loop feedback control using high-stability capacitive RH sensors and real-time thermal management to ensure consistent low-humidity conditions without manual regeneration or consumable replacement. Its compact footprint (600 × 695 × 642 mm external) makes it suitable for benchtop integration in QC labs, R&D cleanrooms, and controlled storage alcoves compliant with ISO 14644-1 Class 8 environments.
Key Features
- Fully automated humidity regulation via Humidity Manager V2 firmware—enabling continuous monitoring, setpoint maintenance, and deviation alarms
- Digital温湿度 curve logging (temperature & RH vs. time) stored internally for audit-ready traceability; exportable via USB interface
- Double-sealed tempered glass door with magnetic gasket and anti-fog coating for visual inspection without compromising chamber integrity
- Two adjustable stainless-steel shelves (555 × 580 × 20 mm), load-rated to 15 kg per shelf, with corrosion-resistant finish
- Peltier-driven dehumidification system—no refrigerants, no oil, zero ozone impact; designed for continuous 24/7 operation
- Low-power operation (55 W avg., 155 W max) meeting IEC 61000-3-2 harmonic emission limits for laboratory-grade equipment
- Front-panel LCD display showing real-time temperature (±1°C accuracy), RH (±3% RH accuracy), system status, and alarm history
Sample Compatibility & Compliance
The X2E-290 accommodates standard laboratory containers including glass desiccators, polypropylene trays, aluminum weighing boats, and sealed vials up to 120 mm in height. Its interior chamber is constructed from electrophoretically coated steel with non-outgassing epoxy finish, minimizing volatile organic compound (VOC) emissions—validated per ASTM D5116 for low-emission material testing. The cabinet meets mechanical safety requirements per IEC 61010-1 (Edition 3.1) for electrical equipment used in laboratory environments. While not certified as explosion-proof, its design excludes ignition sources and complies with EN 60079-0 general requirements for use in non-hazardous zones. For regulated pharmaceutical applications, the Humidity Manager V2 supports configurable user access levels and electronic audit trails aligned with FDA 21 CFR Part 11 Annex 11 expectations for data integrity.
Software & Data Management
Humidity Manager V2 firmware provides embedded data logging at user-selectable intervals (1 min to 24 h), storing up to 30 days of timestamped temperature and RH records onboard. Data export is supported via standard USB-A port to FAT32-formatted drives; CSV files include columns for datetime, chamber T (°C), RH (%), setpoint, and system state flags (e.g., “Dehumidifying”, “Standby”, “Alarm”). No proprietary software installation is required—raw logs open directly in Excel or LabArchives-compatible ELN platforms. Optional RS-485 Modbus RTU interface (available upon request) enables integration into centralized BMS or MES systems for facility-wide environmental monitoring per ISO/IEC 17025 clause 5.4.6.
Applications
- Storage of USP reference standards requiring ≤5% RH per USP packaging guidance
- Pre-conditioning of samples prior to Karl Fischer titration or gravimetric moisture analysis (ASTM E1064)
- Protection of NIST-traceable weights during calibration interval holding (ISO/IEC 17025:2017 clause 7.8.2)
- Stabilization of humidity-sensitive sensors (e.g., MEMS accelerometers, piezoresistive pressure transducers)
- Holding zone for PCB assemblies pre-reflow to prevent popcorning failure (IPC-J-STD-033D)
- Secure archival of archival-grade film, microfiche, and forensic evidence under controlled dry conditions (NARA Bulletin 2012-02)
FAQ
What is the recommended calibration frequency for the built-in RH sensor?
Annual verification against NIST-traceable portable hygrometer (e.g., Rotronic HC2A-S) is advised per ISO/IEC 17025; factory calibration certificate is provided with each unit.
Can the cabinet operate continuously at ambient temperatures above 35°C?
Yes—thermal management system maintains ≤5% RH up to 35°C ambient; performance degrades gradually above 38°C ambient due to reduced Peltier efficiency; consult technical datasheet for derating curves.
Is remote monitoring possible without physical USB access?
Not natively; however, optional RS-485 Modbus RTU interface allows connection to third-party SCADA or building automation systems for real-time telemetry and alarm notification.
Are replacement shelves available as spare parts?
Yes—stainless-steel shelves (P/N: X2E-SH-555×580) and leveling feet (P/N: X2E-FEET-SET) are stocked and ship within 3 business days globally.
Does the unit comply with EU RoHS and REACH directives?
Yes—full material declarations and RoHS 2 (2011/65/EU) compliance documentation are supplied with CE Declaration of Conformity.